Intravenous lipid emulsion interference in coagulation testing: an ex vivo analysis

, , , , , & show all
Pages 441-445 | Received 10 Dec 2023, Accepted 13 Jun 2024, Published online: 03 Jul 2024
 

Abstract

Introduction

Intravenous lipid emulsion is used in the rescue treatment of certain poisonings. A complication is interference with laboratory analyses. The aim of this study was to determine the impact of intravenous lipid emulsion on routine laboratory analysis of coagulation parameters ex vivo and determine if any of the analytical techniques remain reliable.

Methods

Samples were obtained from 19 healthy volunteers and divided in triplicate. One sample served as a control, and the other two were diluted to simulate the treatment of an average adult with Intralipid® 20 per cent Fresenius Kabi 100 mL (dilution-1) or 500 mL (dilution-2). Coagulation tests performed were prothrombin time, activated prothrombin time, D-dimer concentration and fibrinogen. Coagulation testing was performed by three techniques. Test-1 was performed on a Sysmex CN6000 analyzer. Test-2 was performed with a manual mechanical endpoint method using the semi-automated Stago KC4 Delta. Test-3 involved high-speed centrifugation before repeat testing on the Sysmex CN6000 analyzer.

Results

For test-1, only nine (47 per cent) samples in dilution-1 could be analyzed for coagulation tests, and no coagulation tests could be analyzed for dilution-2 because of lipaemia. For test-2 and test-3, all samples could be analyzed, and all results of both testing methods fell within the limits of the laboratory reference range.

Discussion

Difficulties in laboratory analysis of patients having received intravenous lipid emulsion are due to multiple factors. Most automated coagulation analyzers use optical measurements, which can be unreliable in the presence of a high intravenous lipid concentration. By altering the lipaemia in the testing solution using high-speed centrifugation or by using manual mechanical endpoint detection, we were able to obtain reliable results. These findings are limited by the use of an ex vivo method and healthy volunteers.

Conclusions

This ex vivo model confirms that Intralipid® interferes with routine coagulation studies. It is important that clinicians are aware and inform their laboratories of its administration.

Disclosure statement

No potential conflict of interest was reported by the authors.

Data availability statement

Research data will be made available on reasonable request.

Additional information

Funding

The authors reported there is no funding associated with the work featured in this article.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 1,501.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.