Table 1 BP ranges for entry BP
Table 2 Age, arm circumference distribution, and BP values for all four devices populations
Table 3 Results of the OMRON M1 Plus (HEM-4011C-E) device
Figure 1 Plots for systolic blood pressure (SBP) difference between the test device readings and the mean of the two observer readings in 33 participants (n = 99) versus the mean of the devices and the mercury sphygmomanometer readings: (a) Omron M1 Plus (HEM 4011C-E), (b) Omron M6 Comfort (HEM-7000-E), (c) Spengler KP7500D, (d) Microlife BP A100 Plus.
![Figure 1 Plots for systolic blood pressure (SBP) difference between the test device readings and the mean of the two observer readings in 33 participants (n = 99) versus the mean of the devices and the mercury sphygmomanometer readings: (a) Omron M1 Plus (HEM 4011C-E), (b) Omron M6 Comfort (HEM-7000-E), (c) Spengler KP7500D, (d) Microlife BP A100 Plus.](/cms/asset/98ab6144-5151-40dc-bb58-cebdf8cce38f/dvhr_a_12187365_f0001_b.jpg)
![Figure 1 Plots for systolic blood pressure (SBP) difference between the test device readings and the mean of the two observer readings in 33 participants (n = 99) versus the mean of the devices and the mercury sphygmomanometer readings: (a) Omron M1 Plus (HEM 4011C-E), (b) Omron M6 Comfort (HEM-7000-E), (c) Spengler KP7500D, (d) Microlife BP A100 Plus.](/cms/asset/c4a682fa-a152-423b-84a0-7daffd94b304/dvhr_a_12187365_f0001a_b.jpg)
![Figure 1 Plots for systolic blood pressure (SBP) difference between the test device readings and the mean of the two observer readings in 33 participants (n = 99) versus the mean of the devices and the mercury sphygmomanometer readings: (a) Omron M1 Plus (HEM 4011C-E), (b) Omron M6 Comfort (HEM-7000-E), (c) Spengler KP7500D, (d) Microlife BP A100 Plus.](/cms/asset/e8e073ba-594f-491c-b9bd-7142c2bc1564/dvhr_a_12187365_f0001b_b.jpg)
![Figure 1 Plots for systolic blood pressure (SBP) difference between the test device readings and the mean of the two observer readings in 33 participants (n = 99) versus the mean of the devices and the mercury sphygmomanometer readings: (a) Omron M1 Plus (HEM 4011C-E), (b) Omron M6 Comfort (HEM-7000-E), (c) Spengler KP7500D, (d) Microlife BP A100 Plus.](/cms/asset/74fdf8e4-7791-4611-9f6a-3f709c4cd46c/dvhr_a_12187365_f0001c_b.jpg)
Figure 2 Plots for diastolic blood pressure (DBP) difference between the test device readings and the mean of the two observer readings in 33 participants (n = 99) versus the mean of the devices and the mercury sphygmomanometer readings: (a) Omron M1 Plus (HEM 4011C-E), (b) Omron M6 Comfort (HEM-7000-E), (c) Spengler KP7500D, (d) Microlife BP A100 Plus.
![Figure 2 Plots for diastolic blood pressure (DBP) difference between the test device readings and the mean of the two observer readings in 33 participants (n = 99) versus the mean of the devices and the mercury sphygmomanometer readings: (a) Omron M1 Plus (HEM 4011C-E), (b) Omron M6 Comfort (HEM-7000-E), (c) Spengler KP7500D, (d) Microlife BP A100 Plus.](/cms/asset/e888bf09-859b-4811-a0c2-09749d63886d/dvhr_a_12187365_f0002_b.jpg)
![Figure 2 Plots for diastolic blood pressure (DBP) difference between the test device readings and the mean of the two observer readings in 33 participants (n = 99) versus the mean of the devices and the mercury sphygmomanometer readings: (a) Omron M1 Plus (HEM 4011C-E), (b) Omron M6 Comfort (HEM-7000-E), (c) Spengler KP7500D, (d) Microlife BP A100 Plus.](/cms/asset/cd51158b-4827-489e-b23c-34e30f3fe14f/dvhr_a_12187365_f0002a_b.jpg)
![Figure 2 Plots for diastolic blood pressure (DBP) difference between the test device readings and the mean of the two observer readings in 33 participants (n = 99) versus the mean of the devices and the mercury sphygmomanometer readings: (a) Omron M1 Plus (HEM 4011C-E), (b) Omron M6 Comfort (HEM-7000-E), (c) Spengler KP7500D, (d) Microlife BP A100 Plus.](/cms/asset/f0e00648-01cc-442c-bb6f-e9058046d3d9/dvhr_a_12187365_f0002b_b.jpg)
![Figure 2 Plots for diastolic blood pressure (DBP) difference between the test device readings and the mean of the two observer readings in 33 participants (n = 99) versus the mean of the devices and the mercury sphygmomanometer readings: (a) Omron M1 Plus (HEM 4011C-E), (b) Omron M6 Comfort (HEM-7000-E), (c) Spengler KP7500D, (d) Microlife BP A100 Plus.](/cms/asset/d358edac-4046-4404-bf37-b2ff9448d17a/dvhr_a_12187365_f0002c_b.jpg)
Table 4 Results of the OMRON M6 Comfort (HEM 7000-E) device
Table 5 Results of the Spengler KP7500D device
Table 6 Results of the Microlife BP A100 Plus device