Effect of central corneal thickness on intraocular pressure and comparison of Topcon CT‐80 non‐contact tonometry with Goldmann applanation tonometry

, MS & , PhD
Pages 206-212 | Received 24 Mar 2017, Accepted 19 Jun 2017, Published online: 21 Apr 2021
 

Abstract

Background

To compare intraocular pressure (IOP) measurements obtained with the Topcon CT‐80 non‐contact tonometer (NCT) and Goldmann applanation tonometer (GAT), in different ranges of IOP in normal and glaucoma subjects, and to assess the influence of central corneal thickness (CCT) on the IOP measurements in Asian Indian eyes.

Methods

Four hundred and two eyes of 402 subjects (193 newly diagnosed primary open angle glaucoma [POAG] and 209 normal) were enrolled for this prospective study. For each eye, IOP was measured with GAT by a glaucoma specialist and NCT by a trained optometrist. The IOP values were compared among the tonometers in the three different IOP ranges (≤ 12 mmHg, 13–20 mmHg and ≥ 21 mmHg) using Bland–Altman graphs. Correlation between GAT and NCT was assessed by Pearson correlation co‐efficient. CCT was measured with ultrasound pachymetry and its correlation with GAT and NCT was analysed using linear regression analysis.

Results

The mean paired difference of IOP between NCT and GAT was 1.556 ± 2.69 mmHg (r = 0.26, p = 0.006) at IOP range of ≤ 12 mmHg, −1.665 ± 2.6 mmHg (r = 0.51, p < 0.0001) in IOP range of 13–20 mmHg and −2.202 ± 3.44 mmHg (r = 0.82, p < 0.0001) in the IOP range of ≥ 21 mmHg. Linear regression analysis showed a mean IOP variation of 0.27 mmHg per 10 μm change in CCT for NCT (p < 0.0001) and IOP change of 0.19 mmHg per 10 μm change in CCT for GAT (p = 0.01).

Conclusion

In this study of normotensive and POAG subjects, the Topcon CT‐80 NCT showed an overestimation of IOP at the lower range and underestimation of IOP in normal and higher ranges of IOP. Clinicians should keep in mind that CCT influences IOP measurement with both types of tonometer and that the IOP readings obtained with these tonometers are not interchangeable.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Purchase Issue

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 84.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.