HPLC METHODS FOR ASPIRIN-CAFFEINE-BUTALBITAL AND ACETAMINOPHEN-CAFFEINE-BUTALBITAL MIXTURES IN TABLET DOSAGE FORMS USING NON-POROUS OCTADECYLSILANE COLUMNS

&
Pages 769-779 | Received 01 Aug 1999, Accepted 14 Sep 1999, Published online: 17 Aug 2006
 

Abstract

High performance liquid chromatography procedures using non-porous ODS columns were developed for the assay of aspirin-caffeine-butalbital (mixture 1) and acetaminophen-caffeine-butalbital (mixture 2). The separation and quantitation of Mixture 1 were achieved on a 3.0 μm non-porous silica ODS column at ambient temperature using a mobile phase of 98:2 v/v 50 mM phosphate buffer pH 3.0-acetonitrile at a flow rate of 1.5 mL/min with detection at 220 nm. The method showed linearity for aspirin-caffeine-butalbital in the 325–6500, 40–800, and 50–1000 ng/mL ranges, respectively. Intra- and inter-day RSD values were 0.19–1.72% and 1.30–1.49% for aspirin, 0.08–1.17% and 0.06–1.09% for caffeine, and 0.09–1.55% and 0.07–2.10% for butalbital, respectively. Accuracy of intra and inter-day were in the 0.70–1.27% and 0.20–1.13% for aspirin, 0.05–0.06% and 0.004–0.09% for caffeine, and 0.02–0.09 and 0.02–0.05% for butalbital, respectively.

The limits of detection for aspirin, caffeine, and butalbital were 5, 5, and 10 ng/mL, respectively, based on a signal noise ratio of 3 and a 50 μL injection. The separation and quantitation of mixture 2 were achieved on a 3.0 μm non-porous silica ODS column at ambient temperature using a mobile phase of 97:3 v/v 50 mM phosphate buffer pH 2.5-methanol at a flow rate of 2.0 mL/min with detection at 220 nm. The method showed linearity for acetaminophen, caffeine, and butalbital in the 650–6500, 100–800, and 125–1000 ng/mL ranges, respectively. Intra- and inter-day RSD values were 0.06–2.64% and 2.18–4.19% for acetaminophen, 2.24–4.76% and 4.09–4.12% for caffeine, and 1.65–3.94% and 3.33–3.40% for butalbital, respectively. Accuracy of intra and inter-day were in the 1.13–2.86% and 0.08–0.10% for acetaminophen, 1.12–4.62% and 0.04–2.75% for caffeine, and 0.34–3.48% and 0.02–2.30% for butalbital, respectively. The limits of detection for acetaminophen, caffeine, and butalbital were 20, 20, and 35 ng/mL, respectively, based on a signal noise ratio of 3 and a 10 μL injection.

ACKNOWLEDGMENT

The authors thank MICRA Scientific, Inc. for the gifts of the nonporous ODS columns used in this study.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 61.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 583.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.