1,259
Views
10
CrossRef citations to date
0
Altmetric
Regular Articles

Smoke Characterization and Feasibility of the Moment Method for Spacecraft Fire Detection

, , , , , , & show all
Pages 299-309 | Received 11 Sep 2014, Accepted 17 Feb 2015, Published online: 08 Apr 2015

Abstract

The Smoke Aerosol Measurement Experiment (SAME) has been conducted twice by the National Aeronautics and Space Administration and provided real-time aerosol data in a spacecraft micro-gravity environment. Flight experiment results have been recently analyzed with respect to comparable ground-based experiments. The ground tests included an electrical mobility analyzer as a reference instrument for measuring particle size distributions of the smoke produced from overheating five common spacecraft materials. Repeatable sample surface temperatures were obtained with the SAME ground-based hardware, and measurements were taken with the aerosol instruments returned from the International Space Station comprising two commercial smoke detectors, three aerosol instruments, which measure moments of the particle size distribution, and a thermal precipitator for collecting smoke particles for transmission electron microscopy (TEM). Moment averages from the particle number concentration (zeroth moment), the diameter concentration (first moment), and the mass concentration (third moment) allowed calculation of the count mean diameter and the diameter of average mass of smoke particles. Additional size distribution information, including geometric mean diameter and geometric standard deviations, can be calculated if the particle size distribution is assumed to be lognormal. Both unaged and aged smoke particle size distributions from ground experiments were analyzed to determine the validity of the lognormal assumption. Comparisons are made between flight experiment particle size distribution statistics generated by moment calculations and microscopy particle size distributions (using projected area equivalent diameter) from TEM grids, which have been returned to the Earth.

Copyright 2015 American Association for Aerosol Research

INTRODUCTION

SAME Experiment

Appropriate design of fire detection systems requires knowledge of both the expected signature of the events to be detected and the background levels. Ambient aerosols in spacecraft include significantly larger particles than on the Earth, as gravitational settling is absent; consequently, smoke detectors must optimally distinguish between background aerosols and smoke in order to prevent false alarms. Terrestrial fire detection systems have been developed based on extensive study of terrestrial fires (Bukowski and Mulholland Citation1978; Bukowski et al. Citation2003). Unfortunately, there is no corresponding dataset for spacecraft fires, and consequently the fire detectors in current spacecraft were developed based upon terrestrial designs. There are a number of factors that could be expected to affect the particle size distribution of the smoke from spacecraft fires. In low gravity, buoyant flow is negligible, which increases particle residence time in microgravity fires and increases the transport time from the reaction zone to detectors (Brooker et al. Citation2007). Microgravity fires can have significantly different structure from their 1-g counterparts, which can change the formation history of smoke particles. Finally, the materials used in spacecraft are different from typical terrestrial environments where smoke properties were previously evaluated. All of these effects can influence the smoke particle size distribution. The objective of Smoke Aerosol Measurement Experiment (SAME) was to make sufficient measurements of smoke in space to enable improved design of future fire detectors.

It is critically important to detect a fire in its early phase before a flame is established, given the constrained volume on any spacecraft. Consequently, the primary target for spacecraft fire detection is pyrolysis products rather than soot. Therefore, SAME was designed to characterize smoke from overheating samples (oxidative pyrolysis) rather than from flaming combustion. Detectors used on the Space Shuttle were based upon ionization fire detector technology, the most advanced technology available at the time and used an inertial separator designed to eliminate particles larger than 1–2 μm. The International Space Station (ISS) smoke detectors use near-infrared (IR) forward scattering, rendering them most sensitive to particles larger than 1 μm outside the range of sensitivity of the shuttle detector.

The SAME was developed to obtain smoke particle size distribution parameters on orbit without returning samples to the Earth. This is a challenging endeavor because existing aerosol instruments are typically large and incompatible with spacecraft experiment constraints. Space experiments cannot require extensive crew training, equipment calibration, or maintenance and instruments must have low power requirements, be compact, lightweight, and easily assembled and disassembled. The approach for SAME was to use three commercial off-the-shelf instruments to measure different moments of the smoke particle size distribution. Using these moments, different moment average diameters can be calculated (some of which require assumption of a lognormal distribution) and the smoke aerosol can be characterized for the benefit of future smoke detector design. The measurements were made on smoke generated by overheating materials commonly found on spacecraft with controlled sample temperatures, flow rates, and particle aging times. Materials tested include Teflon®,Footnote Kapton®, cotton lamp wick (cellulose, representative of paper, wood, and fabric), silicone rubber, and Pyrell®, a polyurethane foam. The experiment was designed to measure fresh and aged pyrolysis smoke because the likely origin of a spacecraft fire would be electronics in an avionics enclosure or other poorly ventilated region. In such a scenario, the smoke concentration would increase in the confined space before escaping into the cabin where large-scale forced turbulence would slowly dilute the smoke. Thus, the properties of early and aged smoke should be known for optimal fire detector design. The experiment was performed in space in 2007 and 2010 on the ISS. The purpose of this article is two-fold: (1) Report the pyrolysis smoke characteristics of common spacecraft materials to inform future fire detector design, and (2) evaluate the feasibility and limitations of using combined moments for measuring smoke aerosol size distribution parameters in low gravity, particularly the validity of the lognormal assumption. A companion article (CitationMulholland et al. 2015) discusses other aspects of the SAME smoke, such as pyrolysis rate, smoke plume structure, yield, and particle structure.

