Transfer Functions and Penetrations of Five Differential Mobility Analyzers for Sub-2 nm Particle Classification

, , , , , , , & show all
Pages 480-492 | Received 11 May 2010, Accepted 04 Aug 2010, Published online: 15 Jan 2011

Abstract

The transfer functions and penetrations of five differential mobility analyzers (DMAs) for sub-2 nm particle classification were evaluated in this study. These DMAs include the TSI nanoDMA, the Caltech radial DMA (RDMA) and nanoRDMA, the Grimm nanoDMA, and the Karlsruhe-Vienna DMA. Measurements were done using tetra-alkyl ammonium ion standards with mobility diameters of 1.16, 1.47, and 1.70 nm. These monomobile ions were generated by electrospray followed by high resolution mobility classification. Measurements were focused at an aerosol-to-sheath flow ratio of 0.1. A data inversion routine was developed to obtain the true transfer function for each test DMA, and these measured transfer functions were compared with theory. DMA penetration efficiencies were also measured. An approximate model for diffusional deposition, based on the modified Gormley and Kennedy equation using an effective length, is given for each test DMA. These results quantitatively characterize the performance of the test DMAs in classifying sub-2 nm particles and can be readily used for DMA data inversion.

1. INTRODUCTION

The differential mobility analyzer (DMA) classifies particles according to their electrical mobility. The history of mobility analyzers can be traced back to measurements of atmospheric ions in the early 20th century (CitationFlagan 1998). The DMA developed in the University of Minnesota Particle Technology Laboratory was a breakthrough both in terms of apparatus design (CitationLiu and Pui 1974) and theory (CitationKnutson and Whitby 1975). Similar to the instrument used by CitationHewitt (1957), this DMA has a coaxial cylinder arrangement with two inlets and two outlets. Since the mid 1970s, DMAs have been used routinely to produce aerosol standards of known size, concentration, and composition, to measure aerosol size distributions, and to measure aerosol properties with tandem measurement instrument systems (CitationPark et al. 2008). The transfer function that was first derived by CitationKnutson and Whitby (1975) was the key development in enabling quantitative applications of DMAs for such purposes. Over the past decades, several other types of DMAs have been developed and used in aerosol research, including the Vienna type (CitationWinklmayr et al. 1991), the French radial type (RDMA, CitationMesbah 1994; CitationOoghe et al. 1994; CitationFissan et al. 1998), the Caltech radial type (CitationZhang et al. 1994, Citation1995; CitationBrunelli et al. 2009), and the TSI nanoDMA (CitationChen et al. 1998). The performance of such DMAs in the 3 to 200 nm size range has been discussed in previous studies (CitationFissan et al. 1996; CitationBirmili et al. 1997; CitationKarlsson and Martinsson 2003). More recently, supercritical cylindrical DMAs that remain laminar in the classification region to Reynolds numbers much higher than 2000, and DMAs of unconventional geometries have been developed to improve the performance and were reviewed by CitationFernández de la Mora (2011).

FIG. 1 Experimental setup for generating monomobile ion standards and for testing DMA performances.

FIG. 1 Experimental setup for generating monomobile ion standards and for testing DMA performances.

The classification of nanoparticles down to molecular dimensions is of special interest for current research. However, the transfer functions of most DMAs for sub-2 nm particle classification have not been systematically evaluated, largely due to the lack of suitable sub-2 nm particle standards. The tandem DMA (TDMA) method has often been used to evaluate DMA transfer functions (CitationStolzenburg and McMurry 1988; CitationFissan et al. 1996; CitationStratmann et al. 1997). In the 1–2 nm size range, diffusional broadening can significantly degrade the DMA resolution such that it is difficult to generate pseudo-monodisperse particles using a conventional DMA as the first DMA (CitationKousaka et al. 1986; CitationStolzenburg 1988). Recent developments in high-flow DMAs significantly improved the DMA resolution in the nanometer size range (CitationRosell-Llompart et al. 1996; CitationFernández de la Mora et al. 1998; CitationRosser and Fernández de la Mora 2005), and enabled the generation of molecular monodisperse mobility standards (tetra-alkyl ammonium ions) (CitationUde and Fernández de la Mora 2005). These monomobile ions, which have mobility diameters of 1–2 nm, have been used as size standards to test DMA resolution (CitationRosser and Fernández de la Mora 2005; CitationSteiner et al. 2010), diffusional particle losses (CitationHeim et al. 2010), and the transmission factors of mass spectrometers (CitationZhao et al. 2010). Another newly developed size standard is oven-generated fullerene (C60) vapor. Monodisperse C60 monomer has a geometrical size around 1 nm and has been used to test DMA performance (CitationTanaka and Takeuchi 2003; CitationIida et al. 2009).

In this study, the transfer functions and penetrations of five differential mobility analyzers for sub-2 nm particle classification were evaluated using tetra-alkyl ammonium ion standards. These DMAs included the TSI nanoDMA (CitationChen et al. 1998), the Caltech RDMA (CitationZhang et al. 1995) and nanoRDMA (CitationBrunelli et al. 2009), the Grimm nanoDMA, and the Karlsruhe-Vienna DMA, an optimized version of the classical Vienna type DMA (CitationWinklmayr et al. 1991; CitationHeim et al. 2005). Calibration was performed using singly charged tetra-propyl ammonium (N+[C3H7]4), tetra-heptyl ammonium (N+[C7H15]4), and tetra-dodecyl ammonium (N+[C12H25]4) ions with mobility diameters of 1.16, 1.47, and 1.70 nm, respectively. These monomobile ions were generated by electrospray and classified by the high resolution DMA (CitationFernández de la Mora et al. 1998; CitationUde and Fernández de la Mora 2005). Based on Stolzenburg's diffusive transfer function (CitationStolzenburg 1988; CitationStolzenburg and McMurry 2008), a data inversion routine was developed to obtain the true transfer function for each test DMA. In addition to transfer functions, we report the DMA penetration efficiencies.

