Understanding the Nuances of Microwave Ablation for More Accurate Post-Treatment Assessment

Pages 1755-1764 | Received 28 Dec 2017, Accepted 29 Jan 2018, Published online: 14 Feb 2018

Figures & data

Figure 1. Underlying mechanism of action and heat transfer for microwave ablation and radiofrequency ablation.

microwave ablation A & C; radiofrequency ablation B & D.

Figure 1. Underlying mechanism of action and heat transfer for microwave ablation and radiofrequency ablation.microwave ablation A & C; radiofrequency ablation B & D.

Table 1. Parameters that affect ablation volumes using microwave ablation and radiofrequency ablation technologies.

Table 2. Summary of average percent differences of in vivo short and long axis diameters relative to ex vivo references for the AMICA and Certus PR Microwave Ablation Systems at various power and time settings.

Figure 2. Comparison of in vivo and ex vivo reference ablation diameters for the AMICA and Certus PR applicators at a specified wattage with respect to time.
Figure 2. Comparison of in vivo and ex vivo reference ablation diameters for the AMICA and Certus PR applicators at a specified wattage with respect to time.
Figure 3. Ex vivo versus in vivo parameters that inherently make accurate comparisons a challenge with the use of MWA technology.
Figure 3. Ex vivo versus in vivo parameters that inherently make accurate comparisons a challenge with the use of MWA technology.