Reliability, responsiveness, and validity of handheld dynamometry for assessing quadriceps strength in total knee arthroplasty

, , , , &
Pages 3070-3077 | Received 08 Jun 2018, Accepted 12 Feb 2020, Published online: 06 Mar 2020
 

Abstract

Purpose

To determine the reliability, responsiveness, validity, and agreement of handheld dynamometry, relative to electromechanical dynamometry, for assessing quadriceps strength surrounding total knee arthroplasty.

Methods

Fifty-six patients (48% female) undergoing total knee arthroplasty were assessed prior to surgery and at 2 and 6 weeks following surgery. Maximum isometric quadriceps force was assessed at each time point by handheld dynamometry and electromechanical dynamometry. Within-session test–retest reliability was determined by intraclass correlation coefficient (ICC). Standard error of measurement and relative standard error (RSE) values were calculated. Standardized response means were used to describe responsiveness. Pearson's correlations examined construct validity. Agreement was assessed by the Bland–Altman method.

Results

Both handheld dynamometry and electromechanical dynamometry demonstrated excellent test–retest reliability (ICC >0.90) and RSE (<15%). Both methods were responsive, with large postoperative standardized response means of 1.57 (handheld dynamometry) and 1.37 (electromechanical dynamometry). Pearson's correlations were moderate to strong. The Bland–Altman analysis revealed underestimation of force by handheld dynamometry, although this effect was diminished in the early postoperative period.

Conclusions

Our results suggest handheld dynamometry is a promising tool for monitoring quadriceps strength in patients pre- and post-total knee arthroplasty.

    IMPLICATIONS FOR REHABILITATION

  • This study provides evidence of high reliability and responsiveness of handheld dynamometry for assessing quadriceps strength throughout a clinically relevant time frame for patients with total knee arthroplasty.

  • There is construct validity of handheld dynamometry for measuring quadriceps strength, with moderate to high correlations between handheld dynamometry and electromechanical dynamometry, both pre- and post-total knee arthroplasty.

  • Rehabilitation professionals should be aware that handheld dynamometry and electromechanical dynamometry do not agree in terms of the force measurement itself; particularly at higher force values, handheld dynamometry underestimates force output relative to electromechanical dynamometry.

  • Overall, this study supports the use of handheld dynamometry for monitoring quadriceps strength in clinical settings for patient with total knee arthroplasty.

Disclosure statement

No potential conflict of interest was reported by the author(s).

Additional information

Funding

This project was funded in part by the National Institutes of Health [T32-AG000279] and approved by the University of Colorado Institutional Review Board (Aurora, CO, USA).

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 374.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.