Abstract
A sensitive, rapid and reproducible liquid chromatography-positive electrospray ionization- tandem mass spectrometry (LC-MS/MS) method was developed for the simultaneous quantification of Clomipramine and its active metabolite N-desmethyl Clomipramine in human plasma using Clomipramine-d3 and N-desmethyl Clomipramine d3 as an internal standard (IS). The analytes were extracted with solid phase extraction using cation exchange cartridges and analyzed on a BDS Hypersil C18 (100 mm × 4.6 mm, 5 μm) column. The mobile phase was composed of methanol: acetonitrile: 10 mM ammonium acetate (35:35:30 v/v/v). Clomipramine, N-desmethyl Clomipramine, Clomipramine d3 and N-desmethyl Clomipramine d3 detected in the multi-reaction monitoring (MRM) mode using LC-MS/MS (API 5500) productions of transitions as Q1→Q3 m/z 315.2→86.2, m/z 301.1→72.1, m/z 318.1→89.3 and m/z 304.2→75.2 respectively. The matrix effect, recovery, sensitivity, linearity, accuracy and precision were all evaluated over the concentration range of 0.500-200.000 ng/mL for both Clomipramine and N-desmethyl Clomipramine. In summary, the method was developed and validated successfully for intended to be use for the analysis of plasma samples of Bioavailability and Bioequivalence study of Clomipramine Tablets.