2. EXPERIMENTS

As shown in , an electrospray was coupled to a high resolution differential mobility analyzer (HRDMA) to generate monodisperse size standards (tetra-alkyl ammonium ions). The electrospray system used here has been described in detail by CitationUde and Fernández de la Mora (2005). Particle-free compressed air was used as the carrier gas. The organic salts used to produce these ions (tetra-propyl ammonium iodide (TPAI, N[C3H7]4I), tetra-heptyl ammonium bromide (THABr, N[C7H15]4Br), and tetra-dodecyl ammonium bromide (TDDABr, N[C12H25]4Br)) were dissolved in ultrapure methanol with low concentrations. All the chemicals were purchased and used as received from Sigma Aldrich. Rather than precisely controlling the salt concentrations (CitationUde and Fernández de la Mora 2005), serial dilutions were made to a solution of unknown concentration until the electrospray was stable and the peaks of monomer (A+; A denotes the tetra-alkyl ammonium) and dimer (A+(AB)1; B denotes the halide) were somewhat higher than other peaks. Highly mobile positively charged ions generated by electrospray were then transported to a high-resolution differential mobility analyzer (CitationFernández de la Mora et al. 1998). An inverse mobility distribution of electrosprayed THABr ions measured by scanning HRDMA voltage while using an aerosol electrometer to measure the concentrations of mobility-classified ions is shown in . The HRDMA cleanly separates the monomer (A+), dimer (A+(AB)1), and trimer (A+(AB)2). The identities of these molecular ions have been confirmed by mass spectrometry (CitationUde and Fernández de la Mora 2005). Therefore, monodisperse molecular ions having known chemical structures (e.g., monomer or dimer) can be selected by fixing the HRDMA voltage at the value corresponding to the peak currents shown in . The electrical mobilities of electrosprayed tetra-alkyl ammonium ions have been studied extensively by Fernández de la Mora and coworkers. The equations used by CitationUde and Fernández de la Mora (2005) are used to convert their electrical mobilities to mobility diameters. The monomers of TPAI, THABr, and TDDABr have mobility diameters of 1.16, 1.47, and 1.70 nm, respectively. These ions were used in this study to test DMA performance. For some DMAs, the dimer and trimer of tetra-heptyl ammonium ions were also used. Their mobility diameters are 1.78 and 1.97 nm, respectively. For molecular ions and particles in this size range, mobility diameters differ from the estimated geometric diameters based upon their masses. CitationTammet (1995) discussed the theoretical relationships between these sizes. CitationKu and Fernández de la Mora (2009) carried out measurements and found that for particles with mobility diameters down to 1.6 nm, geometric diameters are 0.3 nm smaller than their mobility diameters. Recent tests show that to within 15%, the 0.3 nm correction applies to geometric diameter down to 0.68 nm (CitationLarriba et al. 2011).

FIG. 2 (a) The measured inverse mobility distribution of electrosprayed THABr ions; (b) Example of measured concentration ratios and fitted N 2/N 1 values using Stolzenburg's diffusive transfer function. The TSI nanoDMA was operated at a flow ratio of 1.5/15 lpm while classifying 1.70 nm TDDABr monomer; (c) The transfer function of the TSI nanoDMA with a fixed DMA voltage for 1.70 nm TDDABr monomer; (d) Measured concentration ratios as a function of the centroid mobility (varying DMA voltage).

FIG. 2 (a) The measured inverse mobility distribution of electrosprayed THABr ions; (b) Example of measured concentration ratios and fitted N 2/N 1 values using Stolzenburg's diffusive transfer function. The TSI nanoDMA was operated at a flow ratio of 1.5/15 lpm while classifying 1.70 nm TDDABr monomer; (c) The transfer function of the TSI nanoDMA with a fixed DMA voltage for 1.70 nm TDDABr monomer; (d) Measured concentration ratios as a function of the centroid mobility (varying DMA voltage).

The selected monodisperse ions were then transported to the test DMAs to measure their transfer functions and penetrations. As shown in , both the upstream ion concentration (N 1) and the downstream ion concentration (N 2) of the test DMAs were monitored by aerosol electrometers. The downstream to upstream electrometer ratios without the test DMA present were experimentally calibrated for every ion and flowrate combination to properly account for the difference in readings between two aerosol electrometers, and for the transport losses from the union cross (P 1) to the test DMA inlet (P 2), from the test DMA outlet (P 3) to the downstream aerosol electrometer inlet (P 4), from the union cross (P 1) to the inlet (P 5) of the upstream aerosol electrometer, and inside aerosol electrometers. The reported concentration ratios (N 2/N 1) account for these corrections. A bypass flow was used to adjust the aerosol outlet flowrate in the HRDMA. The concentration of monodisperse ions before the test DMAs was controlled by adjusting the electrospray operational conditions and the bypass flowrate. Measurements were made over a sufficient range of concentrations to confirm that data were not affected by space charge (CitationCamata et al. 2001). During the DMA calibration, the voltage for the HRDMA was fixed and the voltage for the test DMA was scanned continuously. The DMA voltage was increased slowly to ensure pseudo-steady-state operation. A typical scan took approximately 15 min. Most measurements were done at an aerosol-to-sheath flow ratio of 0.1. Several other aerosol and sheath flow combinations were also examined. Balanced flows were used for all test DMAs, i.e., the aerosol inlet flowrate was equal to the aerosol sampling outlet flowrate. A closed-loop was used for DMA sheath and excess flows to ensure they were exactly equal (CitationRogak et al. 1993; CitationJokinen and Makela 1997).

TABLE 1 Geometry parameters and tested flow conditions for various DMAs

The geometry parameters and the tested flowrates for each DMA are summarized in . Designs of the TSI nano- DMA, the Caltech RDMA and nanoRDMA are described in detail in the references given in . The description of the Karlsruhe-Vienna DMA and the Grimm nanoDMA are not readily available in literature. We discuss the difference between the Karlsruhe-Vienna DMA and the original Vienna DMA in Section 4.5. The Grimm nanoDMA is a commercial instrument, and we were unable to obtain a detailed description of its pertinent design features. The French radial DMA (CitationOoghe et al. 1994) was also tested. However, the use of a high voltage insulator inside this DMA causes high electrostatic transport losses, so that no ions were observed at the outlet. Such electrostatic losses have been reported previously with other DMAs and they become increasingly important as size decreases (CitationKousaka et al. 1986; CitationChen et al. 1998). Similar losses were observed for a copy of the Vienna type DMA (referred to as “DMA 1/40” in CitationWinklmayr et al. 1991). The penetration through the Vienna type DMA was significantly improved by replacing the original insulator with a Delrin insulator, though the results of these two DMAs are not included in this study. The Grimm nanoDMA and the Karlsruhe-Vienna DMA were tested at the Karlsruhe Institute of Technology. The Caltech RDMA and nanoRDMA were evaluated at the University of Minnesota. The TSI nanoDMA was tested at both the University of Minnesota and Karlsruhe Institute of Technology.