Moment Method

The approach used by the SAME experiment is termed the “moment method” for convenience (Cleary et al. Citation2003). Three moments of the smoke particle size distribution (zeroth, first, and third moments) were measured, and using the properties of the lognormal distribution, the geometric mean diameter and the standard deviation of the aerosol were calculated. The following two assumptions are made when using this method: The aerosol particles have a spherical (or nearly spherical) shape, and the size distribution is lognormal.

We provide a detailed overview of the moment method in the online supplemental information (SI). Relevant formulas used for data in this study are repeated here without elaboration. The zeroth moment (M0) is equal to the total number concentration, Ntot. When particles can be characterized as spherical, the first moment (M1) is equal to the total diameter length concentration, or the integrated diameter per unit volume, Ltot, and the third moment (M3) is proportional to the total volume and/or mass concentration (Mtot = πρM3/6), which includes the particle density.

Thus, one can obtain the commonly used count mean diameter (simple average), dav, and the diameter of average mass, dm: [1] [2]

There is no assumption about the form of the size distribution for Equations (1) and (2). However, to determine the geometric standard deviation, σg, of the size distribution, or other moment diameters by the moment method, requires that the size distribution be lognormal and that the particles be spherical. The lognormal distribution is widely used for describing aerosols, including non-flaming smoke because for most smoke aerosols, the bulk of the number concentration is associated with smaller particles (Raabe Citation1971; Reist Citation1984). Many studies have assumed that pyrolysis and combustion smoke aerosols from various fuels have a lognormal size distribution (Chen et al. Citation1990; Li and Hopke Citation1993; Zai et al. Citation2006; Xie et al. Citation2007; Chakrabarty et al. Citation2010; Janhäll et al. Citation2010; Mack et al. Citation2010).

The general formula for the pth moment average of the qth moment distribution is [3] where σg is the same for particle number and volume distributions, and dg is equal to the count median diameter of the distribution and is the same as the geometric mean diameter if the lognormal assumption is satisfied.

If a histogram of particle sizes is available, the diameter of an average property proportional to (dp)p can be calculated for i bins with the following formula: [4]

With this formula, binned data from a reference instrument can be used to verify results from moment method calculations.

Aerosol instruments are limited in their range of measurements, and the accuracy of the measurements may vary over the range as well. If the moments of the particle size distribution are determined by instruments that are not identical in their ranges of particle size measurement, we can quantify the truncated moment value normalized by the total moment value. This relative value indicates how much of an actual signal is captured in the limited detection range of an instrument. The formula for a bounded moment measurement, which assesses the uncertainty induced by an instrument omitting particles above or below a certain diameter D, is based on the pth moment cumulative function of a lognormal distribution with dg and σg. If D is the particle size below which no signal can be detected, the relative cumulative pth moment is, [5]

The derivation of this equation is given in the online Supplemental Information. Note that the limiting diameter D is normalized by dg. This relative cumulative moment function, Mp,rel gives the percentage of the pth moment instrument signal that is captured when particles smaller than a diameter D cannot be detected, assuming a lognormal distribution with dg and σg. Conversely, when Mp,rel is subtracted from 1, it gives the percentage of the instrument signal that is lost due to lack of instrument range beyond diameter D.

SAME Aerosol Instruments

The SAME flight and ground test experiment measurements were made using three commercial instruments, which had been ruggedized and re-packaged for space flight. Two are industrial hygiene instruments and one is a residential smoke detector. These instruments were chosen because of their simplicity, low power needs, and small size. Unfortunately, they all show material or size-dependent behavior. Detailed empirical calibrations were performed with these instruments, which are described in the SI along with additional information on the instruments.

The zeroth moment instrument is a P-Trak™ (TSI, Shoreview, MN, USA), which is a condensation particle counter that was modified for use in space because the isopropanol condensate does not flow downwards to the wick in low gravity (Urban et al. 2005). To mitigate this issue, very small grooves were added to the walls of the condensing section of the device to improve conductance of condensate back to the wick. The first moment instrument is the ionization chamber from a residential smoke detector. This device uses an alpha-particle emitter to generate ions in a region within a DC electric field. Drifting ions in the electric field results in a current, and the presence of aerosol particles reduces the current as a result of the attachment of ions to particles. The mobility of the charged aerosol is too small for it to be collected on the ionization chamber electrode. Required minimum particle concentrations are of the order of 105 particles/cm3 and no sample dilution is required. The SAME third moment instrument is the DustTrak™ (TSI, Shoreview, MN, USA), which is a nephelometer using a 90° light scattering signal with a wide acceptance angle and output calibrated to quantify the aerosol mass concentration of Arizona Test Dust (ISO 12103-1). Material-specific calibrations and corrections were needed to account for the range of particle sizes, shapes, and refractive indices in the SAME experiment, and the particle density was required to compute M3 from the mass concentration. While some studies have shown that the DustTrak™ response is not proportional to mass (Moosmüller et al. Citation2001; Maricq Citation2013), after applying calibration factors, the DustTrak™ was found to correlate with the mass concentration. The calibration factors with uncertainty are given in the companion article by CitationMulholland et al. (2015). Some smoke could be sent to an autonomously operated thermal precipitator in which smoke particles are deposited on transmission electron microscope (TEM) grids. The SAME software command caused a valve to open, diverting smoke into one of the 12 isolated ducts containing a heated Kanthal wire above the TEM grid. Additional details on the thermal precipitator design are in the online Supplemental Information. After the space flight experiments, six thermal precipitators were returned to the Earth, and the grids were examined in a TEM to observe particle morphology and to obtain particle size distributions by microscopy. Characterization of particle morphology is a key to determining whether the moment method is valid for obtaining dg and σg from dav, and dm.

