Radiation Dose Reduction to Breast and Thyroid During MDCT: Effectiveness of an In-Plane Bismuth Shield

, , , , , , , & show all
Pages 562-567 | Accepted 03 Mar 2006, Published online: 09 Jul 2009
 

Abstract

Purpose: To evaluate dose reduction and image deterioration using in-plane bismuth breast-shielding and thyroid-shielding for MDCT.

Material and Methods: Skin and organ doses of thyroid and breast were measured with thermoluminescent dosimeters in a female Alderson-Rando Phantom with and without a 4-ply in-plane bismuth shield. Routine neck (120 kVp, 150 mAseff; 16×1.5 mm) and chest (120 kVp, 100 mAseff; 16×1.5 mm) scan protocols were simulated on a 16-row MDCT scanner in three different settings: without shielding, with the shield directly on the surface, and with a 1-cm-thick cotton spacer between surface and shield. Image noise was quantified and compared using the t test.

Results: On average, shielding resulted in a 47% organ-dose reduction for the thyroid and 32% for the breast. Placement of the spacer between shield and surface had no significant impact on the measured doses, but significantly decreased the image noise (P<0.05).

Conclusion: In-plane bismuth breast and thyroid shielding significantly decreases radiation dose in MDCT without deteriorating image quality.

Log in via your institution

Log in to Taylor & Francis Online

There are no offers available at the current time.

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.