Assessing Wear of the Acetabular Cup Using Computed Tomography: an ex vivo Study

, , , , , , & show all
Pages 852-857 | Accepted 07 Jul 2005, Published online: 09 Jul 2009
 

Abstract

Purpose: To validate a clinically useful method for measuring acetabular cup wear using computed tomography (CT).

Material and Methods: Eight uncemented acetabular cups were scanned twice ex vivo using CT. The linear penetration depth of the femoral component head into the cup and the thickness of the remaining polyethylene liner were measured in the CT volumes using dedicated software. Two independent examiners twice assessed each volume. The CT measurements were compared to direct measurements using a coordinate measuring device and micrometer measurements.

Results: Accuracy of wear measurements expressed as penetration depth was ±0.6 and ±1.0 mm for the two examiners, respectively, with no significant differences between examiners, trials, and CT scans. Accuracy of measurements of remaining polyethylene was ±1.3 and ±1.0 mm, respectively, for the two examiners. Systematic differences between examiners were found, but no significant differences between trials and CT scans. These differences were due to different interpretations of metal artifacts in the volumes.

Conclusion: The proposed CT method for evaluating wear as head penetration depth allows for reliable wear detection at a clinically relevant level. Measurements of remaining polyethylene on CT volumes are not as reliable as wear measurements owing to metal artifacts.

Log in via your institution

Log in to Taylor & Francis Online

There are no offers available at the current time.

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.