Abstract
A fast, reliable, and low cost biamperometric flow‐injection method, with an error of 1.3% and an analytical throughput of 55 samples h−1, for determination of tetracycline hydrochloride in pharmaceuticals capsules is proposed. The analytical curve was linear (r=0.998) in the range 10 to 50 mg l−1 using Fe(CN)6 3− and NaOH solutions as reagent and carrier stream/supporting electrolyte, respectively. A relative standard deviation of 1.6% (10 sequential injections of 30.0 mg l−1) was verified with detection and quantification limits of 0.6 and 3.4 mg l−1, respectively.
All authors are honoured with CAPES and CNPq scholarship, respectively.