474
Views
7
CrossRef citations to date
0
Altmetric
Theme: Interventional Cardiology - Editorial

Optical coherence tomography: an arrow in our quiver

, &
Pages 539-541 | Published online: 10 Jan 2014

The rapid advances in the diagnosis of coronary artery disease during the second half of the 20th century may be fairly attributed to the introduction of new imaging techniques. Noninvasive techniques have great specificity and sensitivity for the identification of flow-limiting lesions. It is, however, the non-flow-limiting lesions with a large lipid burden and severe inflammation (i.e., high-risk plaque) that are responsible for the majority of acute coronary syndromes Citation[1,2]. These high-risk lesions are mostly identified with invasive imaging modalities. The era of invasive coronary imaging has enormously expanded over the last 10 years with the development of novel imaging techniques, enabling us to understand the complex aspects of the natural history of coronary artery disease while creating new therapeutic opportunities. Optical coherence tomography (OCT) has been an emerging and a quite promising invasive technique with a unique ability to visualize microstructural changes of the intima Citation[3].

How it works

OCT is a catheter-based, low-coherence interferometric technique utilizing near-infrared electromagnetic radiation that originates from a fiberoptic core. The technique is based on the calculation of the echo time delay of laser light waves that are reflected from biological tissues Citation[4]. OCT provides real-time ­high-resolution 2D cross-sectional images of coronary arteries with an axial resolution of 10–15 µm and a lateral resolution of 20–40 µm.

There are two types of OCT – time-domain OCT and frequency-domain OCT – representing the first and second generation of OCT, respectively. Frequency-domain OCT allows faster and better signal acquisition compared with time-domain OCT. A typical OCT pullback at a speed of 20–40 mm/s with a frame rate of approximately 100–200 frames/s has the potential of scanning nearly 60 mm of coronary arteries in less than 5 s. Such a rapid pullback eliminates the need for balloon occlusion and minimizes the amount of saline or contrast flush required for the removal of red blood cells, resulting in reduced risk of ­myocardial ischemia and clearer images Citation[5].

What has OCT got to offer us?

OCT is the only clinically available technique capable of depicting the arterial lumen and intima with high resolution.Studies have shown the ability of OCT to assess the presence of thrombus, plaque rupture, plaque neovascularization and, most importantly, high-risk plaque characteristics, such as lipid-rich necrotic core and the presence of macrophages in a thin fibrous cap Citation[3,6–10]. Furthermore, the use of OCT in post-stent implantation has allowed the early identification of stent malaposition, lumen dissection, stent protrusion and neointimal hyperplasia Citation[11–13]. In addition, OCT has provided important insights into late drug-eluting stent thrombosis Citation[14].

Relative limitations

Although OCT is a powerful modality for the imaging of coronary arteries, it has several limitations. Poor tissue penetration is the most important one. The penetration's depth is largely dependent on the tissue characteristics. Lipid-rich plaques and thrombus attenuate OCT signal, thereby diminishing our ability to evaluate the arterial wall beyond the internal elastic lamina Citation[4]. Sharing common disadvantages with other intravascular techniques, the OCT requires total clearance of red blood cells for fine-quality image acquisition. The relatively high cost of the procedure along, with the limited reproducibility and subjectivity in the interpretation of the results, are additional limitations. Multicenter OCT studies are now underway and they are anti­cipated for the formulation of consensus statements concerning the integration of OCT into diagnostic and therapeutic guidelines.

What is there to come?

Recently developed techniques related to OCT include polarization-sensitive OCT, a method providing a quantitative measure of birefringence. Birefringence constitutes a tissue property related to macromolecules and proteins, such as actin and collagen. Collagen depletion in the intima overlying a lipid-laden core is a critical component of plaque vulnerability Citation[15]. Pilot human studies are currently underway in the area of ophthalmology, testing the polarization-sensitive OCT Citation[16,17]. The application of this new technique in cardiovascular medicine appears to be quite challenging.

Geometrically correct 3D OCT is a novel imaging tool providing accurate 3D representation of coronary arteries in space by fusion of OCT and biplane angiography Citation[18]. The technique allows analysis of plaque anatomic characteristics, including fibrous cap thickness and lipid core size, as well as functional characteristics, including vascular remodeling and assessment of local endothelial shear stress. Studies have shown that low endothelial shear stress leads to atherogenesis and high-risk plaque development Citation[19–21]. 3D OCT-based calculation of local endothelial shear stress in combination with anatomic plaque characteristics such as fibrous cap thickness and lipid core size may facilitate the early ­identification and prompt treatment of high-risk plaques.

Micro-OCT is another recently developed OCT-related modality. Micro-OCT utilizes a very broad-bandwidth light source and common-domain OCT technology to provide images with 1-µm axial resolution and 2-µm lateral resolution. Understanding of the mechanisms involved in the formation and potential rupture of vulnerable plaque requires deeper tissue penetration in the cellular and subcellular level, and micro-OCT appears to have the potential to shed light towards this direction Citation[22]. Currently, micro-OCT is exclusively applied on an experimental level; its clinical application, however, will most likely expand our ­understanding of the vulnerable plaque pathophysiology.

A combination of OCT with other interventional and non-interventional modalities comprises another fascinating perspective creating new opportunities of local interventions for the management of coronary artery disease.

In conclusion, OCT has been an ‘arrow in our diagnostic quiver’ opening new perspectives in the diagnosis and management of coronary artery disease. However, whether this new tool has a role in reducing cardiovascular morbidity and mortality is unclear. As with every new diagnostic tool, rigorous clinical research is needed to show that the appropriate use of this method can further improve patient outcomes.

