Socket wall addition device in the treatment of recurrent hip prosthesis dislocation : Good outcome in 12 patients followed for 4.5 (1–9) years

, &
Pages 87-91 | Received 14 Dec 2004, Accepted 13 Jun 2005, Published online: 08 Jul 2009

Abstract

Background Recurrent dislocation in total hip replacement is difficult to treat and causes severe morbidity.

Patients and methods 12 patients suffering dislocations were reoperated with a socket wall addition device (anti-luxation ring) for the Lubinus SPII prosthesis, and were followed up after a mean of 4.5 (1–9) years with regard to redislocation, function and radiographic loos-ening.

Results 1 of the patients suffered a redislocation after almost 7 years of use. There was no loosening during the follow-up time. A Harris hip score of 87 (60–100), a health-related quality of life (EQ-5D) index of 0.8 (0.6–1.0) and total range of motion of 145° (125–165) indicate that the patients had a level of function comparable to that of age-matched hip surgery patients with no com-plications.

Interpretation The anti-luxation ring shows promising mid-term results and seems to provide an alternative to more extensive revision surgery for selected patients. ▪

Dislocation is one of the most common complica-tions of total hip replacement (THR), with an inci-dence of 0.6–4% (Masonis et al. Citation2002). Risk factors include indication for THR (Lee et al. Citation1998), surgical approach (Masonis et al. Citation2002), femoral head size (Woo and Morrey Citation1982), positioning of the prosthetic components (McCollum and Gray Citation1990, Barrack Citation2003), and experience of the surgeon (Hedlundh et al. Citation1996, Katz et al. Citation2001). Recurrent dislocations cause severe morbidity, but have been less well documented. After a primary dislocation, the incidence of recurrent dislocation has been reported to be 16–36% (Woo and Morrey Citation1982, Turner Citation1994, Joshi et al. Citation1998). The treatment for recurrent dislocations is a challenge. In the case of malpositioning of a prosthetic component, revision surgery is the best alternative (Soong et al. Citation2004). In cases where components are not obviously malpositioned, however, the treatment strategy becomes complex. Olerud and Karlström were the first to report treatment with a socket wall addition device (Citation1985). Since then, others have reported various results using similar techniques (). Since 1994, we have been using a socket wall addition device, the anti-luxation ring 180° (ALR) (Waldemar Link GmbH, Hamburg, Ger-many) (Figure), together with the Lubinus SPII prosthesis in selected cases. We studied the long-term functional and radiographical results of using this device.

Table 1.  Previous studies on socket wall addition devices

Patients and methods

14 patients who underwent revision surgery at the Department of Orthopedics, Gällivare Hospital, during 1994–2002, with application of ananti-luxation ring (ALR) as the only procedure, were identified in the computerized patient journal system and were invited to participate in the study. 12 patients (6 men) gave their written consent. The mean age of the patients at the time of the THR was 69(58–83) years. The indications for the THR had been primary years. The indications for the THR had been primary osteoarthrosis, revision due to aseptic loosening, femoral neck fracture and secondary osteoarthrosis. The indications for revision surgery with an ALR ring were recurrent dislocations, subdislocations, dislocation that was not reducible by closed reduction, and dislocation with suspected retroversion of the acetabular component. The mean time between the THR and the ALR surgery was 2 (0–4) years (). The patients were interviewed and given a questionnaire including the Harris hip score (HHS) (Harris Citation1969) and EQ-5D (Brooks Citation1996). Clinical examination was performed and radiographs were recorded. For loose femoral components, the Mullroy and Harris definition and 100% radiolucent lines were used (Mullroy and Harris Citation1997). For loose acetabular components, Hodkinson type 3 and 4 was used (Hodkinson et al. Citation1988). For statistical analysis, a paired Student's t-test was used at a rejection level of 5%.

Table 2.  Background data on patients, and results

Results

1 patient (no. 4) suffered a redislocation. Thispatient was operated on because of recurrent sub-dislocations, and has had the ALR for almost 7 years. None of the patients fulfilled the criteria used for a loose femoral or acetabular component. 8 of the patients had an HHS that was good (80–90) or excellent (90–100) (Harris Citation1969). Range of motion was limited (p < 0.05) in the ALR-operated hip compared to the contralateral side. Detailed results are presented in .

