A multicenter study of a contrast-enhanced ultrasound diagnostic classification of breast lesions

, , , , , , , , , , , , , , , , & show all
Pages 2163-2170 | Published online: 15 Mar 2019

Abstract

Purpose

To evaluate a classification model of contrast-enhanced ultrasound (CEUS) and examine the characteristics of patients with false-negative diagnosis.

Patients and methods

A retrospective secondary analysis of a multicenter trial of CEUS for breast cancer diagnosis (from August 2015 to April 2017) was undertaken. Patients (n=1,023) with Breast Imaging Reporting and Data System 4–5 lesions on B-mode ultrasound underwent CEUS. Pathological diagnoses were available from surgical or biopsy specimens for correlation. Lesion maximum diameter (LMD), distance to the papilla (DtP), distance from the superficial edge of the lesion to the skin (DtS), distance from the deep edge of the lesion to the pectoralis muscle (DtPM), and body mass index (BMI) were evaluated.

Results

Median age and BMI were 48.0 and 41.2 years and 23.2 and 22.4 kg/m2 for patients with malignant and benign lesions, respectively. Overall sensitivity, specificity, and accuracy of CEUS for malignancy were 89.4%, 65.3%, and 75.8%, respectively. The patients with true-positive and false-negative diagnosis (ie, with malignant lesion) were older than those with false-positive and true-negative diagnosis (ie, with benign lesion). Patients with true-positive and false-positive diagnoses had higher BMI than patients with true-negative and false-negative diagnoses (P=0.004). Patients with true-positive and false-negative diagnoses had larger LMD and DtP, as well as smaller DtS and DtPM.

Conclusion

Older age, higher BMI, larger LMD and DtP, and smaller DtS and DtPM were associated with malignant lesions on CEUS. Patients with these characteristics should undergo further imaging.

Introduction

Breast cancer is the most common cancer in women and the fifth cause of death from cancer. Citation1 Breast cancer can be classified as invasive or in situ based on whether or not it invades the basement membrane of the duct of lobule. Citation2 In the US, about 95% of the new cases of breast cancer occur in women >40 years of age. Citation3 The worldwide incidence is about 43.3 new cases per 100,000 women, for a cumulative lifetime risk of 4.6%. Citation3 In the US, mortality from breast cancer ranges from 11.3 to 30 cases per 100,000, depending upon ethnic groups. Citation4 Early diagnosis of breast cancer is primordial to enable appropriate treatments and improve prognosis. Many countries offer breast cancer screening programs, and these programs have been shown to decrease breast cancer mortality by 15%–30%. Citation5 On the other hand, screening is associated with risks of false-positive, false-negative, and over-diagnosis. Citation6,Citation7 Ultrasound is often used as an adjunct to mammography for screening to enable early diagnosis of breast cancer worldwide. An optimal screening tool should have both high sensitivity and specificity, but both mammography and B-mode ultrasound have sensitivity <75%. Citation8 Newer and emerging technologies such as tomosynthesis, contrast-enhanced mammography, elastography, and contrast-enhanced ultrasound (CEUS) are promising. Citation9Citation11

Ultrasound characterizes a breast lesion using morphological descriptors such as shape, margin, and orientation. In Western countries, ultrasound is used as an adjunct to mammography. In Asia, it is also used as adjunct to mammography, but the frequency of use is probably higher because Asian women have relatively small and dense breasts, complicating the interpretation of mammography. Citation12,Citation13 Therefore, this study aimed to determine whether CEUS could improve the diagnostic rate and hence reduce the numbers of unnecessary biopsies. Furthermore, compared with mammography, ultrasound requires a less complex setting, and is portable and inexpensive, making it a modality of choice for developing countries with hard-to-reach areas in Asia without a permanent setting for breast cancer screening. In many countries, breast lesions are categorized using the Breast Imaging Reporting and Data System (BI-RADS), in which the malignant risk of lesions with BI-RADS 4 ranges from 2% to 95%, resulting in unnecessary invasive biopsy for many women. Raza et alCitation14 showed that the positive predictive value of BI-RADS 4 lesions was only 15.5%–20.0%.