Different versions of high-resolution differential mobility analyzers have been developed by Fernández de la Mora and coworkers. The HRDMA used in Minneapolis is also referred as the Herrmann DMA (CitationHerrmann et al. 2000), while the HRDMA used in Karlsruhe is also referred as the Attoui DMA (CitationFernández de la Mora and Attoui 2007; CitationHeim et al. 2010). Both HRDMAs were operated in an open-loop for sheath and excess flows and in a safe mode, i.e., the DMA outer electrode is grounded. The use of a “leaky” insulator (polyoxymethylene, Ensinger-HYDE, Germany) at the aerosol outlet facilitates the safe mode operation (CitationFernández de la Mora and Attoui 2007). This insulating material is slightly conductive which helps to dissipate the electric charges built up on the insulator surface due to charged particle depositions.

3. DATA INVERSION

The results of a typical measurement, where the TSI nano- DMA was operated at a flow ratio of 1.5/15 lpm while classifying 1.70 nm TDDABr monomer, are shown in . Since monodisperse particles were sent into the test DMA, the concentration ratio (N 2/N 1) can be written as,

where ηpen(dp) is the penetration efficiency of particles of diameter dp through the entrance and exit regions of the DMA and Ω(V,Z(dp)) is the DMA transfer function. The transfer function is defined as the fraction of particles of electrical mobility Z(dp) that penetrate through a DMA classifying region operated at voltage V. As shown in , the transfer function is characterized by two parameters: the height (Ω(V,Z*)) and the normalized full-width at half-maximum (NFWHM), or the resolution (R). The height of the transfer function is the penetration fraction of particles with the centroid electrical mobility (Z*), i.e., the mobility that corresponds to the DMA voltage V through the voltage-mobility equation (CitationKnutson and Whitby 1975; CitationStolzenburg and McMurry 2008). The NFWHM of the transfer function is,
where ΔZFWHM is the transfer function width at half height (). The instrument mobility resolution is defined as the reciprocal of the NFWHM (CitationFlagan 1999),
and is a figure of merit for DMA performance.

To measure the true DMA transfer function (), ideally one will fix the DMA voltage and send monodisperse particles of different sizes to the DMA to obtain Ω(V,Z(dp )). Since it is impractical to generate many different monodisperse size standards under the narrow non-zero width of the transfer function, an alternative method is used—the particle size is fixed and the DMA voltage is scanned, as shown for example in . If the DMA transfer functions at different voltages are the same, the measured Ω at different voltages can be directly projected back to . Particle diffusion can broaden the DMA transfer function significantly when classifying sub-2 nm particles. The NFWHM of the DMA transfer function also becomes a function of the DMA voltage (the centroid mobility) due to diffusion. Therefore, the results obtained by measuring concentration ratios as a function of the test DMA voltage for monodisperse particles are not exactly equal to the transfer function, i.e., the diffusional broadening effect increases as the voltage is decreased in the course of the transfer function measurement. For all DMAs tested here, this effect was accounted for when inverting the data so that the transfer functions reported here correspond to the transfer function that would be measured at constant voltage if a sufficient range of calibration particles were available. The resolution of the Caltech nanoRDMA in the 1–2 nm size range was previously reported without considering this effect (CitationBrunelli et al. 2009). In that earlier study, the resolution was estimated based on ΔZ*FWHM and Z*peak in .

In this study, a data inversion scheme was developed based on Stolzenburg's diffusive transfer function. CitationStolzenburg (1988) extended the non-diffusive transfer function developed by CitationKnutson and Whitby (1975) to account for particle diffusion. A dimensionless diffusional broadening parameter, σtheo , was defined and an expression for the diffusive transfer function was developed. This theory is described in detail elsewhere (CitationStolzenburg 1988; CitationFlagan 1999; CitationStolzenburg and McMurry 2008) and is summarized in the Supplemental Information. In addition to particle diffusion, the transfer function might also be affected by instrument nonidealities, electrode imperfections and misalignments, and other factors that may influence the flow or electric fields (CitationStolzenburg 1988; CitationZhang and Flagan 1996; CitationChen and Pui 1997; CitationEichler et al. 1998; CitationKarlsson and Martinsson 2003; CitationHeim et al. 2005). To account for these potential effects, a multiplicative factor (fσ ) is used and the broadening parameter in the diffusive transfer function is rewritten as,

If no other broadening exists, the factor (fσ ) will be equal to unity. We assume that fσ is constant across the narrow non-zero width of the transfer function. Particle diffusion will increase the transfer function NFWHM and decrease the height, subsequently lowering the peak value of N 2/N 1. The losses at the entrance and exit regions of the DMA will affect N 2/N 1 without changing the NFWHM (CitationMartinsson et al. 2001). Therefore, two unknowns are needed to calculate the concentration ratio using EquationEquation (1): fσ and ηpen(dp). A least squares fitting method was used to solve for these two unknowns by matching the theoretically predicted concentration ratios with the measured N 2/N 1. For the data shown in , the theoretical calculated concentration ratios agreed with the experimental data very well when fσ is 1.07. The transfer function height (Ω(V,Z*)) and the penetration efficiency (ηpen ) were 33.8% and 23.1%, respectively. Their product is the peak value of N 2/N 1, 7.8% (). Another way to look at the additional broadening is to define it as σdistor which can be estimated from (CitationStolzenburg 1988; CitationRosell-Llompart et al. 1996; CitationFlagan 1999),

When comparing DMA performance, the migration Peclet number is often used (CitationZhang and Flagan 1996; CitationFissan et al. 1998; CitationFlagan 1999; CitationKarlsson and Martinsson 2003). It is defined as the ratio of the electrophoretic flux to the diffusive flux of particles during mobility classification. For each test DMA, the migration Peclet number allows the results obtained at different aerosol and sheath flowrates but the same aerosol-to-sheath ratio, β, to be conveniently presented on the same plot. In this study, we used the definition given by CitationFlagan (1999),

where q is the charge carried by particles (one elementary charge for all ion standards used in this study), V is the DMA voltage, k is Boltzmann's constant, T is the absolute temperature, and f is a DMA geometry factor that accounts for the nonuniformity of the DMA electric field. For radial DMAs, this geometry factor is unity. For cylindrical DMAs, this geometry factor is,
where R 1 and R 2 are the inner radius and the outer radius of the DMA, respectively. The f factors for all test DMAs are summarized in .

