A Comparison of Paired Calibration Check Results of Alco-Sensor IV- RBT IV and Intoxilyzer® 5000 C in Real Cases

Pages 112-119 | Published online: 22 Nov 2013
 

Abstract

More than 1600 paired-control tests data obtained on Alco-Sensor IV—RBTIV and Intoxilyzer® 5000 C have been reviewed. A quasi-normal distribution around the target value of 100 mg/100 mL was observed. Moreover, not only are instruments accurate, but they are also stable. In sequences of tests in real cases, paired data comparison of first (CTRL1) and second (CTRL2) control results showed a difference (CTRL1—CTRL2) of zero in about one-third (⅓) of assays. The majority of cases (98.5%) were found to have a cumulative absolute difference of 3 mg/100 mL. In just one case, the difference between CTRL1 and CTRL2 was 10 mg/100 mL, the maximum acceptable in Quebec. Thus, although calibration checks of 95 to 105 mg/100 mL are acceptable, measurement uncertainty of these instruments should be considered well below 10 mg/100 mL in the great majority of cases.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.