Breast cancers typically have rich blood supply,Citation15 making them candidates for the use of contrast-based imaging modalities. CEUS involves the use of a contrast agent confined to the vascular lumen and visible by ultrasound. CEUS can significantly improve the diagnostic accuracy of breast lesions compared to conventional ultrasound, in particular for BI-RADS 4a and 4b lesions. Citation16,Citation17 Unfortunately, there is no unified classification criteria for CEUS analysis. Our team has previously established a CEUS classification based on six enhancement patterns, and this system has been shown to yield a diagnostic accuracy of 83.4%–86.4% in differentiating benign and malignant breast lesions. Citation17

The aim of the present multicenter study was to further evaluate the performance of this model using a large cohort, and to examine the characteristics of patients with different classes of diagnosis (true negative, true positive, false negative, and false positive). The results could help inform about applicability of the model for breast cancer diagnosis.

Methods

Study design and subjects

This was a retrospective secondary analysis of a prospective multicenter (Fujian Cancer Hospital, Sichuan Provincial People’s Hospital, Yunnan Cancer Hospital, Qujing City First People’s Hospital, Huangshi Central Hospital, Tangdu Hospital, Chengdu First People’s Hospital, and Yanan University Affiliated Hospital) trial of CEUS for breast cancer diagnosis, originally carried out between August 2015 and April 2017. The study has been approved by the ethics committee of Sichuan Provincial People’s Hospital (No 2016 14-1). The patients provided written informed consent for the original study and any eventual secondary data analysis. The original study and this substudy were carried out according to the principles of the Declaration of Helsinki.

All centers that participated in the original trial have screening programs for breast cancer. Before enrollment, all patients underwent B-mode ultrasound. Only those ≥14 years of age and with BI-RADS 4 and 5 breast lesions on B-mode ultrasound were recruited for the original CEUS trial. The exclusion criteria were: 1) contraindication or allergy to the contrast agent; 2) pregnant or breast-feeding; 3) coronary heart disease; 4) pulmonary or respiratory diseases; 5) any previous treatment for breast lesions (malignant or benign); 6) BI-RADS six lesions; or 7) diffuse lesions that cannot be measured or evaluated properly.

B-mode ultrasound and CEUS

The participating centers had different ultrasound systems (Philips iU22, Philips iU-Elite, Mylab 90, and GE LOGIQ E9), and each patient underwent ultrasound using a single system. B-mode ultrasound was performed with 5–12 MHz linear array transducers. CEUS was performed with 3–9 MHz linear transducers. The contrast agent used was Son-oVue (Bracco SpA, Milan, Italy) reconstituted with 5 mL of sterile saline. The ultrasound parameters were: mechanical index ≤0.13 MHz, single focus at the bottom of the image, and probe stabilized manually without pressure.

The ultrasound physicians from each center had a minimum of 15 and 2 years experience in breast ultrasound and CEUS, respectively. The following parameters were measured and recorded: lesion maximum diameter (LMD), distance to the papilla (DtP) (), distance from the superficial edge of the lesion to the skin (DtS) (), and distance from the deep edge of the lesion to the pectoralis muscle (DtPM) ().

Figure 1 Measurement of the distance to the papilla (DtP). The distance shown in the figure is 1.17 cm.

Figure 1 Measurement of the distance to the papilla (DtP). The distance shown in the figure is 1.17 cm.

Figure 2 Measurement of the distance from the superficial edge of the lesion to the skin (DtS) and the distance from the deep edge of the lesion to the pectoralis muscle (DtPM).

Note: The distances shown in the figure are DtS =0.752 cm and DtPM =0.191 cm.

Figure 2 Measurement of the distance from the superficial edge of the lesion to the skin (DtS) and the distance from the deep edge of the lesion to the pectoralis muscle (DtPM).Note: The distances shown in the figure are DtS =0.752 cm and DtPM =0.191 cm.

During examination, the maximum imaging plane with rich blood supply or irregular shape of the mass including its surrounding normal tissue was selected for CEUS. In the presence of multiple lesions in a single patient, only the largest one was evaluated by CEUS. After manual bolus injection of 4.8 mL of SonoVue via an antecubital vein, the selected plane was kept unchanged and real-time imaging was recorded for 2 minutes. All static and dynamic images were stored in the ultrasound systems and exported as JPEG and DICOM (digital imaging communications in medicine) files.

