Benefit of directional microphones for unilateral, bilateral and bimodal cochlear implant users

, ORCID Icon, &
Pages 147-157 | Published online: 13 Feb 2019
 

Abstract

Objective: To compare the standard T-Mic setting to UltraZoom and StereoZoom in 10 unilateral cochlear implant (CI) users, 10 bimodal device users and 10 bilateral CI users as well as a normal hearing (NH) reference group (n = 10).

Method: Speech reception thresholds were measured using the Oldenburg sentence test in noise. Speech was presented from the front at 0°, noise was presented from five loudspeakers spaced at ±60°, ±120°, 180° (setup A) or from four loudspeakers in the front hemisphere at ±30°, ±60° and one at 180° (setup B).

Results: There was a significant advantage for UltraZoom and StereoZoom for all groups in both setups. The largest advantage was for StereoZoom in the bilateral group (setup A, 5.2 dB, P < 0.001 and B, 3.4 dB, P < 0.001) There was a significant advantage for StereoZoom over UltraZoom in the bimodal group (setup A, P < 0.01 and B, P < 0.05) and in the bilateral group (P < 0.01, setup B only). The bilateral group performed as well as the normally hearing group in both setups and the bimodal group performed as well in setup A. There was a significant benefit of 1.8 dB for ClearVoice over UltraZoom alone for the unilateral group.

Conclusions: UltraZoom and StereoZoom provided a clinically and statistically significant benefit over the T-Mic condition. The largest gain was shown for StereoZoom in the bimodal and bilateral groups. The use of StereoZoom enabled the bilateral group to perform as well as the normally hearing group in both the challenging speaker setups. However, real life environments might provide an even greater challenge than the conditions tested here.

Acknowledgements

We would like to thank Mrs Keller for her help with the measurements.

Disclaimer statements

Contributors None.

Funding This study was funded by Advanced Bionics GmbH.

Conflicts of interest We would highlight the fact that the third author is employee of Advanced Bionics, the manufacturer of the device under investigation in this report.

Ethics approval Approval was given by the Charité Medical University Berlin (Universitätsmedizin Berlin) ethics committee. Approval: EA1/124/15.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 380.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.