TABLE 2 Estimated DMA transfer function parameters and effective lengths

FIG. 3 Performances of four DMAs in classifying THABr monomer (1.47 nm): (a) measured concentration ratios by scanning the DMA voltage (varying centroid mobility); (b) the inverted transfer functions of four DMAs with a fixed DMA voltage for 1.47 nm THABr monomer. For the TSI nanoDMA, 15 lpm transport flow was used to reduce the losses through the entrance region (CitationChen et al. 1998).

FIG. 3 Performances of four DMAs in classifying THABr monomer (1.47 nm): (a) measured concentration ratios by scanning the DMA voltage (varying centroid mobility); (b) the inverted transfer functions of four DMAs with a fixed DMA voltage for 1.47 nm THABr monomer. For the TSI nanoDMA, 15 lpm transport flow was used to reduce the losses through the entrance region (CitationChen et al. 1998).

The losses at the entrance and exit regions of the test DMAs were simulated using the laminar diffusional deposition model for cylindrical tubes, and an effective length, Leff , for each test DMA is reported. Though these losses are probably more complicated (e.g., electrostatic deposition likely contributes to observed losses, and flow channels are not all cylindrical), it has been shown that the modified Gormley and Kennedy equation (CitationGormley and Kennedy 1949; CitationCheng 2002) provides a reasonable model for explaining size-dependent DMA penetration efficiencies (CitationReineking and Porstendorfer 1986; CitationKarlsson and Martinsson 2003; CitationIida et al. 2006). More accurate estimation of the diffusion losses in DMA entrance and exit regions might be obtained by dividing them into different sections that are each analyzed using the approach originally developed for the corresponding heat transfer problem (CitationGraetz 1885) and electrostatic deposition can also be included (CitationZhang and Flagan 1996; CitationKarlsson and Martinsson 2003). However, the effective length approach is simpler and provides results that are more convenient to be included in DMA data inversion. The dimensionless deposition parameter, μ, can be written as,

where D is the diffusion coefficient of particles with diameter dp and Q is the flow rate through the cylindrical tube. The penetration efficiency is given by (CitationCheng 2002),

An effective length can be derived by fitting the calculated penetration efficiencies using the above equations to the measured penetrations. A least squares method was used for the fitting.

The errors in determining the DMA transfer function (resolution/NFWHM and height) and penetration efficiency were estimated to be smaller than 7%. Uncertainties are due primarily to the uncertainties in DMA flowrates and particle concentration measurements. As shown in , the fitted concentration ratios mostly agree with the measured N 2/N 1 very well. Though it was not observed under the conditions shown in , an abnormal tail at higher voltage side (lower mobility) occurred under certain conditions for all test DMAs. This was also observed by CitationHeim et al. (2010) in a TDMA experiment using THABr monomer and trimer, but not for the dimer. Their results for the monomer disagree with the proposed fragmentation and clustering hypothesis (CitationRosell-Llompart et al. 1996; CitationFernández de la Mora et al. 2005). The origin of this abnormal tail remains to be identified. Their N 2/N 1 values are often much smaller than the peak value. The abnormal tail was not included in the inversion fitting which may cause some additional uncertainty in the reported resolution and transmission.

FIG. 4 TSI nanoDMA: (a) the resolution as a function of instrument Peclet number; (b) the NFWHM; (c) the height, and the penetration efficiency through DMA entrance and exit regions as a function of particle mobility diameter for the flow ratio of 1.5/15 lpm; (d) the measured penetration efficiency versus the fitted penetration efficiency.

FIG. 4 TSI nanoDMA: (a) the resolution as a function of instrument Peclet number; (b) the NFWHM; (c) the height, and the penetration efficiency through DMA entrance and exit regions as a function of particle mobility diameter for the flow ratio of 1.5/15 lpm; (d) the measured penetration efficiency versus the fitted penetration efficiency.

4. RESULTS AND DISCUSSION

Estimated DMA transfer function parameters and effective lengths are summarized in . The performance of the TSI nanoDMA, the Grimm nanoDMA, the Caltech nanoRDMA, and the Karlsruhe-Vienna DMA in classifying 1.47 nm THABr monomer are shown in . The measured transfer function of the Caltech RDMA was very broad and is therefore not included in . Different flowrates were used for these DMAs, but the aerosol-to-sheath flow ratios were similar (∼0.1). In order to present the results from different DMAs in one graph, the corresponding centroid mobilities (Z*) rather than the test DMA voltages are used in . The Caltech nanoRDMA has the highest peak N 2/N 1 value, followed by the TSI nanoDMA, the Grimm nanoDMA, and the Karlsruhe-Vienna DMA. Their inverted transfer functions, obtained by fixing voltage at a value corresponding to the centroid mobility diameter of 1.47 nm, are shown in . The resolution of the Caltech nanoRDMA was the best, followed by the Karlsruhe-Vienna DMA, the Grimm nanoDMA, and the TSI nanoDMA. Since the multiplicative factors (fσ ) for both the TSI nanoDMA and the Grimm nanoDMA were close to unity, there was no additional broadening for these two DMAs. Both the Caltech nanoRDMA and the Karlsruhe-Vienna DMA had additional broadening under these conditions. The resolution and peak N 2/N 1 value of the Grimm nanoDMA were very close to those of the Karlsruhe-Vienna DMA.

The flowrates of a DMA are critical in determining its performance and different DMAs are designed to operate at different flowrates. Instrument characterization should be based on the optimum operation of each DMA such that a direct comparison of the different DMA performances is only suitable to a limited degree (CitationFissan et al. 1996; CitationBirmili et al. 1997; CitationFlagan 1999). In the following discussion, we focused on a comparison between the measured performances and the theoretical predictions for each test DMA.

