Failing blood gas measurement due to methemoglobin forming hemoglobin variants: a case report and review of the literature

, &
Pages 167-170 | Published online: 21 Apr 2016
 

Abstract

Introduction: We present a case of an arterial blood gas sample analysis from a 33-year old woman where no oximetry results could be obtained using the Radiometer ABL800 FLEX device. Clinical history of this patient learned that she was carrier of a methemoglobin forming hemoglobin variant type Hyde Park (HbM Hyde Park) and raised the question whether or not this variant could be the cause of the errors obtained during analysis.

Materials and methods: A literature search was performed, focusing on methemoglobin forming hemoglobin variants and their influence on oxygenation measurements. An overview of the currently described methemoglobin forming hemoglobin variants is also included.

Results and discussion: In the presence of dyshemoglobins such as methemoglobin, techniques used to obtain parameters that reflect the patient oxygenation status, such as pulse oximetry and CO-oximetry can be influenced. In these cases, CO-oximetry is the preferred technique because it can compensate for this, in contrast to pulse oximetry. In case of the presence of methemoglobin originating from a hemoglobin variant, it is possible that CO-oximetry data cannot be calculated because the absorbance spectrum of this methemoglobin can differ from regular methemoglobin. Moreover, pulse oximetry devices are actually prone to erroneous results since pulse oximetry data will be calculated in these cases, but unreliable and should be avoided.

Conclusion: Methemoglobin forming hemoglobin variants are rare genetic mutations. However, they can possibly interfere with the calculation of CO-oximetry values. In these cases, pulse oximetry data should be avoided because they could lead to incorrect medical decisions.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 256.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.