Direct Determination of Mercury at 0.2 ppm Level by Flame AA Using a Thermospray Atomizer and a Circular Burner

&
Pages 763-777 | Received 15 Feb 1989, Accepted 16 Mar 1989, Published online: 24 Oct 2006
 

Abstract

A circular flame atomizer system using a quartz T flame adapator has been developed for the direct determination of mercury. A thermospray nebulizer was used as a sample introduction device to the atomizer. This design gave a sensitivity (1% absorption) of 0.20 ppm and detection limits of 10 ppb at 1.0 mL/min solvent flow at 253.7 nm. The absolute sensitivity was 25 ng and absolute detection limits was 1.5 ng.