ABSTRACT
Recent advances in nebulizer technology of the evaporative light scattering detector (ELSD) allow the evaporation of trichlorobenzene (TCB) at much lower temperatures than the previous instruments, thus avoiding the sample loss in the lower molecular weight domain. Therefore, the new ELSD opens the possibility to correctly evaluate the molecular weight distribution (MWD) of polymers by gel permeation chromatography, after linearizing the ELSD signal intensity in function of concentration. To find the correct exponent parameter for linearization, it is necessary to take into account not only the peak area, but each point of the chromatogram. The evaluation method for this exponent parameter, found to be 1.61 for the analysis in TCB of polyolefin and polystyrene samples, is presented in this study. This value was verified by the excellent correlation found between the obtained MWD of a high-density polyethylene both with the ELSD and with the traditional differential refractive index detector.