980
Views
25
CrossRef citations to date
0
Altmetric
Scientific Articles

Assessment of different methods to estimate bovine colostrum quality on farm

, , , , , & show all
Pages 263-267 | Received 19 Aug 2015, Accepted 05 Apr 2016, Published online: 26 May 2016
 

Abstract

AIMS: To evaluate two different hydrometers and an optical and a digital Brix refractometer for the assessment of bovine colostrum quality, in terms of accuracy and precision compared with the measurement of IgG concentrations using radial immunodiffusion (RID), and to evaluate the reliability and repeatability of the Brix refractometers.

METHODS: To determine reliability and repeatability, 145 colostrum samples were tested by two independent observers twice, using the optical and digital Brix refractometers. A further 193 colostrum samples from Holstein cows were collected on one commercial dairy farm at first milking and tested with two hydrometers and an optical and digital Brix refractometer. An aliquot of each sample was frozen for RID measurement of IgG concentrations and samples were classified as poor (≤50 g IgG/L) or good (>50 g IgG/L) quality colostrum. Intraclass correlation coefficients (ICC) were used to determine inter- and intra-observer reliability and repeatability. Optimised cut-off values for the four devices were determined using receiver operating characteristics (ROC) analysis with the RID results as the reference. Using these cut-offs, sensitivities and specificities for determining good quality colostrum were calculated.

RESULTS: The ICC for inter-observer reliability was 0.98 for the optical Brix refractometer, and for intra-observer repeatability was 0.97 and 0.98 for the optical and the digital Brix refractometers, respectively. For the 193 colostrum samples, 67 (34.7%) had concentrations of IgG ≤50 g/L determined by RID. Optimised cut-off values evaluated by ROC analysis were higher for all devices compared with manufacturer reference or previously published values. Using these values, the sensitivities for the two hydrometers, and the optical and the digital Brix refractometers were 0.73, 0.71, 0.56 and 0.79, respectively; specificities were 0.72, 0.61, 0.90 and 0.69, respectively.

CONCLUSIONS AND CLINICAL RELEVANCE: The Brix refractometers provided the most accurate assessment of colostrum quality of the devices evaluated, and demonstrated excellent precision in terms of repeatability. To provide optimal health for newborn calves, a sufficient intake of good quality colostrum is essential. The Brix refractometers provide rapid, convenient tools for classification of colostrum quality.

AUC=

Area under the curve

ICC=

Intraclass correlation coefficient

RID=

Radial immunodiffusion

ROC=

Receiver operating characteristics

Acknowledgements

The authors would like to thank Marie-Christine Wiedrich and Dorottya Ivanyos for their assistance in sample collection and Andreas Zaiser for his assistance in laboratory work. Furthermore, the authors would like to thank First Farms A/S (Slovakia) for assistance during the study. The authors gratefully acknowledge the financial support by the Austrian Association for Buiatrics (ÖBG, Vienna, Austria).

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 61.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 213.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.