Image analysis

Image analysis was blinded and in accordance to the CEUS mode previously described. Citation17 This classification consists of three benign () and three malignant () classes: (A) hyper-enhancement with enlarged strength range, with or without irregular shape; B) hyper-centripetal enhancement with perfusion defect, with or without enlarged range; C) rapid or synchronous wash-in with hyper- or iso-enhancement, presence of penetrating vessels or crab claw-like pattern, with or without perfusion defect; ) rapid wash-in with hyper-enhancement, clear margin after enhancement without enlarged size; B) synchronous or slow wash-in with iso-enhancement, without distinction of margin and shape after enhancement; and C) synchronous or slow wash-in with hypo-enhancement.

Figure 3 The benign sub-models of contrast-enhanced ultrasound for breast lesions. (A) Rapid wash-in with hyper-enhancement and clear margin after enhancement without enlarged size. (B) Synchronous or slow wash-in with iso-enhancement, and no difference in margin and shape after enhancement. (C) Synchronous or slow wash-in with hypo-enhancement.

Figure 3 The benign sub-models of contrast-enhanced ultrasound for breast lesions. (A) Rapid wash-in with hyper-enhancement and clear margin after enhancement without enlarged size. (B) Synchronous or slow wash-in with iso-enhancement, and no difference in margin and shape after enhancement. (C) Synchronous or slow wash-in with hypo-enhancement.

Figure 4 The malignant sub-models of contrast-enhanced ultrasound for breast lesions.

Notes: (A) Hyper-enhancement with enlarged range, with or without irregular shape. (B) Hyper-centripetal enhancement with perfusion defect, with or without enlarged range. (C) Rapid or synchronous wash-in with hyper- or iso-enhancement, presenting penetrating vessels (as indicated by the white arrows) or crab claw-like pattern, with or without perfusion defect.

Figure 4 The malignant sub-models of contrast-enhanced ultrasound for breast lesions.Notes: (A) Hyper-enhancement with enlarged range, with or without irregular shape. (B) Hyper-centripetal enhancement with perfusion defect, with or without enlarged range. (C) Rapid or synchronous wash-in with hyper- or iso-enhancement, presenting penetrating vessels (as indicated by the white arrows) or crab claw-like pattern, with or without perfusion defect.

The lesions that met any of the malignant or benign classes were categorized as malignant or benign. The lesions that met both of benign and malignant classes were categorized as malignant. Any disagreements were solved by the director of the ultrasound department of each center.

Pathological data

All patients underwent surgery or 16G core needle biopsy. The specimens were examined by two pathologists (at least 3 years of experience in breast diagnosis) at each center. Diagnosis was based on the WHO criteria. Citation2

Diagnostic classification

Based on the CEUS classification and the pathological diagnosis, each patient was classified as true positive, true negative, false positive, and false negative. Sensitivity, specificity, and accuracy were calculated.

Data collection

Age, body mass index (BMI; calculated as kg/m2), number of lesions, lesion diagnosis, LMD, DtP, DtS, and DtPM were collected in the original trial.

Statistical analysis

Statistical analysis was conducted using SPSS 23.0 for Windows (IBM Corporation, Armonk, NY, USA). Continuous variables were analyzed using ANOVA (or the Kruskal– Wallis test) and the Bonferroni’s post hoc test. Categorical variables were analyzed with the chi-squared test. The null hypothesis was rejected at an α level of 5% (P<0.05).

Results

Characteristics of the patients and lesions

The study group included 1,023 female patients with 1,023 breast lesions evaluated by CEUS. presents the characteristics of the patients and their lesions.

Table 1 Characteristics of the patients and lesions

Diagnostic performance of CEUS

The overall sensitivity, specificity, and accuracy of CEUS for the diagnosis of breast lesions were 89.4%, 66.4%, and 75.8%, respectively.

