Verifying measurements on Siemens Atellica® instruments using clinically acceptable analytical performance specifications

, , , , & ORCID Icon
Pages 408-416 | Received 14 Jun 2023, Accepted 26 Aug 2023, Published online: 06 Sep 2023
 

Abstract

Measurements on clinical chemistry analysers must be verified to demonstrate applicability to their intended clinical use. We verified the performance of measurements on the Siemens Atellica® Solution chemistry analysers against the clinically acceptable analytical performance specifications, CAAPS, including the component of intra-individual biological variation, CVI. The relative standard uncertainty of measurement, i.e. analytical variation, CVA, was estimated for six example measurands, haemoglobin A1c in whole blood (B-HbA1c), albumin in urine (U-Alb), and the following measurands in plasma: sodium (P-Na), pancreatic amylase (P-AmylP), low-density lipoprotein cholesterol (P-LDL-C), and creatinine (P-Crea). Experimental CVA was calculated from single-instrument imprecision using control samples, variation between measurements on parallel instruments, and estimation of bias with pooled patient specimens. Each obtained CVA was compared with previously developed CAAPS. The calculated CVA was 1.4% for B-HbA1c (CAAPS 1.9% for single diagnostic testing, CAAPS 2.0% for monitoring after duplicate tests; IFCC units), 10.9% for U-Alb (CAAPS 44.9%), 1.2% for P-Na (CAAPS 0.6%, after triplicate testing 1.5%), 8.2% for P-AmylP (CAAPS 22.9%). The CVA was 4.9% for P-LDL-C (CAAPS for cardiovascular risk stratification 4.9% after four replicates), and 4.2% for P-Crea (CAAPS 8.0%). Three of the six measurands fulfilled the estimated clinical need. Results from P-Na measurements indicate a general need for improving the P-Na assays for emergency patients. It is necessary to consider CVI when creating diagnostic targets for laboratory tests, as emphasised by the CAAPS estimates of B-HbA1c and P-LDL-C.

Disclosure statement

No potential conflict of interest was reported by the authors.

Siemens Healthineers did not influence in the design or the interpretation of this study. No external funding was obtained for this verification which was a part of a developmental process at HUS Diagnostic Centre laboratories. The authors have no competing interests to declare.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 200.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.