Functional Evaluation of a Novel Microwave Surgical Device in a Canine Splenectomy Model

, , , &
Pages 164-171 | Published online: 10 Jun 2019
 

Abstract

Aim: Partial splenectomy remains a difficult procedure even with the assistance of a sealing system. Recently, a scissor-like microwave energy device (Acrosurg) that provides seamless coagulation and manual cutting has been applied in surgery. The objective of this study is to evaluate the surgical capabilities of Acrosurg in partial splenectomy. Materials and Methods: Acrosurg and Harmonic Focus (HF) were employed for partial splenectomy in dogs without prior ligation of vessels. Cutting time, cutting area, and bleeding were recorded. Lateral thermal injury (LTI), possible complications was observed immediately and after 4 weeks. Results: The manual cutting mechanism of Acrosurg required a longer cutting time compared with the automatic cutting mode of HF (301.5 [243.2–527] vs. 114.5 [106.0–135.0] sec, p < 0.01). There was no statistical difference in cutting area or bleeding cases. Acrosurg achieved complete hemostasis in all cases, whereas the HF group failed to stop bleeding in two of the eight cases. The Acrosurg group exhibited a similar LTI compared with the HF group (3.0 [2.4–3.4] vs. 2.7 [2.3–2.9] mm, p = 0.151), but the LTI of the Acrosurg group tended to shrink more after 4 weeks (1.2 [1.0–1.3] vs. 1.7 [1.3–1.9] mm, p < 0.05). Conclusions: A microwave energy device enabled a partial splenectomy without vessel ligation. The combination of manual cutting and sealing capability helps not only provide an appropriate seal time by adjusting cutting timing adaptively but also potentially stop bleeding by using a microwave heating process unlike other energy devices.

ACKNOWLEDGMENTS

The authors would like to express their gratitude to Ms. Ikuko Arikawa and Ms. Miho Yamamoto, laboratory technicians at the Department of Surgery of Shiga University of Medical Science, for their assistance in sample treatment.

Declaration of Interest

Tohru Tani declares that he is the inventor of the microwave coagulation surgical instrument (MWCX), and his company receives royalties provided by the intellectual property of the MWCX. The other authors declare no conflict of interest. The authors alone are responsible for the content and writing of the article.

Additional information

Funding

This research received financial support from Nikkiso Co., Ltd. (Tokyo, Japan).

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access
  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart
* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.