ABSTRACT
Background: The prothrombin time may be used to monitor the plasma concentration of rivaroxaban. However, there is variability in the responsiveness of rivaroxaban to different thromboplastins. We aimed to develop a rivaroxaban-monitoring method using the prothrombin time to reduce the differences in the sensitivity among reagents.
Methods: Rivaroxaban-spiked pooled normal plasma at a 0–1000 ng/ml concentration was used to generate a rivaroxaban-adjusted sensitivity index (SI) values, and was tested with three thromboplastins. The warfarin-adjusted international sensitivity index (ISI-warfarin), rivaroxaban-adjusted sensitivity index (SI-rivaroxaban), international normalized ratio (INR) calculated with ISI-warfarin, normalized ratio (NR) calculated with SI-rivaroxaban, and their coefficient of variances (CVs) were compared. The NR-rivaroxaban value was compared with the results of an anti-Xa assay.
Results: The ISI-warfarin and SI-rivaroxaban using different thromboplastins were 1.02 and 1.88, respectively, with Thromborel S, 0.90 and 1.00 using Recombiplastin 2G, and 1.30 and 1.15 using Neoplastin CI-plus. Between-thromboplastin variability expressed as CV were 6.3%–25.1% when expressed as INR-warfarin and 1.7%–4.7% when expressed as NR-rivaroxaban. CVs for the NR-rivaroxaban with another laboratory were significantly lower than those for INR-warfarin. Anti-Xa assay v NR-rivaroxaban correlation coefficients were 0.97–0.99.
Conclusion: Using a rivaroxaban-specific NR effectively minimises inter-thromboplastin variability. By utilizing a NR-rivaroxaban, standardized prothrombin time results could be rapidly obtained, especially useful in standardizing the therapeutic effect of rivaroxaban.
Summary Table
Disclosure statement
No potential conflict of interest was reported by the authors.