Characteristics of the lesions according to the diagnostic class

shows the characteristics of the lesions according to the diagnostic class. The patients with pathologically malignant lesions (true-positive and false-negative diagnoses) were older (true positive: 49.5 years; false negative: 47.2 years) than those with pathologically benign lesions (false positive: 41.2 years; true negative: 41.2 years) (P<0.05). The patients with true-positive and false-positive diagnoses had a higher BMI (true positive: 23.1 kg/m2; false positive: 23.3 kg/m2) than patients with true-negative (22.5 kg/m2) and false-negative (22.4 kg/m2) diagnoses (P=0.004). The patients with true-positive and false-negative diagnoses also had larger LMD and DtP, as well as smaller DtS and DtPM compared with patients with true-negative and false-positive diagnoses (LMD: 23.4 and 19.6 vs 14.8 and 17.1 mm, respectively, P<0.05; DtP: 26.3 and 26.5 vs 20.4 and 23.1 mm, respectively, P<0.05; DtS: 5.9 and 4.8 vs 6.5 and 6.8 mm, respectively, P<0.05; DtPM: 2.4 and 3.4 vs 3.8 and 4.4 mm, respectively, P<0.05). These results suggest that lesions with false-negative diagnosis share the same characteristics than true-positive lesions, ie, high LMD and DtP, and small DtS and DtPM.

Table 2 Characteristics of the patients according to the type of final diagnosis

Discussion

CEUS has better accuracy than B-mode ultrasound for breast cancer diagnosis, but CEUS is limited by image analysis subjectivity. Therefore, this study aimed to evaluate the performance of a classification model of CEUS and examine the characteristics of patients with different classes of diagnosis (true positive, true negative, false negative, and false positive), in particular the characteristics of the patients with false-negative diagnosis. The results showed that the CEUS classification for breast lesions is promising. Age, BMI, LMD, DtP, DtS, and DtPM were associated with diagnostic accuracy. Lesions with false-negative diagnosis share the same characteristics as true-positive lesions, ie, high LMD and DtP, and small DtS and DtPM. These results suggest that lesions with a negative diagnosis but with these specific characteristics should potentially be investigated more thoroughly since there is a risk of positive findings at pathologic examination. Nevertheless, there is also risk of false-positive diagnosis and this concern should be studied further.

Early diagnosis of breast diseases allows for earlier treatment and better prognosis. Citation18 CEUS represents a method with good prospects for the early diagnosis of breast diseases and could be helpful to optimize the BI-RADS category. Citation16,Citation19 The enhancement patterns of lesions may be influenced by multiple factors such as microvessel density (MVD) and the intercellular substance structure. Citation20 How to improve the diagnostic performance of CEUS is still an issue in the clinical application of CEUS. In the present study, the accuracy of the proposed CEUS classification for the diagnosis of malignant lesions was higher than that for benign lesions. It is noteworthy that the CEUS features of benign and malignant breast lesions can overlap. Model A is strongly predictive for malignancy, as confirmed in previous studies. Citation19,Citation21 Nevertheless, in the present study, 4.6% of the women had false-negative diagnosis, and many benign lesions also show enlarged size after enhancement. Our results showed that 72.7% of atypical hyperplasia lesions, 69.8% of granulomatous mastitis lesions, and 60.0% of complex sclerosing adenosis lesions were misdiagnosed. Further study on the enhancement features could help refine the models and improve accuracy.

In the present study, the overall sensitivity, specificity, and accuracy of CEUS for the diagnosis of breast lesions were 89.4%, 66.4%, and 75.8%, respectively. Sensitivity was similar to that reported by a meta-analysis of studies of CEUS, with 86% sensitivity, but the specificity was lower than in the meta-analysis (79%). Citation22 Similar observation can be made with B-mode ultrasound (87% sensitivity and 72% specificity). Citation23 This could possibly be due to differences in image analysis and may suggest that our classification method could be improved. Nevertheless, the diagnostic performance of CEUS in the present study was similar to that of magnetic resonance imaging (92% sensitivity and 70% specificity), but better than mammography in terms of sensitivity (75% sensitivity and 71% specificity). Citation24 Additional studies in various populations are necessary to determine the exact value of the classification system suggested in the present study.

