Agreement between intraoperative anterior segment spectral-domain OCT and 2 swept-source OCT biometers

, , , , &
Pages 387-393 | Received 11 Feb 2021, Accepted 16 Mar 2021, Published online: 30 Mar 2021
 

ABSTRACT

Purpose: To evaluate the agreement of different biometric parameters obtained using intraoperative spectral-domain optical coherence tomography (SD-OCT) and two swept-source optical coherence tomography (SS-OCT) based biometers.

Methods: 102 eyes were assessed using the intraoperative SD-OCT integrated into the Catalys femtosecond-laser, and the IOLMaster 700 and Anterion SS-OCT-based-biometers. Central corneal thickness (CCT), anterior chamber depth (ACD), white-to-white (WTW), and lens thickness (LT) were measured.

Results: There were statistically significant differences for CCT, ACD, WTW and LT between devices (p < 0.001). The mean difference for ACD ranged from −0.067 to −0.250 mm, with the largest mean difference being between the IOLMaster 700 and Catalys. CCT mean differences ranged from 7 to 32 µm, with the largest mean difference being between the Anterion and Catalys. For WTW, the comparison between the IOLMaster 700 vs Catalys showed the largest mean difference (0.38 mm). However, the mean differences for LT from all three devices were quite similar, ranging from −0.02 to −0.08 mm.

Conclusions: SS-OCT biometers showed good agreement for ACD, CCT, WTW and LT. The SD-OCT showed ACD, CCT and WTW values that do not seem to be interchangeable with the SS-OCT biometers; however, this device did show excellent agreement in the case of LT.

Article highlights

  • The purpose of the present study was to analyze the agreement of several parameters obtained using an intraoperative spectral-domain OCT integrated into the Catalys Precision Laser System with those obtained with two SS-OCT biometers, the IOLMaster 700 and the Anterion.

  • A Bland–Altman graphical method was applied to enable pairwise comparisons with the average difference, 95% confidence interval and 95% limits of agreement.

  • SS-OCT biometers showed good agreement for ACD, CCT, WTW and LT. The SD-OCT of the Catalys showed ACD, CCT and WTW values that do not seem to be interchangeable with the SS-OCT biometers; however, this device did show excellent agreement in the case of LT.

Declaration of interest

All authors report research grants from Heidelberg Engineering GmbH and Carl Zeiss Meditec AG outside the submitted work. The authors have no other relevant affiliations or financial involvement with any organization or entity with a financial interest in or financial conflict with the subject matter or materials discussed in the manuscript apart from those disclosed.

Reviewer Disclosures

Peer reviewers on this manuscript have no relevant financial or other relationships to disclose.

Additional information

Funding

This paper was not funded.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 99.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 570.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.