Abstract
Hip resurfacing arthroplasty is an alternative to traditional hip replacement that can conserve proximal bone stock and has gained popularity but bone resorption may limit implant survival and remains a clinical concern. The goal of this study was to analyze bone remodelling patterns around an uncemented resurfacing implant and the influence of ingrowth regions on resorption. A computed tomography-derived finite element model of a proximal femur with a virtually implanted resurfacing component was simulated under peak walking loads. Bone ingrowth was simulated by six interface conditions: fully bonded; fully friction; bonded cap with friction stem; a small bonded region at the stem-cup intersection with the remaining surface friction; fully frictional, except for a bonded band along the distal end of the cap and superior half of the cap bonded with the rest frictional. Interface condition had a large influence on remodelling patterns. Bone resorption was minimized when no ingrowth occurred at the bone-implant interface. Bonding only the superior half of the cap increased bone resorption slightly but allowed for a large ingrowth region to improve secondary stability.
Acknowledgements
This study was partially funded by the Natural Sciences and Engineering Research Council of Canada [grant number 342890]. The authors thank T. Elgin for technical assistance in preparing the finite element models.