Biomechanical properties of the cornea measured by the Ocular Response Analyzer and their association with intraocular pressure and the central corneal curvature

, PhD & , PhD
Pages 469-475 | Received 12 Feb 2009, Published online: 15 Apr 2021
 

Abstract

Background:  The aim of this study was to investigate the biomechanical properties of the cornea and their association with intraocular pressure (IOP), central corneal thickness (CCT) and the central corneal radius of curvature (Rc).

Methods:  Eighty‐three eyes were divided into two groups. The biomechanical properties of the cornea were measured in 63 normal eyes and in 20 post‐laser in situ keratomileusis (LASIK) eyes. The IOP, corneal hysteresis (CH) and corneal resistance factor (CRF) were measured by the Ocular Response Analyzer (ORA). The Rc and CCT were measured using the corneal topographer Medmont E‐300 and the Tomey SP‐100 Handy ultrasonic pachymeter. Other parameters measured by the ORA, such as TimeIn and TimeOut, were also studied.

Results:  A mean corneal hysteresis of 10.8-mmHg and CRF of 10.6-mmHg were recorded for the normal eyes. There was no significant association with central curvature. All parameters measured by the ORA showed a significant correlation with the CCT, except for the corneal‐compensated intraocular pressure (IOPcc). Both IOPs measured by the ORA had the same values for the mean CH and CRF. For the post‐LASIK eyes, the CH and CRF were lower than in the normal non‐operated eyes. The TimeIn and the TimeOut also presented lower values for the post‐LASIK eyes, suggesting that additional data can be obtained with the ORA measurements.

Conclusions:  The results of this study indicate that there is no correlation between the parameters measured with the Ocular Response Analyzer and central corneal radius of curvature. Some of the biomechanical properties of the cornea studied were found to differ in the normal eyes compared to the post‐LASIK eyes.

ACKNOWLEDGEMENTS

The authors thank Optometron for the loan of the Ocular Response Analyzer and Celeste Krauss MD from MedEdit for the English editing of the manuscript.

Neither author has a financial or proprietary interest in any material or method mentioned in this study.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Purchase Issue

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 84.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.