4.1. TSI nanoDMA

Both the resolution and the penetration efficiency of the TSI nanoDMA are shown in . Monodisperse ions with diameters of 1.16, 1.47, and 1.70 nm were tested using flow ratios of 1.5/15 and 2/20 lpm. In both cases, 15 lpm transport flow was used to reduce the losses through the nanoDMA entrance region (CitationChen et al. 1998). Literature data on the TSI nanoDMA performance in classifying particles larger than 3 nm are also included in (CitationChen et al. 1998; CitationLi et al. 2006). As shown in , the experimentally determined resolution for this instrument agreed very well with its ideal resolution (fσ = 1.0) in both diffusive and non-diffusive ranges. No additional broadening due, for example, to imperfections in flow profiles, electric fields, or machining tolerances, was observed under these conditions. At the test flowrates, the NFWHM of the TSI nanoDMA transfer function in classifying sub-2nm particles is 2∼4 times wider than the non-diffusive value (0.1).

shows the transfer function heights (Ω(V,Z*)) for classifying different sized particles at a flow ratio of 1.5/15 lpm and the penetration efficiencies (ηpen ) of sub-2 nm particles through the nanoDMA entrance and exit regions. Literature data for particles larger than 3 nm were also analyzed using the same method, i.e., by separating Ω(V,Z*) and ηpen . Together the penetration efficiencies for different sized particles were fitted with the modified Gormley and Kennedy equation. If the flowrate of 1.5 lpm was used (without considering the 15 lpm transport flow at the entrance region), the effective length of the TSI nanoDMA was found to be 1.58 m. As shown in , good agreement between the experimental data and the fitted penetrations is observed.

To include the transport flow, the entrance and exit regions can be fitted separately and two effective lengths will be reported. The flowrate through the first length (the entrance region) is then the aerosol flow plus the transport flow, while the flowrate through the second length (the exit region) is only the aerosol flow. Both results from this study and literature data were used for the fitting. There are total 45 data points and their information are summarized here—this study: 1.5 and 2 lpm (15 lpm transport flow), 1–2 nm; CitationChen et al. (1998): 1.5 lpm (with and without 15 lpm transport flow), 3–50 nm; CitationLi et al. (2006): 0.75 and 1.5 lpm (no transport flow), 6–50 nm; CitationHeim et al. (2007): 2 lpm (15 lpm transport flow), 3–10 nm. The effective lengths for the entrance and exit regions were found to be 2.24 and 1.40 m, respectively. As shown in , the fitted penetration efficiencies agree well with the measured ones.

FIG. 5 Caltech nanoRDMA: (a) the resolution as a function of instrument Peclet number; (b) the NFWHM; (c) the height, and the penetration efficiency as a function of particle mobility diameter for the flow ratio of 0.6/6 lpm.

FIG. 5 Caltech nanoRDMA: (a) the resolution as a function of instrument Peclet number; (b) the NFWHM; (c) the height, and the penetration efficiency as a function of particle mobility diameter for the flow ratio of 0.6/6 lpm.

4.2. Caltech nanoRDMA

The results for the Caltech nanoRDMA are shown in . Monomobile ions with mobility diameters of 1.16, 1.47, and 1.70 nm were tested using flow ratios of 0.6/6, 1/10, and 1.5/15 lpm. At the test flowrates, the NFWHM of the Caltech nanoRDMA transfer function in classifying sub-2 nm particles was equal to or less than 0.2 (relatively close to the non-diffusive value of 0.1). The measured resolution for this instrument agreed very well with its ideal resolution at the flow ratio of 0.6/6 lpm. However, the measured resolution fell below its ideal resolution at the flow ratio of 1/10 lpm and deviated further at higher flowrates (1.5/15 lpm). Its resolutions at the flow ratio of 0.6/10 lpm were also found to be lower than the ideal values, which is consistent to previous studies (CitationBrunelli et al. 2009). As shown in , σdistor increases with the increasing flowrate through the nanoRDMA. Additional tests indicated that turbulence occurred inside the nanoRDMA for a sheath flowrate of 18 lpm and particles were detected at the aerosol outlet when there was no voltage applied. Similar resolution degradation by flow effects was observed for the Caltech RDMA (CitationZhang and Flagan 1996) and was attributed to the flow disturbances near the entrance slot or at the classified aerosol exit. The shorter classification length of the nanoRDMA might cause additional non-uniformities in flow and electric field, e.g., in addition to the observed resolution degradation, as confirmed in this study, an empirical correction factor of 0.874 is needed for the conversion between the DMA voltage and the centroid mobility (CitationBrunelli et al. 2009). Both the transfer function heights (Ω(V,Z*)) for classifying different sized particles and the penetration efficiencies (ηpen ) of sub-2 nm particles at a flow ratio of 0.6/6 lpm are shown in . The effective length of the Caltech nanoRDMA was found to be 0.60 m. If ηpen for all the tested flowrates (0.6, 1.0, and 1.5 lpm) were fitted together, the effective length was found to be 0.87 m.

4.3. Caltech RDMA

The results for the Caltech RDMA are shown in . Monomobile ions with diameters of 1.47 and 1.70 nm were tested using flow ratios of 1.5/15 and 2/20 lpm. Literature data on the Caltech RDMA performance in classifying particles larger than 3 nm are also included in (CitationZhang and Flagan 1996; CitationBirmili et al. 1997). As shown in , the measured Caltech RDMA resolution was lower than its ideal resolution in both diffusive and non-diffusive size ranges. By using a constant additional broadening (σdistor = 0.05), theoretically calculated RDMA resolutions agreed very well with measured resolutions for particles larger than 3 nm, while σdistor was estimated to be ∼0.13 for sub-2 nm classification. Again, flow disturbances inside the DMA might account for the observed resolution degradation for the Caltech RDMA.

The transfer function heights (Ω(V,Z*)) at σdistor of 0.05 and 0.13 are shown in . The penetration efficiencies (ηpen ) for sub-2 nm particles and literature penetration efficiencies for large particles (CitationZhang and Flagan 1996) are also included in . For these literature data, ηpen and Ω(V,Z*) were already separated from each other. There was no single effective length to achieve good fitting with the penetration efficiencies of particles in the whole size range. When fitted separately, the effective length was found to be 1.77 m for sub-2 nm particles and less than 0.1 m for particles larger than 3 nm. If ηpen of sub-2 nm particles tested at all the flowrates (1.5 and 2 lpm) were fitted together, the effective length was found to be 1.99 m. Based on the same RDMA data set for large particles, an effective length of 0.083 m was reported by CitationKarlsson and Martinsson (2003). It remains to be determined why the resolution/ NFWHM and penetration of RDMA is different for small particles (<2 nm) and for large particles (≥3 nm). This difference may indicate that different mechanisms, with different size-dependence, are responsible for losses in the two regimes, possibly electrophoretic and diffusive deposition (CitationKousaka et al. 1986; CitationZhang and Flagan 1996), but additional work would be required to resolve the cause for the variable loss rates. It has been observed that the performance of Vienna DMA copies can be different from one instrument to the next due to machining inconsistencies (CitationKarlsson and Martinsson 2003). It remains to be tested whether this also contributes to the observed discrepancy for the RDMA.

