Abstract
Hemoglobin A1c (HbA1c) is a frequently requested laboratory test and there is thus a need for high throughput instruments for this assay. We evaluated a new automated multicapillary zone electrophoresis instrument (Capillarys 3 Tera, Sebia, Lisses, France) for analysis of HbA1c in venous samples. Routine requested HbA1c samples were analyzed immunologically on a Roche c6000 instrument (n = 142) and then with the Capillarys 3 Tera instrument. The Capillarys 3 Tera instrument performed approximately 70 HbA1c tests/hour. There was a strong linear correlation between Capillarys 3 Tera and Roche Tina-Quant HbA1c Gen 3 assay (y = 1.003x – 0.3246 R2 = .996). The total CV for the 12 capillaries varied between 0.8 and 2.2% and there was a good agreement between duplicate samples (R2 = .997). In conclusion, the Capillarys 3 Tera instrument has a high assay capacity for HbA1c. It has a good precision and agreement with the Roche Tina-Quant HbA1c method and is well suited for high volume testing of HbA1c.
Acknowledgements
The HbA1c reagents for the Capillarys 3 Tera evaluation was generously provided by ACC Nordic.
Disclosure statement
The authors report no conflicts of interest. The authors alone are responsible for the content and writing of this article.