Abstract
We present a sensitive method for simultaneous determination of bisphenol A (BPA), benzophenone-3 (BP-3), 4-tert-octylphenol (t-OP), ortho-phenylphenol (OPP), four parabens (methyl, ethyl, propyl, butyl parabens) and five chlorophenols (2,4-dichlorophenol (2,4-DCP), 2,5-dichlorophenol (2,5-DCP), 2,4,5-trichlorophenol (2,4,5-triclorophenol), 2,4,6-trichlorophenol (2,4,6-TCP), and triclosan (TCS)), in human urine by high-pressure liquid chromatography (HPLC) mass spectrometry (MS). Samples were processed using enzymatic deconjugation of glucuronides followed by solid phase extraction (SPE) on a C18 cartridge and the eluate was concentrated. Analytes were separated by reversed-phase HPLC and then detected by atmospheric pressure chemical ionisation (APCI) MS and quantified by isotope dilution method. We describe details for optimisation of each step of the procedure. The sample treatment steps are straightforward and not labour-intensive and, therefore, permit a high sample throughput with excellent prospects for automation. This method shows low inter-day variation, and detection limits for most of the compounds are below 1 ng/mL in 1 mL of urine. The method accuracy was also verified by the analysis of proficiency testing urine samples.
Acknowledgements
Partial funding for this work was provided by the CDC (Grant 5U38EH000481-03) to the California Environmental Contaminant Biomonitoring Program (or Biomonitoring California), which also is supported by the State of California.