FIG. 6 Caltech RDMA: (a) the resolution as a function of instrument Peclet number; (b) the NFWHM; (c) the height, and the penetration efficiency as a function of particle mobility diameter for the flow ratio of 1.5/15 lpm.

FIG. 6 Caltech RDMA: (a) the resolution as a function of instrument Peclet number; (b) the NFWHM; (c) the height, and the penetration efficiency as a function of particle mobility diameter for the flow ratio of 1.5/15 lpm.

4.4. Grimm nanoDMA

The results for the Grimm nanoDMA are shown in . THABr monomer and dimer with diameters of 1.47 and 1.78 nm were tested using a flow ratio of 2/21.9 lpm. The measured Grimm nanoDMA resolutions agreed very well with its ideal resolutions for sub-2 nm particles and no additional broadening was observed under these conditions. The penetration efficiencies (ηpen ) of the Grimm nanoDMA were previously tested using particles larger than 3 nm (CitationHeim et al. 2007). Together the penetration efficiencies for different sized particles were fitted with the modified Gormley and Kennedy equation. The effective length of the Grimm nanoDMA was found to be 3.87 m.

FIG. 7 Grimm nanoDMA: (a) the NFWHM, (b) the height, and the penetration efficiency as a function of particle mobility diameter for the flow ratio of 2/21.9 lpm.

FIG. 7 Grimm nanoDMA: (a) the NFWHM, (b) the height, and the penetration efficiency as a function of particle mobility diameter for the flow ratio of 2/21.9 lpm.

4.5. Karlsruhe-Vienna DMA

The results for the Karlsruhe-Vienna DMA are shown in . THABr monomer, dimer, and trimer with diameters of 1.47, 1.78, and 1.97 nm were tested using a flow ratio of 6/61.4 lpm. The measured resolution for this DMA agreed very well with its ideal resolution for THABr dimer and trimer. For the 1.47 nm monomer, the measured transfer function was wider than the ideal transfer function and the multiplicative factor (fσ ) was 1.22. This additional broadening is believed to be related to the abnormal tail that occurred at lower mobility side and was more severe for the Karlsruhe-Vienna DMA than other four test DMAs. By fitting the abnormal tail as a secondary peak and then excluding its contribution to the primary peak, fσ for the THABr monomer was found to be 1.12, which is close to the values for the dimer and trimer, i.e., 1.07. shows both Ω(V,Z*) and ηpen at the flow ratio of 6/61.4 lpm. The penetration efficiencies of particles larger than 3 nm are also included (CitationHeim et al. 2007). The effective length of the Karlsruhe-Vienna DMA was found to be 11.53 m. If the effect of the abnormal tail on ηpen was included, the effective length was found to be 11.04 m.

FIG. 8 Karlsruhe-Vienna DMA: (a) the NFWHM, (b) the height, and the penetration efficiency as a function of particle mobility diameter for the flow ratio of 6/61.4 lpm.

FIG. 8 Karlsruhe-Vienna DMA: (a) the NFWHM, (b) the height, and the penetration efficiency as a function of particle mobility diameter for the flow ratio of 6/61.4 lpm.

The Karlsruhe-Vienna DMA replaces the tangential flow arrangement in the original Vienna DMA (CitationWinklmayr et al. 1991) with axisymmetric sheath inlet and excess outlet. The sheath flow is introduced into the center of a plenum chamber through a cylindrical tube. The flow is distributed radially via 40 holes around the circumference before entering the classification zone through a laminarization screen. To ensure flow uniformity, the excess outflow passes through 30 holes around the insulator circumference followed by four excess outlet ports. The tangential aerosol inlet used in Vienna type DMAs can cause asymmetric flow, and may result in DMA transfer function broadening (CitationHeim et al. 2005). The DMA inlet blade was redesigned to increase the pressure drop through the annular gap in order to achieve a homogenous aerosol flow into the classifying region. The flow uniformity was confirmed even when the aerosol flowrate is as low as 0.3 lpm. In the aerosol outlet, a “leaky” insulator is used to prevent a penetration decrease over time due to the accumulation of charged particles on the insulator. The DMA classification length is 2.5 cm which can be easily extended to 6.5 cm by changing the inner electrode. Unlike previously tested Vienna DMA versions whose resolutions were found to be lower than ideal values (CitationFissan et al. 1996; CitationBirmili et al. 1997), the resolution of the Karlsruhe-Vienna DMA is very close to theoretical values.

5. CONCLUSIONS

Transfer functions and penetrations for sub-2 nm particles were evaluated for five DMAs. The Caltech nanoRDMA shows the best resolution in this size range, followed by the Karlsruhe-Vienna DMA, the Grimm nanoDMA, the TSI nanoDMA, and the Caltech RDMA. The penetration efficiencies of each test DMA were fitted using the modified Gormley and Kennedy equation to determine an effective length. The effective lengths for the Caltech nanoRDMA, the Caltech RDMA, the TSI nano- DMA, the Grimm nanoDMA, and the Karlsruhe-Vienna DMA are 0.87, 1.99, 3.64, 3.87, and 11.53 m, respectively. The Caltech nanoRDMA shows the highest penetration efficiency. This study also demonstrates that Stolzenburg's diffusive transfer function accurately predicts transfer functions for particles down to the smallest tested size (1.16 nm). The resolutions of the TSI nanoDMA, the Grimm nanoDMA, the Caltech nanoRDMA (at low flow rates), and the Karlsruhe-Vienna DMA (for 1.78 and 1.97 nm particles) agree very well with the diffusive transfer function. Additional broadening was observed for the Caltech RDMA, the Caltech nanoRDMA (at high flow rates), and the Karlsruhe-Vienna DMA for 1.47 nm particles. Results are presented in a form that is useful for DMA data inversion.

Supplemental material

uast_a_546819_sup_17050484.zip

Download Zip (40.3 KB)

Acknowledgments

This work was partially supported by a grant from the US NSF (ATM-0506674) and US DOE (DE-FG-02-05ER63997). JJ thanks Dr. Kenjiro Iida for his help on the data inversion program. A NSF graduate fellowship to NAB is gratefully acknowledged. We thank the two anonymous reviewers for helping to improve the clarity of our article.