A schematic of the SAME hardware appears in . Space experiments are ideally autonomous, with minimal astronaut intervention beyond initial assembly. Hardware with programmable experiment parameters decreases crew-training requirements and increases the quantity and reliability of the resulting data. Software controlled all aspects of the experiment once the crew inserted the fuel sample carousel and commenced the test sequence. For the space experiments, smoke was generated by overheating a small sample of material in the smoke generation duct for approximately 60 seconds. During this interval, a rising piston drew smoke into a 6-liter aging chamber, where it could be held for a predetermined aging duration, allowing the particles to coagulate. Half of the smoke was pushed by the piston into moment instruments almost immediately for unaged smoke measurements by moment instruments. After a period of aging, the remaining smoke was measured. Additional information on the sample heating sequence and temperatures is given in the companion article (CitationMulholland et al. 2015).

FIG. 1. SAME flight hardware schematic, shown with additional ground testing apparatus within the dotted line. During ground tests, some smoke is diverted from the SAME setup to fill one of the two drums, which hold the diluted smoke for SMPS measurements. Two drums were needed to contain and measure both unaged and aged smoke.

FIG. 1. SAME flight hardware schematic, shown with additional ground testing apparatus within the dotted line. During ground tests, some smoke is diverted from the SAME setup to fill one of the two drums, which hold the diluted smoke for SMPS measurements. Two drums were needed to contain and measure both unaged and aged smoke.

SAME Smoke-in-Drums Ground-Based Experiment

In order to assess whether the size distribution of a particular smoke is lognormal, detailed particle size distributions were measured with a reference instrument. This cannot be accomplished in low gravity, so this investigation was performed with the ground-based engineering SAME hardware, which is identical to the setup on the ISS, incorporating the flight aerosol instruments that had been returned to the Earth. A Scanning Mobility Particle Sizer (SMPS) Spectrometer (3936, TSI, Shoreview, MN, USA) was used as the reference instrument in the validation experiment. The SMPS requires a two-minute scan through a range of voltages to acquire a high-resolution particle size distribution; however, the duration of smoke supplied from the SAME aging chamber is at most 30 seconds. Therefore, the smoke was collected in an intermediate container, which served two purposes. The first purpose was to sufficiently dilute the smoke from the SAME chamber to effectively stop coagulation (aging) of smoke particles during the SMPS scans. The second purpose was to have a large enough quantity of diluted smoke for multiple SMPS scans. A 55-gallon drum was chosen for this purpose and the SAME smoke-in-drums setup was developed to enable SMPS measurements on a portion of smoke output from the SAME piston chamber. The configuration is shown in , which shows the original SAME hardware outside the dashed outline. One DustTrak™ was removed from the original SAME configuration and its portion of the smoke sample was diverted from the setup to one of the two drums which hold the diluted smoke during multiple two-minute SMPS measurements. One drum collected fresh smoke from the heated sample material and the other was filled after a controlled aging period in the piston chamber. Thus, both aged and unaged smoke could be measured with SMPS. Unfortunately, one or more of the flight moment instruments was not functioning properly during these ground-based tests, so a majority of the resulting moment data were not reliable. Therefore, the analysis of the drum test data is exclusively on SMPS results, particularly to assess whether smoke from different test materials can be assumed to have a lognormal particle size distribution. While a comparison of the moment data with the SMPS reference data would have been preferable, lognormality is a fundamental assumption of that approach, and needs to be confirmed or refuted before spacecraft fire detection systems are further developed.

RESULTS

The results of the various aerosol measurements are presented in the following order: TEM images of low-gravity particle morphology which influences interpretation of the results of the moment instruments. SMPS smoke particle size distributions from ground testing provide a visual means of determining whether the smoke particle sizes are lognormally distributed, and discrete SMPS particle bin data are used to further test for lognormality. TEM particle size distributions from flight tests are given for materials with sufficient particle deposition in the thermal precipitator and provide an independent reference for moment method results.

TEM Particle Morphology Results from Flight Tests

Summaries are given on the morphology of all five materials tested, and details on the pyrolysis and formation are in the companion article (CitationMulholland et al. 2015). Kapton particles are the smallest of the five materials tested and are rarely agglomerated. The spherical shape and uniform density indicate growth by condensation in the saturated vapor of pyrolysis products. shows the effect of aging, with the unaged particles () having a higher population of very small particles, and the aged ones () appearing only slightly larger.

FIG. 2. TEM images showing morphology of smoke particles from ISS testing, all with a reference length scale = 5 μm, with the exception of image (f). (a) Unaged Kapton 574°C, (b) aged Kapton 574°C, (c) Unaged lamp wick, 265°C, (d) Unaged Pyrell, 242°C, (e) Unaged Teflon, 514°C, and (f) ISS residual unaged silicone smoke particles, 380°C, and reference length scale = 2 μm.

