Intrasession repeatability and agreement of the anterior corneal assessment provided by a multidiagnostic device

, , , & ORCID Icon
Received 07 Oct 2023, Accepted 23 Feb 2024, Published online: 17 Mar 2024
 

ABSTRACT

Clinical relevance

Multidiagnostic systems have recently appeared on the market. Knowledge of the repeatability and validity of any instrument is mandatory before its introduction in clinical practice.

Background

The aim of this work is to examine the intrasession repeatability of anterior pole measurements provided by the multidiagnostic device Wave Analyzer Medica 700 (WAM700) and agreement with Pentacam measurements in normal eyes.

Methods

In the right eyes of 113 participants, three repeat measurements of central keratometry, central corneal thickness, anterior chamber depth and corneal eccentricity were made with the WAM700 and Pentacam in random order. Intrasession repeatability and agreement were determined.

Results

Employing WAM700, intrasession repeatability for keratometry, central corneal thickness and anterior chamber depth was good (ICCs ≥ 0.992; CV 0.48–0.98%), yet worse than the values obtained for the Pentacam (ICCs ≥ 0.998; CV 0–0.33%). WAM700 showed excellent intrasession repeatability when used to measure the anterior chamber depth (Sw 0.03 mm). However, the repeatability of this device was inferior for central corneal thickness (Sw 4.24 μm) and keratometry measurements (Sw < 0.21 D) and was poor for corneal eccentricity (Sw 0.07; ICC 0.908; CV 14.58%). Agreement between WAM700 and Pentacam showed a high ICC for the keratometry measurements, central corneal thickness and anterior chamber depth (>0.972) but lower for corneal eccentricity (ICC 0.762).

Conclusions

In healthy eyes, the WAM700 multidiagnostic device showed good intrasession repeatability for keratometry, central corneal thickness and anterior chamber depth measurements. Agreement between WAM700 and Pentacam was good for the anterior chamber depth measurement. However, these instruments cannot be considered interchangeable for keratometry, central corneal thickness and eccentricity readings.

Disclosure statement

No potential conflict of interest was reported by the author(s).

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Purchase Issue

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 84.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.