Discussion

The good mid-term results after this simple operation are somewhat surprising, with only 1 patient suffering redislocation during the 4.5 years of follow-up. The EQ-5D index was 0.8 and can be compared with the figure of 0.73 which was reported from a 6-year follow-up of 1,401 THR patients in southern Sweden, also including patients in all Charnley categories (A, B and C) (Charnley Citation1972, Swedish Total Hip Replacement Register Citation2002). Also, the HHS of 87, and 8 of 12 with good or excellent outcome, indicate that our patients do have a level of function that is similar to that of other age-matched hip surgery patients. The range of motion was reduced compared with the contralateral side. This is to be expected, because of the fact that the ALR itself limits the range of motion, and also probably because the hip has undergone surgery several times; scarring of the soft tissues may limit range of motion, but this seems to be of minor importance.

Different constrained designs limit range of motion and thereby increase forces over the bone-prosthesis interface. This may in turn lead to a risk of early loosening. Cobb et al. (Citation1996) reported a lower incidence of dislocation when using an elevated rim acetabular liner, but they also had con-cerns about the long-term results, because elevated acetabular liners can increase impingement and thereby increase component wear—with the possi-bility of osteolysis. However, in a follow-up study, these authors could not demonstrate any difference in the cumulative incidence of revisions due to loosening, when comparing elevated and normal acetabular rims (Cobb et al. Citation1997). None of the studies in describe any loose femoral or acetabular components, although only 3 of them have specifically addressed this issue. Likewise, we found no cases of loosening.

Recurrent dislocation of hip prostheses remains a problem. Since the first report of using a socket wall addition device by Olerud and Karlström in Citation1985, others have reported promising results using the same or similar techniques (Mogensen et al. Citation1986, Güngör and Hallin Citation1990, Bradbury and Mulligan Citation1994, Nicholl et al. Citation1999, Charlwood et al. Citation2002, Madan et al. Citation2002), whereas others have reported less promising results (Gie et al. Citation1989, Reikerås et al. Citation1989, Williamson et al. Citation1989, Watson et al. Citation1991). Various technical issues have been discussed, such as the thickness of the socket wall addition device (Graham et al. Citation1988) and the number of fixation screws (Watson et al. Citation1991). A special device has been developed for some prostheses. Several reports have been published on the use of the Wroblewski wedge augmentation (DePuy International Ltd, Leeds, UK) developed for the Charnley prosthesis (Bradbury et al. Citation1994, Nicholl et al. Citation1999, Madan et al. Citation2002). The ALR we used was developed for the Lubinus SPII pros-thesis, and we have found no other reports on the use of this device.

This study was supported by grants from Norrbotten Research Institute, Sweden.

No competing interest declared.

Author contributions

AE, JM clinical examination, radiographic analyzing and design of study. OS design of study.

