On‐Line Flow‐Injection Detection System and Its Application to the Measurement of Ethanol Transfer and Osmosis Rate in Centrifugal Precipitation Chromatography

&
Pages 2979-2985 | Received 10 Jun 2004, Accepted 15 Jul 2004, Published online: 11 Jun 2009
 

Abstract

An on‐line flow‐injection detection system is newly designed and successfully used in the basic research on determination of the mass transfer and osmosis rates of ethanol through the membrane in centrifugal precipitation chromatography. The system improves the linearity range of the calibration curve, which showed linearity coefficient of correlation of 0.998 between 10 and 80% ethanol solution.