Fixation of trapezial implants in a trapeziometacarpal total joint prosthesis tested in a model of porcine bone

, , &
Pages 263-266 | Accepted 30 Mar 2011, Published online: 18 Jan 2012
 

Abstract

High aseptic loosening rates have been reported in total joint prostheses of the carpometacarpal joint of the thumb, particularly in the trapezial component. The primary fixation of new implants may be tested in cadaver bones, but the anatomy of the pig is in many ways similar to that of the human, so we compared the central carpal bone from the forefoot of 6-month-old pigs, which has a saddle joint surface similar to the trapezium, to the trapezium in patients with carpometacarpal osteoarthritis. The mean (SD) bone mineral density of the 13 pig forefoot bones was 0.88 (0.12) g/cm2 compared with 0.63 (0.16) g/cm2 in the 31 human trapeziums. The measured size of the porcine bones was slightly larger than that of the human trapeziums. The similarity in form, size, and bone mineral density means that the central bone of the forefoot of 6-month-old pigs may be used for fixation tests in trapezial implants.

Declaration of interest: The authors report no conflicts of interest. The authors alone are responsible for the content and writing of the paper.

Log in via your institution

Log in to Taylor & Francis Online

There are no offers available at the current time.

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.