  • Barrack R L. Dislocation after total hip arthroplasty implant design and orientation. J Am Acad Orthop Surg 2003; 11: 89–99
  • Bradbury N, Milligan G F. Acetabular augmentation for dislocation of the prosthetic hip. Acta Orthop Scand 1994; 65(4)424–6
  • Brooks R. EuroQol: the current state of play. Health Policy 1996; 37: 53–72
  • Charlwood A P, Thompson N W, Thompson N S, Beverland D E, Nixon J R. Recurrent hip arthroplasty dislocation. Acta Orthop Scand 2002; 73(5)502–5
  • Charnley J. The long-term results of low-friction arthroplasty of the hip performed as a primary intervention. J Bone Joint Surg (Br) 1972; 54(1)61–76
  • Cobb T K, Morrey B F, Ilstrup D M. The elevated-rim acetabular liner in total hip arthroplasty: relationship to postoperative dislocation. J Bone Joint Surg (Am) 1996; 78(1)80–6
  • Cobb T K, Morrey B F, Ilstrup D M. Effect of the elevated-rim acetabular liner on loosening after total hip arthroplasty. J Bone Joint Surg (Am) 1997; 79(9)1361–4
  • Gie G A, Scott T D, Ling R S M. Cup augmentation for recurrent hip replacement dislocation. J Bone Joint Surg (Br) 1989; 71 338
  • Graham G P, Jenkins A I R, Mintowt-Czyz W. Recurrent dislocation following hip replacement: brief report. J Bone Joint Surg (Br) 1988; 70-B: 675
  • Güngör T, Hallin G. Cup re-enforcement for recurrent dislo-cation after hip replacement. J Bone Joint Surg (Br) 1990; 72: 525
  • Harris W H. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. J Bone Joint Surg (Am) 1969; 51(4)737–55
  • Hedlundh U, Ahnfelt L, Hybbinette C H, Weckstrom J, Fredin H. Surgical experience related to dislocations after total hip arthroplasty. J Bone Joint Surg (Br) 1996; 78(2)206–9
  • Hodkinson J P, Shelley P, Wroblewski B M. The correla-tion between the roentgenographic appearance and operative findings at the bone-cement junction of the socket in Charnley low friction arthroplasties. Clin Orthop 1988, 228: 105–9
  • Joshi A, Lee C M, Markovic L, Vlatis G, Murphy J C M. Prognosis of dislocation after total hip arthroplasty. J Arthroplasty 1998; 13: 17–21
  • Katz J N, Losina E, Barrett J, Philips C B, Mahomed N N, Lew R A, Guadagnoli E, Harris W H, Poss R, Baron J. Association between hospital and surgeon procedure volume and outcomes of total hip replacement in the United States medicare population. J Bone Joint Surg (Am) 2001; 83(11)1622–9
  • Lee B P H, Berry D J, Harmsen W S, Sim F H. Total hip arthroplasty for the treatment of acute fracture of the femoral neck: Long time results. J Bone Joint Surg (Am) 1998; 80(1)70–5
  • Madan S, Sekhar S, Fiddian N J. Wroblewski wedge aug-mentation for recurrent posterior dislocation of the Charnley total hip replacement. Ann R Coll Surg Engl 2002; 84: 399–403
  • Masonis J L, Bourne R B. Surgical approach, abductor function, and total hip arthroplasty dislocation. Clin Orthop 2002, 405: 46–53
  • McCollum D E, Gray W J. Dislocation after total hip arthroplasty. Clin Orthop 1990, 261: 159–70
  • Mogensen B, Arnason H, Jonsson G T. Socket wall addition for dislocating total hip. Acta Orthop Scand 1986; 57: 373–4
  • Mullroy W F, Harris W H. Acetabular and femoral fixation 15 years after cemented total hip surgery. Clin Orthop 1997, 337: 118–28
  • Nicholl J E, Bintcliffe I W L, Koka S R, Addison A K. Acebular augmentation for the treatment of unstable total hip arthroplasties. Ann R Coll Surg Engl 1999; 81: 127–32
  • Olerud S, Karlström G. Recurrent dislocation after total hip replacement. J Bone Joint Surg (Br) 1985; 67(3)402–5
  • Reikeräs O. Acetabular arthroplasty to prevent dislocation after total hip replacement. Acta Orthop Scand 1989; 60(Suppl 231)17
  • Soong M, Rubash H E, Macaulay W. Dislocation after total hip arthroplasty. J Am Acad Orthop Surg 2004; 12: 314-21
  • The Swedish Total Hip Replacement Register 2002
  • Turner R S. Postoperative total hip prosthetic femoral head dislocations. Clin Orthop 1994, 301: 196–204
  • Watson P, Nixon J R, Mollan R A B. A prosthesis augmentation device for the prevention of recurrent hip dislocation. Clin Orthop 1991, 267: 79–84
  • Williamson J B, Galasko C S B, Rowley D I. Failure of acetabular augmentation for recurrent dislocation after hip arthroplasty. Report of 3 cases. Acta Orthop Scand 1989; 60(6)676–7
  • Woo Y G, Morrey B F. J Bone Joint Surg (Am) 1982; 64(9)1295–306

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.