REFERENCES

  • Birmili , W. , Stratmann , F. , Wiedensohler , A. , Covert , D. , Russell , L. M. and Berg , O. 1997 . Determination of Differential Mobility Analyzer Transfer Functions Using Identical Instruments in Series . Aerosol Sci. Technol. , 27 : 215 – 223 .
  • Brunelli , N. A. , Flagan , R. C. and Giapis , K. P. 2009 . Radial Differential Mobility Analyzer for One Nanometer Particle Classification . Aerosol Sci. Technol. , 43 : 53 – 59 .
  • Camata , R. P. , Atwater , H. A. and Flagan , R. C. 2001 . Space-Charge Effects in Nanoparticle Processing Using the Differential Mobility Analyzer . J. Aerosol Sci. , 32 : 583 – 599 .
  • Chen , D. R. and Pui , D. Y. H. 1997 . Numerical Modeling of the Performance of Differential Mobility Analyzers for Nanometer Aerosol Measurements . J. Aerosol Sci. , 28 : 985 – 1004 .
  • Chen , D. R. , Pui , D. Y. H. , Hummes , D. , Fissan , H. , Quant , F. R. and Sem , G. J. 1998 . Design and Evaluation of a Nanometer Aerosol Differential Mobility Analyzer (Nano-DMA) . J. Aerosol Sci. , 29 : 497 – 509 .
  • Cheng , Y. S. 2002 . “ Condensation Detection and Diffusion Size Separation Techniques ” . In Aerosol Measurement: Principles, Techniques, and Applications , Edited by: Baron , P. A. and Willeke , K. New York : John Wiley & Sons .
  • Eichler , T. , de Juan , L. and Fernández de la Mora , J. 1998 . Improvement of the Resolution of TSI's 3071 DMA via Redesigned Sheath Air and Aerosol Inlets . Aerosol Sci. Technol. , 29 : 39 – 49 .
  • Fernández de la Mora , J. 2011 . “ Sub-3 nm Aerosol Measurement with DMAs and CNCs ” . In Aerosol Measurement: Principles, Techniques, and Applications , Edited by: Kulkarni , P. , Baron , P. A. and Willeke , K. New York : John Wiley & Sons .
  • Fernández de la Mora , J. and Attoui , M. A DMA Covering the 1–100 nm Particle Size Range with High Resolution Down to 1 nm . European Aerosol Conference , Salzburg, , Austria
  • Fernández de la Mora , J. , de Juan , L. , Eichler , T. and Rosell , J. 1998 . Differential Mobility Analysis of Molecular Ions and Nanometer Particles . Trac-Trends in Anal. Chem. , 17 : 328 – 339 .
  • Fernández de la Mora , J. , Thomson , B. A. and Gamero-Castano , M. 2005 . Tandem Mobility Mass Spectrometry Study of Electrosprayed Tetraheptyl Ammonium Bromide Clusters . J. Am. Soc. Mass Spectrom. , 16 : 717 – 732 .
  • Fissan , H. , Hummes , D. , Stratmann , F. , Buscher , P. , Neumann , S. , Pui , D. Y. H. and Chen , D. 1996 . Experimental Comparison of Four Differential Mobility Analyzers for Nanometer Aerosol Measurements . Aerosol Sci. Technol. , 24 : 1 – 13 .
  • Fissan , H. , Pocher , A. , Neumann , S. , Boulaud , D. and Pourprix , M. 1998 . Analytical and Empirical Transfer Functions of a Simplified Spectrometre de Mobilite Electrique Circulaire (SMEC) for Nano Particles . J Aerosol Sci. , 29 : 289 – 293 .
  • Flagan , R. C. 1998 . History of Electrical Aerosol Measurements . Aerosol Sci. Technol. , 28 : 301 – 380 .
  • Flagan , R. C. 1999 . On Differential Mobility Analyzer Resolution . Aerosol Sci. Technol. , 30 : 556 – 570 .
  • Gormley , P. G. and Kennedy , M. K. 1949 . Diffusion from a Stream Flowing through a Cylindrical Tube . Proc. R. Ir. Acad. , A52 : 163 – 169 .
  • Graetz , L. 1885 . On the Heat Capacity of Fluids . Annalen der Physik. , 25 : 337 – 357 .
  • Heim , M. , Attoui , M. and Kasper , G. Performance and Characterization of a High Flow, High-Resolution DMA: Particle Losses and Filtration Efficiency Measurments for Particles Below 20 nm . European Aerosol Conference , Salzburg, , Austria
  • Heim , M. , Attoui , M. and Kasper , G. 2010 . The Efficiency of Diffusional Particle Collection onto Wire Grids in the Mobility Equivalent Size Range of 1.2–8 nm . J. Aerosol Sci. , 41 : 207 – 222 .
  • Heim , M. , Wild , M. and Kasper , G. Asymmetric Flow Effects in Vienna-Type Differential Mobility Analyzers . European Aerosol Conference , Ghent, , Belgium
  • Herrmann , W. , Eichler , T. , Bernardo , N. and Fernández de la Mora , J. Turbulent Transition Arises at Reynolds Number 35,000 in a Short Vienna Type DMA with a Large Laminarization Inlet . AAAR Annual Conference , St. Louis, , USA
  • Hewitt , G. 1957 . The Charging of Small Particles for Electrostatic Precipitation . AIEE Trans. , 76 : 300 – 307 .
  • Iida , K. , Stolzenburg , M. , McMurry , P. , Dunn , M. J. , Smith , J. N. , Eisele , F. and Keady , P. 2006 . Contribution of Ion-Induced Nucleation to New Particle Formation: Methodology and Its Application to Atmospheric Observations in Boulder, Colorado . J. Geophys. Res. Atmospheres. , 111 : D23201
  • Iida , K. , Stolzenburg , M. R. and McMurry , P. H. 2009 . Effect of Working Fluid on Sub-2 nm Particle Detection with a Laminar Flow Ultrafine Condensation Particle Counter . Aerosol Sci. Technol. , 43 : 81 – 96 .
  • Jokinen , V. and Makela , J. M. 1997 . Closed-Loop Arrangement with Critical Orifice for DMA Sheath Excess Flow System . J Aerosol Sci. , 28 : 643 – 648 .
  • Karlsson , M. N. A. and Martinsson , B. G. 2003 . Methods to Measure and Predict the Transfer Function Size Dependence of Individual DMAs . J. Aerosol Sci. , 34 : 603 – 625 .