Financial & competing interests disclosure

YS Chatzizisis is funded by the Behrakis Foundation, the European Commission (FP7-PEOPLE-2009-RG, Grant 249303) and the Greek State Scholarships Foundation. The authors have no other relevant affiliations or financial involvement with any organization or entity with a financial interest in or financial conflict with the subject matter or materials discussed in the manuscript apart from those disclosed.

No writing assistance was utilized in the production of this manuscript.

References

  • Falk E, Shah PK, Fuster V. Coronary plaque disruption. Circulation 92(3), 657–671 (1995).
  • Stone GW, Maehara A, Lansky AJ et al.; PROSPECT Investigators. A prospective natural-history study of coronary atherosclerosis. N. Engl. J. Med. 364(3), 226–235 (2011).
  • Tearney GJ, Regar E, Akasaka T et al. Consensus standards for acquisition, measurement, and reporting of intravascular optical coherence tomography studies: a report from the International Working Group for Intravascular Optical Coherence Tomography Standardization and Validation. J. Am. Coll. Cardiol. 59(12), 1058–1072 (2012).
  • Low AF, Tearney GJ, Bouma BE, Jang IK. Technology Insight: optical coherence tomography–current status and future development. Nat. Clin. Pract. Cardiovasc. Med. 3(3), 154–162; quiz 172 (2006).
  • Bezerra HG, Costa MA, Guagliumi G, ­Rollins AM, Simon DI. Intracoronary optical coherence tomography: a comprehensive review clinical and research applications. JACC. Cardiovasc. Interv. 2(11), 1035–1046 (2009).
  • Yabushita H, Bouma BE, Houser SL et al. Characterization of human atherosclerosis by optical coherence tomography. Circulation 106(13), 1640–1645 (2002).
  • Jang IK, Bouma BE, Kang DH et al. Visualization of coronary atherosclerotic plaques in patients using optical coherence tomography: comparison with intravascular ultrasound. J. Am. Coll. Cardiol. 39(4), 604–609 (2002).
  • Jang IK, Tearney GJ, MacNeill B et al. In vivo characterization of coronary atherosclerotic plaque by use of optical coherence tomography. Circulation 111(12), 1551–1555 (2005).
  • Tearney GJ, Waxman S, Shishkov M et al. Three-dimensional coronary artery microscopy by intracoronary optical frequency domain imaging. JACC Cardiovasc. Imaging 1(6), 752–761 (2008).
  • Sánchez-Elvira G, Coma-Canella I, Artaiz M, Páramo JA, Barba J, Calabuig J. Characterization of coronary plaques with combined use of intravascular ultrasound, virtual histology and optical coherence tomography. Heart Int. 5(2), e12 (2010).
  • Inoue T, Shite J, Yoon J et al. Optical coherence evaluation of everolimus-eluting stents 8 months after implantation. Heart 97(17), 1379–1384 (2011).
  • Toutouzas K, Vaina S, Riga MI, Stefanadis C. Evaluation of dissection after coronary stent implantation by intravascular optical coherence tomography. Clin. Cardiol. 32(7), E47–E48 (2009).
  • Kedhi E, Joesoef KS, McFadden E et al. Second-generation everolimus-eluting and paclitaxel-eluting stents in real-life practice (COMPARE): a randomised trial. Lancet 375(9710), 201–209 (2010).
  • Guagliumi G, Sirbu V, Musumeci G et al. Examination of the in vivo mechanisms of late drug-eluting stent thrombosis: findings from optical coherence tomography and intravascular ultrasound imaging. JACC Cardiovasc. Interv. 5(1), 12–20 (2012).
  • Suter MJ, Nadkarni SK, Weisz G et al. Intravascular optical imaging technology for investigating the coronary artery. JACC Cardiovasc. Imaging 4(9), 1022–1039 (2011).
  • Yin B, Wang B, Liu S et al. High-speed polarization sensitive optical coherence tomography for retinal diagnostics. Presented at: SPIE Photonics West. San Francisco, CA, USA, 21–26 January 2012 (Poster 82132W).
  • Oh WY, Yun SH, Vakoc BJ et al. High-speed polarization sensitive optical frequency domain imaging with frequency multiplexing. Opt. Express 16(2), 1096–1103 (2008).
  • Chatzizisis YS, Giannoglou GD, Matakos A et al. In-vivo accuracy of geometrically correct three-dimensional reconstruction of human coronary arteries: is it influenced by certain parameters? Coron. Artery Dis. 17(6), 545–551 (2006).
  • Giannoglou GD, Soulis JV, Farmakis TM, Farmakis DM, Louridas GE. Haemodynamic factors and the important role of local low static pressure in coronary wall thickening. Int. J. Cardiol. 86(1), 27–40 (2002).
  • Chatzizisis YS, Jonas M, Coskun AU et al. Prediction of the localization of high-risk coronary atherosclerotic plaques on the basis of low endothelial shear stress: an intravascular ultrasound and histopathology natural history study. Circulation 117(8), 993–1002 (2008).
  • Chatzizisis YS, Baker AB, Sukhova GK et al. Augmented expression and activity of extracellular matrix-degrading enzymes in regions of low endothelial shear stress colocalize with coronary atheromata with thin fibrous caps in pigs. Circulation 123(6), 621–630 (2011).
  • Liu L, Gardecki JA, Nadkarni SK et al. Imaging the subcellular structure of human coronary atherosclerosis using micro-optical coherence tomography. Nat. Med. 17(8), 1010–1014 (2011).

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.