VAN DEEMTER PLOT IN HIGH-SPEED COUNTERCURRENT CHROMATOGRAPHY WITH A FIXED VOLUME OF STATIONARY PHASE

, &
Pages 2929-2935 | Received 04 Jun 2000, Accepted 30 Jun 2000, Published online: 06 Feb 2007
 

Abstract

An adaptability of countercurrent chromatographic data to Van Deemter equation (H = A + B/u + Cu) is examined under a fixed volume of the stationary phase. Three test samples, epigallochatechin gallate (EGCG), gallocatechin-gallate (GCG), and epicatechin gallate (ECG) were separated using a two-phase solvent system of n-hexane/ethyl acetate/water (3.5:1:10, v/v/v) at various linear flow rates (u).

Our mathematical treatment produced a set of formulae, HEGCG = 6.5 + 1/u + 6u; HGCG = 6.5 + 1.3/u + 7u; and HECG = 6.5 + 2.5/u + 9u, which showed excellent fit to the experimental data for each component. Constants B and C show high correlation with the partition coefficient values of three components.

Based on the high correlation between log u and log H when u > 1, the plate height (H) for three components may be expressed by H = Duk, where D is the plate height at u = 1 cm/s and k, the constant for each component.

Acknowledgments

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 61.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 583.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.