Abstract
Purpose
Balance deficits after brain injury, including reactive recovery from unexpected perturbations, can persist well after rehabilitation is concluded. While traditional clinical assessments are practical, the anticipatory nature of the tasks may mask perceptible balance control. Computerized dynamic posturography can directly quantify capacity to respond to unexpected, external perturbations. This study examined the reliability of the computerized dynamic posturography assessment with the device PROPRIO® 4000 in adults with traumatic brain injury and created the minimal detectable change for its standardized test.
Methods
Ten adults (ages 21–55 years) with chronic (average 10 ± 6 years post-injury) severe (loss of consciousness 2–75 days) brain injury performed three trials of the Propriotest® on two separate days. The average of three trials and the best scores were used separately for analysis. Test-retest reliability was verified using Intraclass Correlation Coefficients with 95% confidence interval and standard error of measurement in relation to the Intraclass Correlation Coefficients at 95%. The minimal detectable change was calculated at 95% confidence level (minimal detectable change95) and Bland-Altman plots were created to express agreement between measurement days.
Results
The results exhibited excellent reliability for both average (Intraclass Correlation Coefficient of 0.969, standard error of measurement 50.9 points) and best (Intraclass Correlation Coefficient of 0.985, standard error of measurement 31.3 points) scores, with average and best minimal detectable change95 of 141.0 and 86.7 points, respectively.
Conclusions
Clinicians and rehabilitation researchers can use these findings to determine if a Propriotest® change score represents a true post-treatment effect with adults with chronic brain injury.
After brain injury, balance deficits are common and can persist well after completion of rehabilitation programs.
Computerized dynamic posturography allows for objective quantification of one’s capacity to respond to external perturbations.
The device PROPRIO® 4000 provides reliable quantification of balance deficits of community dwelling individuals who have experienced a severe traumatic brain injury.
The minimal detectable change scores created can assist clinicians and rehabilitation researchers detect whether a change in balance score represents a true effect of an intervention at post-treatment.
IMPLICATIONS FOR REHABILITATION
Disclosure of sources
The authors declare no conflicts of interest, including financial, consultant, institutional or any other conflict involving relationships that might lead to bias towards the submitted work.