FIG. 2. TEM images showing morphology of smoke particles from ISS testing, all with a reference length scale = 5 μm, with the exception of image (f). (a) Unaged Kapton 574°C, (b) aged Kapton 574°C, (c) Unaged lamp wick, 265°C, (d) Unaged Pyrell, 242°C, (e) Unaged Teflon, 514°C, and (f) ISS residual unaged silicone smoke particles, 380°C, and reference length scale = 2 μm.

Lamp wick smoke aerosols () are known to be primarily spherical droplet-type particles that grow by condensation of pyrolysis gases (Mulholland Citation1995). Occasional doublets are seen but most are unagglomerated. Two distinct large particle types are observed: uniformly dense or lighter in the center, which suggests that they arrive at the carbon film of the TEM grid as a liquid. Some TEM images display additional faint particles that covered only one or two pixels.

Pyrell smoke particles () comprise agglomerates made up of primary particles ranging from 30 to 100 nm. Teflon primary particles are much smaller and are agglomerates with a fractal structure (). The darker agglomerates are more electron-dense and indicate that the fainter particles may have partially evaporated in the electron beam. In addition, some particles were not completely adhered to the TEM grid and movement could be observed as the force of the electron beam influenced the loose ends of agglomerates.

Silicone particles were not wholly preserved on the TEM grids owing to the volatile nature of pyrolysis products. Only very small and faint particles remained after the return flight to the Earth, as seen in . Note that the magnification in this figure is nearly double that of the other particle images shown. SMPS ground test data indicate that fresh silicone particles are much larger, as seen in .

FIG. 3. Ground testing size distributions and residual plots for Kapton (a, b), lamp wick (c, d), and silicone (e, f). Upper images are SMPS measurements, open markers for unaged smoke, and solid for aged; solid curves represent the nonlinear least square fits. Size distribution parameters: (a) Kapton® baseline temperature test (510°C), unaged dg = 139 nm, σg = 1.78, aged dg = 209 nm, σg = 1.66. (c) Lamp wick high-temperature test (286°C), unaged dg = 171 nm, σg = 1.98, aged dg = 248 nm, σg = 1.75. (e) Silicone baseline temperature test (342°C), unaged dg = 257 nm, σg = 1.84, aged dg = 382 nm, σg = 1.56. Residual plots of deviations from lognormal fits are aligned below each size distribution.

FIG. 3. Ground testing size distributions and residual plots for Kapton (a, b), lamp wick (c, d), and silicone (e, f). Upper images are SMPS measurements, open markers for unaged smoke, and solid for aged; solid curves represent the nonlinear least square fits. Size distribution parameters: (a) Kapton® baseline temperature test (510°C), unaged dg = 139 nm, σg = 1.78, aged dg = 209 nm, σg = 1.66. (c) Lamp wick high-temperature test (286°C), unaged dg = 171 nm, σg = 1.98, aged dg = 248 nm, σg = 1.75. (e) Silicone baseline temperature test (342°C), unaged dg = 257 nm, σg = 1.84, aged dg = 382 nm, σg = 1.56. Residual plots of deviations from lognormal fits are aligned below each size distribution.

Morphology results show that only Kapton and lamp wick are spherical aerosols, so they are better suited to calculating particle diameters from TEM images. Although the TEM images of silicone do not reflect spherical morphology, it is considered a spherical smoke aerosol as it consists of liquid droplets (Mulholland Citation1995). Regardless of shape, meaningful values of dav can be calculated from moment instrument results, and the material-specific calibration of the DustTrak with fundamental aerosol mass measurements provides moment method values for dm, which are valid for the nonspherical materials, Pyrell and Teflon (by Equations (1) and (2)). No significant discernable difference was noted between the morphology of the pyrolysis particles sampled in low gravity versus normal gravity for typical SAME flow conditions. A specific set of test points were run in low gravity with no flow through the SAME smoke generation duct, which resulted in significantly larger spherical particles. Details of these tests are outlined in the companion article (CitationMulholland et al. 2015).

SMPS Results for Spherical Smoke Aerosols

Five SAME materials were tested at two temperature levels: baseline and high temperature. Typical particle size distributions from the ground testing validation experiments of the unaged and aged smoke for more spherical aerosols are shown in . Pyrell® and Teflon® are not spherical aerosols and are not analyzed here; however, their mobility diameter size distributions appear in the SI. Particle size distributions are shown in the upper plots, Kapton (), lamp wick (), and silicone (). The plotted lines represent a lognormal curve-fit with the MATLAB Statistics Toolbox function “nlinfit,” which performs nonlinear least squares regression with the Levenberg–Marquardt algorithm (MATLAB version R2012a, The MathWorks Inc.). Residual plots are aligned directly below the size distributions showing deviations from lognormal fits. This visual test for “goodness-of-fit” would result in randomly scattered residual points, both above and below zero, for a good lognormal fit. It is common to observe a wedge-shaped spread of residuals, as in Kapton (), where the tails of the distribution have mostly small residual values, with a wider spread of residuals around the peak diameter. In general, the spread of the residuals is more compact for Kapton and lamp wick, indicating a better lognormal fit for these materials. Note that the unaged Kapton residuals are mostly positive up to 70 nm, which indicates that the data are less steep than the lognormal fit. The lamp wick residual plot () also shows a small but systematic deviation from the lognormal fit, which is evident by the change in sign of the unaged data residuals between 500 nm and 600 nm. The silicone residual plot () shows the least randomness, which indicates that the lognormal fit is less valid. The residuals change signs on both sides of the peak, indicating a shoulder in the small sizes (residuals go from positive to negative), the peak is offset from the fit (positive residuals around 400 nm), and the data are steeper than the lognormal fit in large sizes (negative residuals). Furthermore, the silicone residual plot has more noise and negative residuals at higher diameters, which may also be caused by losses from gravitational settling of these larger particles in the SAME aging chamber and/or the 55-gallon drum. Several systematic deviation patterns are observed, for example, where a shoulder in the distribution exists, a corresponding set of all positive residuals show a marked departure from the lognormal curve fit. This could be attributed to an improper multiple charge correction.

