Analysis of reproducibility and field size using gating system in 4D computed tomography

, &
Pages 63-67 | Received 02 Dec 2012, Accepted 13 Jun 2014, Published online: 27 Jun 2014
 

Abstract

Differences exist in each reproduced image since the scan types are different for the images acquired by axial, helical and cine scanning in accordance with computed tomography. This study was aimed to examine reproducibility using drive phantom in order to identify how accurately each scan type of computed tomography describes the movement of table. This study evaluated gating-based four-dimensional (4D)-computed tomography treatment planning by a comparison with common three-dimensional (3D)-computed tomography treatment planning and examining the change in treatment field size. The study used a self-manufactured drive phantom and a respiration gated drive motor to reproduce movements in the upward/downward direction and in the top/bottom direction. Following this, breathing-induced movement was applied to the self-manufactured model phantom and images were taken by Axial, Helical and cine scanning according to each computed tomography scan type. Reproducibility of the model and volume reflecting the movement was compared and analysed for accuracy using the RPT equipment. The scanning conditions according to each computed tomography scan type were set equally as follows: tube voltage 120 kV, tube current 200 mA, slice thickness and gap between slices 2·5 mm, inter scan delay 1·3, and pitch 1. The 3D-computed tomography and gating-based 4D-computed tomography images were obtained from patients who had undergone radiation treatment for chest and abdomen tumours in the oncology department. From the study results, the total volume of the marker was 88·2 cm3 and when a 3 cm shift in the top/bottom direction and a 0·3 cm shift in the upward/downward direction were taken into account, the total volume was 184·3 cm3. Total volumes by different computed tomography scan types were 135 cm3 for axial scanning, 164·9 cm3 for helical scanning and 181·7 cm3 for cine scanning. Regarding the presentation of reproducibility for movement, the cine scan type depicted the marker as closely as possible. The treatment field size was smaller in volume in the case of 4D-computed tomography treatment planning.

Acknowledgement

This work was supported by the 2013 Inje University research grant.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 61.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 305.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.