Large Convoluted Tubing for Scale‐Up of Slow Rotary Countercurrent Chromatograph

, &
Pages 1991-2002 | Received 22 Jan 2003, Accepted 28 Feb 2003, Published online: 02 Jun 2007
 

Abstract

The chromatographic performance of 1.5 cm I.D. convoluted teflon tubing was studied at low rotary speeds ranging from 34 to 105 rpm. Results were as follows: (i) many solvent systems showed excellent retention of the stationary phase even at a high flow rate of the mobile phase; (ii) a good mixing of the two phases gave efficient solute partitioning resulting in high peak resolution; (iii) increasing the sample size caused only minor peak broadening; and (iv) increase in peak width and resolution using longer columns is predictable. Our data show that it is feasible to scale up a slow rotary countercurrent chromatograph with 1.5 cm I.D. convoluted tubing for industrial use.