It is notable that most of the extreme positive and negative values of the residuals for all materials are for unaged smoke (open symbols), which suggests that as smoke ages within the timeframe of this experiment, it becomes more lognormal. As expected, the aged (black symbol) distribution moves to the right as aging increases the geometric mean diameter and σg decreases as the distribution narrows by coagulation. Log-probability plots were also used to assess lognormality of these aerosols in the SI.

The SMPS setting for a 10:1 sheath-to-aerosol flow rate ratio (3.0-lpm sheath, 0.3-lpm aerosol flow) captured the complete size distribution only for unaged Kapton® smoke, whereas other materials had larger size ranges, which were only completely captured by a 5:1 flow rate ratio (1.5-lpm sheath, 0.3-lpm aerosol flow) which extended the measurement range to 1000 nm. Silicone and Teflon® high temperature distributions were not completely captured by SMPS, even with a larger range of up to 1000 nm. Some SMPS distributions had an initial uptick in the small diameter tail, which is believed to be a sampling anomaly in the SMPS, possibly an artifact from the previous sample, as the scans were performed in rapid succession. This anomaly did not have a significant effect on the parameters obtained in the fitting of SMPS data.

TEM Particle Size Distribution Results

Particle size distributions were created by image analysis of the particles captured on TEM grids as an independent verification of particle measurements and moment diameter calculations. The particle-projected area-equivalent diameter was computed, which is considered to be equivalent to mobility diameter in the transition regime, even for nonspherical and agglomerate particles (Rogak et al. Citation1993; Chakrabarty et al. Citation2008). The TEM volume distribution was based on the assumption of spherical particles.

The limitations of Silicone TEM images outlined above preclude the creation of a reliable size distribution by microscopy, and the fractal nature of Teflon particles does not give size distribution statistics, which are directly comparable with spherical aerosols, thus only Kapton, lamp wick, and Pyrell were analyzed. TEM size distributions for Pyrell appear in the SI. Size distributions of a typical high temperature Kapton test are shown in . The unaged smoke has dg = 158 nm and σg = 1.68, and after 12 minutes of aging, dg increases to 210 nm and σg shrinks to 1.63. This supports observations on , which shows aged particles with a narrower size range and uniformly larger diameters.

FIG. 4. TEM particle size distributions for Kapton and lamp wick based on projected area equivalent diameters. (a) The effect of aging is shown for a high-temperature Kapton ISS experiment (574°C), circles are unaged smoke, and squares are aged smoke. (b) Ground test number and volume distribution for lamp wick aged test (286°C) from SMPS (circle symbols) and TEM (square symbols). The open symbols are number distribution, and closed ones are volume distribution. Best fit parameters for SMPS are dgn = 248 nm, dgv = 548 nm, σgn = 1.75 and σgv = 1.59, and for TEM are dgn = 123 nm, dgv = 812 nm, σgn = 2.70, and σgv = 1.65.

FIG. 4. TEM particle size distributions for Kapton and lamp wick based on projected area equivalent diameters. (a) The effect of aging is shown for a high-temperature Kapton ISS experiment (574°C), circles are unaged smoke, and squares are aged smoke. (b) Ground test number and volume distribution for lamp wick aged test (286°C) from SMPS (circle symbols) and TEM (square symbols). The open symbols are number distribution, and closed ones are volume distribution. Best fit parameters for SMPS are dgn = 248 nm, dgv = 548 nm, σgn = 1.75 and σgv = 1.59, and for TEM are dgn = 123 nm, dgv = 812 nm, σgn = 2.70, and σgv = 1.65.

A TEM lamp wick particle size distribution of particles collected during ground testing is compared with SMPS results in . Two different smoke tests are compared but the heating temperatures are within 0.6°C, so the pyrolysis can be considered to be similar. For TEM analysis, 17 images containing 4668 particles were processed. TEM images displayed two types of particles: dark, high contrast particles and smaller, lower contrast particles. The faint particles could be an artifact on the TEM grid or be the result of poor image quality. Another explanation could be a secondary particle formation event from the pyrolysis products. For this size distribution analysis, all particles were counted without thresholding, so the small faint particles were included in the total, which reduced the geometric mean diameter and caused σgn to be significantly larger than the SMPS results. The SMPS and TEM volume distributions have a similar spread but do not agree well in the peak location. The largest size bin of this TEM analysis had 12 particles, which is considered statistically significant for this type of analysis (Hinds Citation1999). The SMPS size distribution ends at 1000 nm, but there is no upper size limit for particles in the TEM size distribution. Additional TEM size distribution results from ISS flight tests are tabulated in the SI.

