ABSTRACT
Objective
To assess the repeatability and kinematic data of Modjaw® recordings from a group of asymptomatic patients.
Methods
Modjaw® mandibular kinematic data were recorded for 22 asymptomatic volunteers (DC/TMD questionnaire). Sagittal (SCI at 3 and 5 mm) and transversal (TCI at 4 mm) condylar inclination values of the instantaneous center of rotation were measured over two recording sessions. The repeatability of recordings was determined from the intraclass correlation coefficient (ICC), which is based on a one-way random-effects ANOVA.
Results
The repeatability of recordings was found to be good to excellent. The mean TCI in mediotrusion was 7.1° ± 5.1, and the mean SCI in protrusion was 51.07° ± 9.43.
Conclusion
The Modjaw® device reliably records the patient’s real hinge axis kinematics during functional mandibular movements. Data can be transferred with the patient’s personal reference plane and digital casts to the CAD/CAM software.
Disclosure statement
No potential conflict of interest was reported by the authors.