The Effects of Pore Diameter and Ligand Chain Length on Fast Liquid Chromatography of Proteins and Peptides

, , &
Pages 633-647 | Received 11 Sep 1994, Accepted 18 Oct 1994, Published online: 23 Sep 2006
 

Abstract

Fast flow separations of proteins and peptides were successfully carried out on microbore columns containing reversed phase supports of three pore diameters. Increasing flowrates from 0.25 to 3 ml/min (1.2 mm/s–14.4 mm/s) did not show any adverse effects on the separations when SynChropak RPP-1000 and RPP-4000, which have 1000Å and 4000Å pore diameters, were used. At 3 ml/min, the 1000Å and 4000Å pore diameters yielded narrower peaks for proteins than the 300Å support, whereas the latter exhibited better resolution for peptides. Analyses in less than five minutes were achieved.

A series of ligand chains (C-4, C-8 and C-18) showed few differences in retention or resolution for either protein or peptide standards. Longterm stability of the 300Å and 4000Å supports was in excess of 25,000 column volumes when run at 3 ml/min with 0.1% trifluoroacetic acid.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.