DISCUSSION

Limitations of Aerosol Instrument Measurement Ranges

Calibration of SAME instruments (described in the SI) was intended to empirically account for differences in the ranges of the instruments. In the ground validation tests, however, the truncated distribution formula (Equation (5)) can shed light on the limitations and uncertainty of the SMPS measurement range for the fuels tested. Particularly when the particle size distributions are converted to surface area or volume distributions, the percentage of the distribution that is lacking can be significant. Furthermore, the upper end of the SMPS size distribution measurement may not be as reliable because it can be affected by poor counting statistics and these bins are more susceptible to multiple charge correction errors. Therefore, it is prudent to compare a more conservative upper SMPS limit of 700 nm along with a full recorded range to 1000 nm, to see the effect of the SMPS measurement range. Thus, if we were to consider the SMPS data to be most reliable one (having the least uncertainty) in the range of 23 to 700 nm, then the truncated distribution formula for Mp,rel can indicate what percentage of the distribution would be captured with this limitation. shows the percentage of the distribution captured by SMPS for two example materials, Kapton and silicone. Kapton is the best candidate for SMPS validation, as the highest percentages of each type of distribution are within the SMPS measurement range. For example, considering both the conservative 700-nm limit and the 1000-nm limit, the bar graph indicates that 98 to 100% of the distribution has been measured. Silicone smoke is not a good candidate for SMPS validation because of the lack of measurement range, particularly when converting to surface and volume size distributions. Notably in the volume distribution in , the black bar representing a 1000-nm upper size limit is only at 40%, indicating that 60% of the distribution is missed by the instrument, but when losses are considered and a range of 700 nm is relied upon, approximately 20% of the distribution is captured (the white bar).

FIG. 5. The percentage of the number, surface area, and volume particle size distributions captured by the SMPS (ground testing) for (a) low-temperature (511°C) unaged Kapton smoke, and (b) high-temperature (370°C) unaged silicone smoke. Black bars represent an SMPS upper limit of 1000 nm (as measured) and white bars represent a more conservative upper limit of 700 nm.

FIG. 5. The percentage of the number, surface area, and volume particle size distributions captured by the SMPS (ground testing) for (a) low-temperature (511°C) unaged Kapton smoke, and (b) high-temperature (370°C) unaged silicone smoke. Black bars represent an SMPS upper limit of 1000 nm (as measured) and white bars represent a more conservative upper limit of 700 nm.

Thus, the truncated distribution formula for Mp,rel can be a useful indicator of the suitability of an aerosol reference instrument and the level of uncertainty in measurements. If enough of the size distribution is known to obtain parameters for a lognormal fit from curve-fitting software, one can determine how comprehensive the size distribution measurement is, and whether conversion of the distribution will produce reliable results.

Comparison of SMPS Data, Discretely Calculated Moment Diameters versus Hatch–Choate Diameters Based on Lognormal Fit Parameters

A useful quantitative measure of the validity of the lognormal assumption is to start with one set of data and compare diameters calculated by two different methods. The SMPS data offer the opportunity to use grouped data discretely, and based on the lognormal fit values of dg and σg, the same diameters can be calculated with the Hatch–Choate equations.

The SMPS bin data (based on a 64-channel per size decade histogram) can be used in Equation (4) formulas to calculate diameters of average properties, which can then be compared with diameters calculated from Equation (3) using the geometric mean diameter and σg from the lognormal fit of the SMPS particle size distribution. This is equivalent to calculating Ntot, Ltot, and Mtot from SMPS binned data to obtain dav and dm, by Equations (1) and (2). Thus, the continuous distribution parameters used in the conversion equations will be compared with the grouped data, and the expectation is that these diameters will be equal if the lognormal assumption is valid. Two examples of these diameter comparisons are shown in , which compares the count mean diameter (also known as the number average, or d50 of the number distribution), surface area diameter, diameter of average mass, mass median diameter (d50 of the volume distribution), and the count median diameter (which is the geometric mean diameter, provided that the distribution is lognormal). As can be seen, there is a good agreement in all diameters for Kapton () but not as good agreement for silicone (). The diameters with the largest deviations are those having to do with the mass. This is not surprising, since these are heavily influenced by the large diameter particles, and often the SMPS raw counts in the upper bins of the tail have fewer than 10 particles, so there is a potential for discrepancies in discrete bin calculations due to insufficient statistics.

FIG. 6. Two examples of the comparison of diameters calculated from SMPS data (ground testing) in two ways: (1) calculated using discrete SMPS bin data (black bars), and (2) converted by the Hatch–Choate conversion equations using SMPS dg and σg (white bars) for (a) low-temperature, unaged Kapton smoke, and (b) low-temperature, unaged silicone smoke.

FIG. 6. Two examples of the comparison of diameters calculated from SMPS data (ground testing) in two ways: (1) calculated using discrete SMPS bin data (black bars), and (2) converted by the Hatch–Choate conversion equations using SMPS dg and σg (white bars) for (a) low-temperature, unaged Kapton smoke, and (b) low-temperature, unaged silicone smoke.

shows the results of all bar graphs from comparison of the three spherical aerosols on one plot (including unaged and aged diameters, at all temperatures tested). The black bars of are the y-axis quantity in and the white bars are the x-axis quantity. Data falling on the 1:1 reference line meet the lognormal assumption, whereas those that differ significantly do not. While some information is lost in this scatter plot versus the bar graphs (e.g., which data marker represents which moment diameter), the graph shows that for Kapton and lamp wick, the diameters calculated by both methods coincide and thus can be considered lognormal. It is evident that diameters above 500 nm, which are the higher moment diameters, and mostly silicone, do not coincide. Overall, the qualitative comparison of these diameters strongly suggests that the smoke particle size distributions for Kapton® and lamp wick can be considered lognormal and silicone should not.