Age is one of the most important risk factors since breast cancer in Western countries occurs around 55–70 years of age,Citation4 but around 45–55 years of age in Asian women. Citation25 Thickness, duct width, and blood flow of the breast are different under the influence of hormones and according to the menopausal status. Citation26 The results showed that patients with malignant lesions were older than those with benign lesions. Kim et alCitation27 suggested that fibroadenoma in young women may enlarge rapidly and be accompanied by ductal epithelial hyperplasia, leading to increased number of peripheral microvessels, which may lead to false-positive CEUS results. Inflammatory responses could be accompanied by local vascular dilation and intraductal papilloma may display high enhancement, both potentially leading to false-positive CEUS diagnosis. Citation28 Therefore, to reduce false-negative diagnosis in the clinical application of CEUS, the physicians should be aware that elderly patients are more likely to have malignant lesions, while young patients are more likely to have benign lesions, as in the present study. Nevertheless, the results of the present study should be interpreted in the context of our specific study population (Chinese women). The results should be confirmed in other populations around the globe.

Overweight or obese women have a higher risk of breast cancer compared with lean women. Citation29 In the present study, patients with malignant lesions had a higher frequency of high BMI compared with women with benign lesions, and BMI was significantly different among the different diagnostic classes of CEUS. These results suggest that high BMI is associated with high risk of breast cancer as adipose tissue is considered as a major source of female hormones. Citation30 Indeed, after menopause, aromatase, the key enzyme in synthesis of estrogens, is mainly expressed in adipose tissue and aromatase inhibitors are part of the armament available against breast cancer. Citation30

Lesion size is a crucial variable during ultrasound examination. It is highly associated with tumor angiogenesis,Citation31 which is essential for tumor growth and infiltration. Citation32 The strength of CEUS enhancement is closely related to MVD. Citation33 In the present study, LMD was the largest in patients with true-positive CEUS diagnosis and the smallest in women with true-negative diagnosis, while those with false-positive and false-negative CEUS diagnosis had intermediate LMD values. This suggests that LMD is a major component of the qualitative analysis of CEUS images. In addition, some malignant lesions have no typical ultrasound features, neo-vascularization is not abundant, and heterogeneity is not obvious. Citation28 It is difficult to determine the strength or type of enhancement of small lesions. As highlighted by Zhang et al,Citation34 subjectivity plays a major role in determining whether there is heterogeneous enhancement of small lesions, and it may lead to high-false negative rates. For malignant lesions, with the enlargement of the lesion size, vessel heterogeneity is more prominent, spatial distribution is unbalanced, terminal fine vessels are increased, and tumor vessel branches are irregular. Citation35 Nevertheless, taken together, these results suggest that large lesions should be at least considered suspicious, even in the presence of negative CEUS diagnosis.

DtP has been known for a long time to be associated with malignancy. DtP >15 mm on mammography and ultrasound implies that the lesion is probably malignant. Citation36 On the other hand, this feature has not been associated with malignancy for shear wave elastography,Citation37 and there is no literature available regarding DtP in CEUS. In the present study, DtP of malignant lesions (true positive and false negative) was larger than for benign lesions. Additional studies are necessary to examine this feature in the context of CEUS.

Our results suggest that DtPM was smaller in malignant lesions (true positive and false negative) than in benign lesions (false positive and true negative). This could be explained by malignant nodules having larger diameter and longitudinal growth. In addition, DtS was smaller in malignant lesions than in benign lesions, but this parameter should be considered with caution because pressure from the transducer can greatly influence microcirculation perfusion, resulting in false-negative diagnosis. There can be important differences in DtS with ultrasound compared with mammography and MRI, possibly from the pressure exerted by the transducer on the breast. Citation38,Citation39 Although DtS is operator-dependent, it should be reproducible for different specific operators and could be used as a parameter for assessing malignancy.

There are some limitations to this study. Because this was a multicenter study, there was variability of equipment and physicians. The number of women <35 years of age was relatively small, and the value of CEUS in those patients still needs to be studied. No other imaging modality was used, and the value of the combination of multiple imaging techniques with CEUS needs to be examined. Finally, the use of different ultrasound systems is an important limitation of this study.

