Comparison of central corneal thickness measurements with standard ultrasonic pachymetry and optical devices

, MD & , MD
Pages 126-130 | Received 02 Jul 2018, Accepted 18 Oct 2018, Published online: 21 Apr 2021
 

Abstract

Background

To compare the repeatability and agreements of central corneal thickness measurements of healthy individuals obtained by Scheimpflug‐Placido topographer (Sirius), anterior segment spectral domain optical coherence tomography (Spectralis) (AS‐OCT), optical biometry (AL‐Scan) and ultrasonic pachymetry.

Methods

Sixty‐four eyes of 32 subjects with no ocular or systemic diseases were included in this study. Central corneal thickness measurements performed with Sirius, AS‐OCT, optical biometry AL‐Scan and ultrasonic pachymetry were compared. Bland–Altman analysis was used to demonstrate agreement between methods. Intra‐examiner repeatability was assessed by using intraclass correlation co‐efficients (ICCs).

Results

Sixty‐four eyes of 32 patients were enrolled (25 male and seven female). The mean age was 23.8-years (range 21–28-years). The mean central corneal thickness was 560.8 ± 37.2 μm, 528.8 ± 32.0 μm, 546.4 ± 33.8 μm, 543.4 ± 35.8 μm for the ultrasonic pachymetry, optical biometry AL‐Scan, Sirius and AS‐OCT values, respectively. The thickest mean central corneal thickness (560.8 ± 37.2 μm) was obtained from ultrasonic pachymetry. The thinnest mean central corneal thickness (528.8 ± 32.0 μm) was obtained from optical biometry AL‐Scan. All four modalities of central corneal thickness measurements correlated closely with each other. Intra‐examiner repeatability was excellent for all devices with ICCs > 0.90.

Conclusions

In conclusion, although measurements obtained by various methods correlate well, the measurements are not directly interchangeable. Between ultrasonic pachymetry and the three optical instruments tested, significant diferences can be seen. Therefore, the same imaging method should be used in corneal thickness follow‐ups.

ACKNOWLEDGEMENTS

No conflicts of interest were identified for any of the authors and no funding was received for this research.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Purchase Issue

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 84.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.