Sebia Capillarys 2 versus the Helena Biosciences V8 capillary electrophoresis analyser for carbohydrate-deficient transferrin measurement: comparison and analytical evaluation

, , , &
Pages 23-27 | Accepted 17 Dec 2014, Published online: 04 Jan 2016
 

ABSTRACT

This study compares two automated capillary electrophoresis (CE) systems, the Capillarys 2 (Sebia, Surrey, UK) and V8 (Helena Biosciences, Tyne and Wear, UK) for the measurement of carbohydrate-deficient transferrin (CDT). Analytical imprecision was calculated for both platforms using internal quality control material from Sebia and Helena Biosciences, while a patient comparison was performed on 150 patient samples with CDT% levels ranging from 0.3% to 23.7%. Inter- and intra-assay imprecision between the two platforms were comparable. The correlation between platforms using patient samples was r2=0.985. However, there was a significant proportional bias at higher CDT concentration ranges, with the Helena system showing negative bias but good correlation over the clinically significant range. Analytical performances from both CE systems have been proven as suitable for routine laboratory use. The V8 CDT results were comparable to the Capillarys 2 in human sera over the clinical range of interest.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.