Corneal layer thickness in keratoconus using optical coherence tomography

, , , ORCID Icon &
Pages 32-39 | Received 28 Feb 2021, Accepted 05 Apr 2023, Published online: 30 Apr 2023
 

ABSTRACT

Clinical relevance

Accurate thickness measurement of corneal layers using anterior segment OCT can be used to improve visual outcomes. Understanding its applications is essential for optometric practices to enhance eye care procedures.

Background

To evaluate the thicknesses of different corneal layers for identifying keratoconus (KCN) and subclinical keratoconus (SKCN) using spectral-domain optical coherence tomography (SD-OCT).

Methods

This prospective study analyzed 60 eyes with KCN, 48 eyes with SKCN, and 53 normal eyes. The central corneal thickness (CCT) and thicknesses of the epithelium, Bowman, stroma, and Descemet-endothelium layers were measured using SD-OCT. One way analysis of variance and the area under the curve (AUC) were used to evaluate the parameters. The Delong method was used to compare AUCs.

Results

In KCN, CCT and thicknesses of epithelium, Bowman, stroma, and Descemet-endothelium layers were 495.5 ± 41.7, 52.6 ± 6.4,11.5 ± 1.4, 415.5 ± 38.9, and 12.3 ± 1.7 µm, respectively. These thickness values were respectively 524.5 ± 33.3, 56.8 ± 6.8, 11.5 ± 1.6, 439.8 ± 30.6, and 12.4 ± 1.7 µm in SKCN and 563.8 ± 37.9, 57.7 ± 6.9, 12.2 ± 1.6, 469.5 ± 33.7, and 12.8 ± 2.1µm in normal group. Total cornea and stroma in KCN and SKCN, and epithelium in KCN were significantly thinner compared to the normal group (P < 0.001). The highest AUC values were observed for CCT in KCN (AUC 0.90) and SKCN (AUC 0.782). The diagnostic accuracy was significantly higher for stromal thickness in KCN (sensitivity 81.7%, specificity 73.6%, AUC 0.871) and SKCN (sensitivity 80.0%, specificity 56.6%, AUC 0.751) than other individual corneal layers (Delong, P < 0.001)                                    .

Conclusion

CCT can accurately distinguish keratoconus from normal eyes. However, central corneal stromal thinning was the most sensitive diagnostic index for early detection of SKCN. Developing standardized stromal maps may be helpful for detecting SKCN.

Disclosure statement

No potential conflict of interest was reported by the author(s).

Additional information

Funding

The author(s) reported there is no funding associated with the work featured in this article.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Purchase Issue

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 84.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.