Abstract
A flow injection chemiluminescence (CL) method is described for the determination of trenbolone acetate based on the CL generated during its reaction with KMnO4 in acidic medium. The CL intensity is greatly enhanced by alizarin yellow R. The CL intensity is linear with trenbolone acetate concentration in the range 0.1–100.0 mg L−1 with a detection limit of 0.05 mg L−1. The sample throughout is about 90 h−1 and the relative standard deviation for 2.0 mg L−1 trenbolone acetate solution is 1.5% (n = 11). The proposed method was applied to the determination of trenbolone acetate in cattle feeds.
Acknowledgments
This paper was submitted as part of a Special Issue on Flow Injection Analysis.
The authors gratefully acknowledge financial support from Xi'an Jiaotong University (Grant No. 08140009).
Notes
a I is the CL signal in presence of 2.5 mg L−1 trenbolone acetate and I b is the blank signal, ΔI = I − I b .
b n = 3.