FIG. 7. Comparison of moment method diameters by different techniques. (a) Ground testing diameters calculated from SMPS data in two ways: converted by the Hatch–Choate conversion equations using SMPS lognormal fit dg and σg plotted against diameters calculated using discrete SMPS bin data. All diameters analyzed as in Figure 6 are combined here for all materials and test conditions: open markers are unaged, solid are aged; gray represents baseline temperature, black represents high temperature tests, and marker shapes are Kapton: circle, lamp wick: square, and silicone: triangle. (b) ISS flight data comparison of diameter of average mass by TEM analysis and the moment method.

FIG. 7. Comparison of moment method diameters by different techniques. (a) Ground testing diameters calculated from SMPS data in two ways: converted by the Hatch–Choate conversion equations using SMPS lognormal fit dg and σg plotted against diameters calculated using discrete SMPS bin data. All diameters analyzed as in Figure 6 are combined here for all materials and test conditions: open markers are unaged, solid are aged; gray represents baseline temperature, black represents high temperature tests, and marker shapes are Kapton: circle, lamp wick: square, and silicone: triangle. (b) ISS flight data comparison of diameter of average mass by TEM analysis and the moment method.

Comparison of Size Distribution Parameters from TEM and Moment Method Measurements from Flight Data

Although meticulous calibrations were performed, a number of smoke aerosols measured in flight tests exceeded the calibration range of ionization detector which measures the first moment. As evident in the uncertainty analysis (see the SI), error in the first moment measurement has the largest influence on the resulting calculations. Kapton consists of smaller particles and was not affected by this shortcoming, so these tests provided the most reliable moment method results. Since the first moment measurement is not used in the calculation of dm, this quantity can be compared for other materials.

shows the comparison of the flight TEM diameter of average mass with the value calculated from the moment instrument data using Equation (2). Since the third moment instrument was calibrated for each smoke type with a direct-reading reference instrument, the measurements are assumed to be relatively shape-independent (see the SI). For the three materials in this graph, the moment method provides a reasonably good measurement for the diameter of average mass. Thus, it can be concluded that the zeroth and third moment instruments maintained their calibration sufficiently to quantify this moment average diameter from flight data.

CONCLUSIONS

The aerosols considered in SAME represent the most likely smokes that a spacecraft fire detector will have to detect. We have characterized these smoke aerosols, and conclude with the following observations:

  1. TEM analysis of the particles from five spacecraft revealed distinct morphologies ranging from nearly spherical (Kapton and lamp wick) to extended aggregates (Pyrell and Teflon).

  2. The silicone particles were not stable enough for TEM analysis.

  3. Successful size distributions from TEM analysis were obtained for the more spherical particles (Kapton and lamp wick), as well as for Pyrell, as the projected area-equivalent diameter is comparable with mobility diameter, even for nonspherical particles.

  4. SMPS measurements were made for all five materials.

  5. Comparison of SMPS and TEM size distribution measurements showed reasonable but not complete agreement.

  6. Comparison of TEM and moment measurement results from the space experiments showed good agreement for the three materials whose morphology was amenable to TEM analysis (Kapton, Pyrell, and lamp wick).

  7. Statistical analysis of SMPS measurements showed that the spherical particles, Kapton and lamp wick, can be characterized as lognormal.

  8. Although a direct comparison of TEM, SMPS, and moment instrument results was not possible, the observed sizes from each system were quite consistent, given the constraints of each measurement type.

The moment method for the measurement of size distribution parameters relies on two assumptions: spherical particles and a lognormal distribution. These conditions were reasonably met in two of the five materials tested (Kapton and lamp wick). However, using the output from the calibrated SAME moment instruments, one is able to partially characterize the aerosol by determining dm and dav for any particle morphology. Within the limitations of spacecraft fire detection, the moment method was considered as a candidate for smoke aerosol measurement and has been proved moderately effective.

The smokes observed for these spacecraft materials cover a broad range in particle size. Ambient aerosols in spacecraft include significantly larger particles than on the Earth, as gravitational settling is absent, and smoke detectors must distinguish between background aerosols and smoke in order to prevent false alarms. Therefore, the typical background aerosols in manned spacecraft should be characterized and taken into account for smoke detector designs. Spacecraft fire detection systems require years of maintenance-free operation. This will be an important challenge for future longer-term space missions, as the expertise and resources necessary to calibrate and/or repair aerosol instruments in flight would not be available.

SUPPLEMENTAL MATERIAL

Supplemental material for this article can be accessed on the publisher's website.

Supplemental material

UAST_1025124_Supplementary_File.zip

Download Zip (2.2 MB)

ACKNOWLEDGMENTS

The SAME experiment team and the crews of ISS increments 10, 13, 15, and 24 are gratefully acknowledged. The SAME project was conducted through the ISS Exploration Research Project of the Exploration Technology Development Program.

Notes

1 Certain commercial equipment, instruments, or materials are identified in this article to foster understanding. Such identification does not imply recommendation or endorsement by the National Aeronautics and Space Administration (NASA) or the National Institute of Standards and Technology, nor does it imply that the materials or equipment identified are necessarily the best available for the purpose.

