Magnetic Resonance Imaging of the Femoral Neck Cortex

, , &
Pages 308-314 | Accepted 09 Nov 2006, Published online: 09 Jul 2009
 

Abstract

Purpose: To determine the in vivo precision and accuracy of magnetic resonance imaging (MRI) in evaluating the thin (˜mm) femoral neck cortex.

Material and Methods: Hip regions of 30 volunteers were scanned with standard 1.5T MRI twice with repositioning, and the images of the narrow-neck cross-sections were obtained. Two radiologists independently delineated periosteal and endosteal cortical boundaries, and several structural variables were determined. For precision, 95% limits of agreement and coefficients of variation were calculated. Accuracy was determined with phantoms.

Results: In vivo precision of the directly delineated measurements varied from 0.9% (periosteal circumference) to 1.6% (total area), while the precision of the calculated variables remained poorer, from 7.3% (cortical to total area ratio) to 15.1% (torsional rigidity). The root-mean-square error of the cortical thickness measurement was 0.15 mm.

Conclusion: Thin cortical bone at the narrowest location of the femoral neck can be delineated precisely and accurately with a standard clinical 1.5T MRI device. MRI provides a feasible tool for the assessment of mechanically important cortical bone at the femoral neck and may be of clinical utility in assessing hip fragility.

Log in via your institution

Log in to Taylor & Francis Online

There are no offers available at the current time.

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.