Abstract
Roche Diagnostics® decided a more restrictive haemolysis index (≤20 HI) for approval the release of serum potassium results. This study examined the risk of overestimating serum potassium results related to the HI according to the Ricos total error (±5.6%) and evaluated the approach of Martinez-Morillo and Alvarez to determine a corrected potassium. According to Ricos’ criteria, our study showed compliant potassium results with HI less than or equal to 75. Between 90 and 100 HI, the results did not show a significant overestimation. The equation to obtain Kcorrected was: Kmeasured – (0.004 × HI). The use of corrective formulas for adjusting results of potassium could help the laboratory to identify patients at increased risk and to repeat the test as soon as possible.
Disclosure statement
No potential conflict of interest was reported by the authors.