REFERENCES

  • Brooker, J. E., Urban, D. L., and Ruff, G. A. (2007). ISS Destiny Laboratory Smoke Detection Model, International Conference on Environmental Systems, 07ICES-113, 2007, Chicago, IL.
  • Bukowski, R. W., and Mulholland, G. W. (1978). Smoke Detector Design and Smoke Properties, NBS Technical Note 973. National Bureau of Standards, Washington, DC.
  • Bukowski, R. W., Peacock, R. D., Averill, J. D., Cleary, T. G., Bryner, N. P., Walton, W. D., Reneke, P. A., and Kuligowski, E. D. (2003). Performance of Home Smoke Alarms, Analysis of the Response of Several Available Technologies in Residential Fire Settings, NIST Technical Note 1455, December 2003.
  • Chakrabarty, R. K., Moosmüller, H., Chen, L.-W. A., Lewis, K. A., Arnott, W. P., Mazzoleni, C., Dubey, M. K., Wold, C. E., Hao, W. M., Kreidenweis, S. M. (2010). Brown Carbon in Tar Balls from Smoldering Biomass Combustion. Atmos. Chem. Phys. 10:6363–6370.
  • Chakrabarty, R. K., Moosmüller, H., Garro, M. A., Arnott, W. P., Slowik, J. G., Cross, E. S., Han, J.-H., Davidovits, P., Onasch, T. B., and Worsnop, D. R. (2008). Morphology-Based Particle Segregation by Electrostatic Charge. J. Aerosol Sci. 39:785–792.
  • Chen, B. T., Namenyi, J., Yeh, H. C., Mauderly, J. L., and Cuddihy, R. G. (1990). Physical Characterization of Cigarette Smoke Aerosol Generated from a Walton Smoke Machine. Aerosol Sci. Technol. 12(2):364–375.
  • Cleary, T. G., Weinert, D. W., and Mulholland, G. W. (2003). Moment Method for Obtaining Particle Size Measures of Test Smokes, NISTIR 7050. National Institute of Standards and Technology, Gaithersburg, MD.
  • Hinds, W. C. (1999). Aerosol Technology, 2nd Ed.. Taylor & Francis, New York, NY.
  • Janhäll, S., Andreae, M. O., and Pöschl, U. (2010). Biomass Burning Aerosol Emissions from Vegetation Fires: Particle Number and Mass Emission Factors and Size Distributions. Atmos. Chem. Phys. 10:1427–1439.
  • Li, W., and Hopke, P. K. (1993). Initial Size Distributions and Hygroscopicity of Indoor Combustion Aerosol Particles. Aerosol Sci. Technol. 19:305–316.
  • Mack, L. A., Levin, E. J. T., Kreidenweis, S. M., Obrist, D., Moosmüller, H., Lewis, K. A., Arnott, W. P., McMeeking, G. R., Sullivan, A. P., Wold, C. E., Hao, W. M., Collett Jr., J. L., Malm, W. C. (2010). Optical Closure Experiments for Biomass Smoke Aerosols. Atmos. Chem. Phys. 10:9017–9026.
  • Maricq, M. M. (2013). Monitoring Motor Vehicle PM Emissions: An Evaluation of Three Portable Low-Cost Aerosol Instruments. Aerosol Sci. Technol. 47(5–7):564–573.
  • Moosmüller, H., Arnott, W. P., Rogers, C. F., Bowen, J. L., Gillies, J. A., Pierson, W. R., Collins, J. F., Durbin, T. D., Norbeck, J. M. (2001). Time-Resolved Characterization of Diesel Particulate Emissions. 1. Instruments for Particle Mass Measurements. Environ. Sci. Technol. 35:781–787.
  • Mulholland, G. W. (1995). Smoke Production and Properties, in SFPE Handbook of Fire, 2nd Edition, DiNenno, P. J., Beyler, R. L., Custer, P., Walton, W. D., Watts, J. C. L. M., Drysdale, D., and Hall, J. R., eds., Society of Fire Protection Engineers, Boston, Mass., pp. 2-217–2-227.
  • Mulholland, G. W., Meyer, M., Urban, D. L., Ruff, G. A., Yuan, Z., Bryg, V., Cleary, T., and Yang, J. (2015). Pyrolysis Smoke Generated under Low-Gravity Conditions, Aerosol Sci. Tech. 49(5):310–321.
  • Raabe, O. G. (1971). Particle Size Analysis Utilizing Grouped Data and the Lognormal Distribution. J. Aerosol Sci. 2:289–303.
  • Reist, P. C. (1984). Introduction to Aerosol Science. Taylor & Francis, New York, NY.
  • Rogak, S. N., Flagan, R. C., and Nguyen, H. V. (1993). The Mobility and Structure of Aerosol Agglomerates. Aerosol Sci. Technol. 18:25–47.
  • Xie, Q., Yuan, H., Song, L., and Zhang, Y. (2007). Experimental Studies on Time-Dependent Size Distributions of Smoke Particles of Standard Test Fires. Build. Environ. 42:640–646.
  • Zai, S., Zhen, H., and Jia-Song, W. (2006). Studies on the Size Distribution, Number and Mass Emission Factors of Candle Particles Characterized by Modes of Burning. J. Aerosol Sci. 37:1484–1496.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.