The application of the Reaction Rate Analyser LKB 8600 as an automatic coagulometer

&
Pages 7-13 | Received 14 Sep 1976, Accepted 16 Nov 1976, Published online: 28 Aug 2009
 

Abstract

The application of the Reaction Rate Analyser LKB 8600 (RRA) as an automatic coagulometer is described. The RRA was slightly modified without interfering with its function as an enzyme reaction rate analyser. The endpoint of coagulation was recorded when the increase in absorbance exceeded 0.047 at δ = 340 nm. The coagulation time was monitored by a counter with automatic print-out or by a recorder. The analytical dispersion, s(x)x, ranged between 0.01 and 0.05. Results (x, y) of the determination of the quantity of plasma coagulation factors (II + VII+X) by RRA and manually by visual recording could be expressed by y =. 93x+0.04, r = 0.96, n = 66 (method of Owren & Aas) or by y = 1.00x + 0.02, r = 0.98, n = 49 (Normotest). Similarly, plasma activated partial thromboplastin time (Activated Thrombofax) could be expressed by y = 0.99x + 5.00, r = 0.99, n = 23.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.