Intercomparison of Instruments Used for Safety and Performance Measurements of Ultraviolet Germicidal Irradiation Lamps

, &
Pages 289-297 | Published online: 22 Jul 2010
 

Abstract

Public health concerns such as multidrug-resistant tuberculosis, bioterrorism, pandemic influenza, and sick building syndrome have brought about increased interest in the use of ultraviolet germicidal irradiation (UVGI) to prevent the spread of airborne infection. UVGI lamps require that radiometric measurements be performed to ensure their safe and effective use. This study evaluates 10 detectors that measure the ultraviolet radiation hazard of low-pressure mercury UVGI lamps, including a polychromator spectroradiometer, narrowband detectors designed to measure the ultraviolet radiation in a short range of wavelengths, and broadband detectors with a varying spectral response designed to follow the UV hazard action spectrum. The angular responses, spectral responses, and linearity of the detectors were measured and compared. The agreement between the measured angular responses and the ideal cosine responses varied widely among the detectors, and in general, the detectors with diffusing optics agreed significantly better with the ideal cosine response. The spectral responses at 254 nm also varied widely among the detectors, and, in general, the narrowband detectors agreed more closely with the 254 nm irradiances measured under the same conditions by a double monochromator spectroradiometer. All detectors displayed good linearity. The angular and spectral response data were then used to develop correction factors for the effective irradiance measurements of two UVGI sources, each measured at 10, 20, and 30 cm. The measured effective irradiances were compared with those measured by a double monochromator spectroradiometer with an integrating sphere input optic. Prior to correction, the effective irradiances measured by the detectors varied widely, ranging from 0.29 to 2.7 times those measured by the spectroradiometer. The application of cosine and spectral response correction factors significantly improved the agreement for the effective irradiances measured by all of the detectors, typically to within 10–20%. Awareness of these detector characteristics can play a key role in ensuring the accuracy of health hazard measurements of UVGI lamps.

ACKNOWLEDGMENTS

This project was supported in part by an appointment to the Internship/Research Participation Program for the U.S. Army Center for Health Promotion and Preventive Medicine administered by the Oak Ridge Institute for Science and Education through an agreement between the U.S. Department of Energy and the USACHPPM. Use of trademarked name does not imply endorsement by the U.S. Army but is intended only to assist in identification of a specific product.

The views expressed in this article are the views of the authors and do not reflect the official policy of the Department of the Army, the Department of Defense, or the U.S. government.

Notes

A The spectral response of the RPS200 was not evaluated.

A The spectral response of the RPS200 was not evaluated.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 61.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 148.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.