Abstract

Introduction: ‘Stenver's’ is the standard plain radiograph view to check the electrode position after cochlear implantation. However, a reproducible alignment for intra-individual comparison of electrode position using a true Stenver's alignment is not always straightforward to achieve, particularly for inexperienced radiographers, or on non-compliant children. In addition, two ionizing exposures in two different positions are required for bilateral cochlear implants.

Aims: To assess the efficacy of other, more simple, single-exposure radiographs for the assessment of electrode position.

Methods: Dummy electrodes and receiver units were bilaterally implanted into a cadaver. Plain radiographs in Stenver's views, and plain antero-posterior (AP) midline radiographic single views incorporating both implants were obtained with the tube at a range of angles tilted caudal and cranial to the perpendicular. Five internationally renowned cochlear implant surgeons were each asked if each radiograph gave adequate information for unilateral and for bilateral implants and they were asked to list in order of their favoured top three views.

Results: No surgeon thought that a single-exposure Stenver's view was adequate for assessment of the contra-lateral side. Consensus was that all AP views were fit for the purpose, with no preference given between each of the AP views. The ipsi-lateral Stenver's was considered to give better depth of insertion information.

Conclusion: There is no apparent advantage of caudal and cranial tilt angles over a straight perpendicular AP. A single-view AP radiograph is an alternative to Stenver's view for a post-unilateral and post-simultaneous bilateral cochlear implant check.

Acknowledgements

We would like to thank Cochlear Ltd for providing the dummy devices and the participating panel of surgeons.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 380.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.