  • Knutson , E. O. and Whitby , K. T. 1975 . Aerosol Classification by Electric Mobility: Apparatus, Theory, and Applications . J. Aerosol Sci. , 6 : 443 – 451 .
  • Kousaka , Y. , Okuyama , K. , Adachi , M. and Mimura , T. 1986 . Effect of Brownian Diffusion on Electrical Classification of Ultrafine Aerosol-Particles in Differential Mobility Analyzer . J. Chem. Eng Japan , 19 : 401 – 407 .
  • Ku , B. K. and Fernández de la Mora , J. 2009 . Relation between Electrical Mobility, Mass, and Size for Nanodrops 1–6.5 nm in Diameter in Air . Aerosol Sci. Technol. , 43 : 241 – 249 .
  • Larriba , C. , Hogan , C. J. Jr , Attoui , M. , Borrajo , R. , Garcia , J. F. and Fernández de la Mora , J. 2011 . The Mobility-Volume Relationship below 3.0 nm Examined by Tandem Mobility-Mass Measurement . Aerosol Sci. Technol. , 45 : 453 – 467 .
  • Li , W. L. , Li , L. and Chen , D. R. 2006 . Technical Note: A New Deconvolution Scheme for the Retrieval of True DMA Transfer Function from Tandem DMA data . Aerosol Sci. Technol. , 40 : 1052 – 1057 .
  • Liu , B. Y. H. and Pui , D. Y. H. 1974 . A Submicron Aerosol Standard and the Primary, Absolute Calibration of the Condensation Nuclei Counter . J. Colloid Interface Sci. , 47 : 155 – 171 .
  • Martinsson , B. G. , Karlsson , M. N. A. and Frank , G. 2001 . Methodology to Estimate the Transfer Function of Individual Differential Mobility Analyzers . Aerosol Sci. Technol. , 35 : 815 – 823 .
  • Mesbah , B. 1994 . “ Le Spectromètre de Mobilité Electrique Circulaire S.M.E.C ” . In Théorie, Performances et Applications—Contamination Contrôlée des Surfaces Granulomètrie de l’aérosol fin , Paris : Université Paris XII .
  • Ooghe , M. , Mesbah , B. , Pourprix , M. and Boulaud , D. Radial Flow Differential Mobility Analyzer (RF-DMA) . International Aerosol Conference , pp. 463 Los Angeles, , USA
  • Park , K. , Dutcher , D. , Emery , M. , Pagels , J. , Sakurai , H. , Scheckman , J. , Qian , S. , Stolzenburg , M. R. , Wang , X. , Yang , J. and McMurry , P. H. 2008 . Tandem Measurements of Aerosol Properties—A Review of Mobility Techniques with Extensions . Aerosol Sci. Technol. , 42 : 801 – 816 .
  • Reineking , A. and Porstendorfer , J. 1986 . Measurements of Particle Loss Functions in a Differential Mobility Analyzer (Tsi, Model 3071) for Different Flow-Rates . Aerosol Sci. Technol. , 5 : 483 – 486 .
  • Rogak , S. N. , Flagan , R. C. and Nguyen , H. V. 1993 . The Mobility and Structure of Aerosol Agglomerates . Aerosol Sci. Technol. , 18 : 25 – 47 .
  • Rosell-Llompart , J. , Loscertales , I. G. , Bingham , D. and Fernández de la Mora , J. 1996 . Sizing Nanoparticles and Ions with a Short Differential Mobility Analyzer . J. Aerosol Sci. , 27 : 695 – 719 .
  • Rosser , S. and Fernández de la Mora , J. 2005 . Vienna-Type DMA of High Resolution and High Flow Rate . Aerosol Sci. Technol. , 39 : 1191 – 1200 .
  • Steiner , G. , Attoui , M. , Wimmer , D. and Reischl , G. P. 2010 . A Medium Flow, High-Resolution Vienna DMA Running in Recirculating Mode . Aerosol Sci. Technol. , 44 : 308 – 315 .
  • Stolzenburg , M. 1988 . An Ultrafine Aerosol Size Distribution Measuring System , Minneapolis : University of Minnesota Twin Cities .
  • Stolzenburg , M. and McMurry , P. 1988 . TDMAFIT User's Manual , Minneapolis : University of Minnesota .
  • Stolzenburg , M. R. and McMurry , P. H. 2008 . Equations Governing Single and Tandem DMA Configurations and a New Lognormal Approximation to the Transfer Function . Aerosol Sci. Technol. , 42 : 421 – 432 .
  • Stratmann , F. , Kauffeldt , T. , Hummes , D. and Fissan , H. 1997 . Differential Electrical Mobility Analysis: A Theoretical Study . Aerosol Sci. Technol. , 26 : 368 – 383 .
  • Tammet , H. 1995 . Size and Mobility of Nanometer Particles, Clusters, and Ions . J. Aerosol Sci. , 26 : 459 – 475 .
  • Tanaka , H. and Takeuchi , K. 2003 . Experimental Transfer Function for a Low-Pressure Differential Mobility Analyzer by Use of a Monodisperse C-60 Monomer . J. Aerosol Sci. , 34 : 1167 – 1173 .
  • Ude , S. and Fernández de la Mora , J. 2005 . Molecular Monodisperse Mobility and Mass Standards from Electrosprays of Tetra-Alkyl Ammonium Halides . J. Aerosol Sci. , 36 : 1224 – 1237 .
  • Winklmayr , W. , Reischl , G. P. , Lindner , A. O. and Berner , A. 1991 . A New Electromobility Spectrometer for the Measurement of Aerosol Size Distributions in the Size Range from 1 to 1000 Nm . J. Aerosol Sci. , 22 : 289 – 296 .
  • Zhang , S. H. , Akutsu , Y. , Russell , L. , Flagan , R. C. and Seinfeld , J. H. Radial Differential Mobility Analyzer . International Aerosol Conference , pp. 465 Los Angeles, , USA
  • Zhang , S. H. , Akutsu , Y. , Russell , L. M. , Flagan , R. C. and Seinfeld , J. H. 1995 . Radial Differential Mobility Analyzer . Aerosol Sci. Technol. , 23 : 357 – 372 .
  • Zhang , S. H. and Flagan , R. C. 1996 . Resolution of the Radial Differential Mobility Analyzer for Ultrafine Particles . J. Aerosol Sci. , 27 : 1179 – 1200 .
  • Zhao , J. , Eisele , F. L. , Titcombe , M. , Kuang , C. and McMurry , P. H. 2010 . Chemical Ionization Mass Spectrometric Measurements of Atmospheric Neutral Clusters Using the Cluster-CIMS . J. Geophys. Res. Atmospheres. , 115 : D08205

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.