Abstract
A method of detection of PCBs in water, based on extraction/concentration with chips cut from an SPE disk-shaped C18 inverse phase coupled with a fluorescence detection and quantification directly on the phase has been evaluated. Experiments were performed with PCBs congeners IUPAC N° 3, 35, 67 and 169. Within the experimental conditions used : Hitachi F-4500 fluorometer, band-paths: 2.5 nm and extraction time: 60 min. the limits of detection were estimated to be respectively : 0.07, 0.4, 0.3 and 0.2 ppb. These values are within the range of PCB concentrations found in highly polluted waters and thus confer some interest to the method. However its sensitivity is not sufficient for application to moderately polluted natural water and the method needs some further improvements. Moreover, fluorescence quenching by humic substances was shown to substantially affect PCB's fluorescence intensity and must be taken into consideration for quantitative analysis.