Conclusion

CEUS can synthesize multiple microvasculature and hemodynamic data of breast lesions. The rate of false-negative diagnosis could be influenced by age, BMI, LMD, DtP, DtS, and DtPM. The results support more features to improve the diagnostic performance of CEUS for breast lesions. Patients with false-negative diagnosis share the same characteristics as true-positive lesions, including high LMD and DtP, and small DtS and DtPM. These results suggest that patients with negative diagnosis but with these characteristics should possibly be investigated more thoroughly since there is a risk of positive findings. A larger international study including women of different nationalities is needed to validate this study.

Acknowledgments

The authors acknowledge the invaluable participation of the women as well as the help from the staff at all participating centers. The work was sponsored by Sichuan Provincial People’s Hospital (#2017HX001), Fujian Provincial Health Technology Project (#2015-1-11), and Science and Technology Program of Fujian Province, China (# 2018Y2003).

Disclosure

The authors report no conflicts of interest in this work.

References

  • BrayFFerlayJSoerjomataramISiegelRLTorreLAJemalAGlobal Cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countriesCA Cancer J Clin201868639442430207593
  • LakhaniSREllisIOSchnittSJTanPHvan de VijverMJWHO Classification of Tumours of the Breast (IARC WHO Classification of Tumours)LyonIARC Press2012
  • American Cancer SocietyBreast Cancer - Facts & Figures 2017–2018 Available from: https://www.cancer.org/content/dam/cancer-org/research/cancer-facts-and-statistics/breast-cancer-facts-and-figures/breast-cancer-facts-and-figures-2017-2018.pdfAccessed January 23, 2019
  • SiegelRLMillerKDJemalACancer statistics, 2018CA Cancer J Clin201868173029313949
  • NjorSNyströmLMossSEuroscreen Working GroupBreast cancer mortality in mammographic screening in Europe: a review of incidence-based mortality studiesJ Med Screen201219Suppl 1_334122972809
  • KorhonenKEWeinsteinSPMcDonaldESConantEFStrategies to increase cancer detection: review of true-positive and false-negative results at digital breast Tomosynthesis screeningRadiographics20163671954196527715711
  • ShahTAGurayaSSBreast cancer screening programs: review of merits, demerits, and recent recommendations practiced across the worldJ Microsc Ultrastruct201752596930023238
  • FioricaJVBreast cancer screening, mammography, and other modalitiesClin Obstet Gynecol201659468870927741212
  • CovingtonMFPizzitolaVJLoransRThe future of contrast-enhanced mammographyAJR Am J Roentgenol2018210229230029064748
  • ZhaoYXLiuSHuYBGeYYLvDMDiagnostic and prognostic values of contrast-enhanced ultrasound in breast cancer: a retrospective studyOnco Targets Ther2017101123112928260926
  • BotticelliAMazzottiEDi StefanoDPositive impact of elastography in breast cancer diagnosis: an institutional experienceJ Ultrasound201518432132726550070
  • RajaramNMariapunSErikssonMDifferences in mammographic density between Asian and Caucasian populations: a comparative analysisBreast Cancer Res Treat2017161235336227864652
  • OhuchiNSuzukiASobueTJ-START investigator groupsSensitivity and specificity of mammography and adjunctive ultrasonography to screen for breast cancer in the Japan strategic anti-cancer randomized trial (J-START): a randomised controlled trialLancet20163871001634134826547101
  • RazaSGoldkampALChikarmaneSABirdwellRLUS of breast masses categorized as BI-RADS 3, 4, and 5: pictorial review of factors influencing clinical managementRadiographics20103051199121320833845
  • YJLiGWenLYangX-LinZhangHeterogeneity of angioarchitecture and their hemodynamic changes in benign and malignant breast tumorsZhonghua Zhong Liu Za Zhi20093112427 Chinese19538864
  • WanCDuJFangHLiFWangLEvaluation of breast lesions by contrast enhanced ultrasound: qualitative and quantitative analysisEur J Radiol2012814e444e45021612882
  • LuoJChenJDChenQPredictive model for contrast-enhanced ultrasound of the breast: is it feasible in malignant risk assessment of breast imaging reporting and data system 4 lesions?World J Radiol20168660060927358688
  • CaplanLDelay in breast cancer: implications for stage at diagnosis and survivalFront Public Health2014228725121080
  • LiuHJiangYXLiuJBZhuQLSunQEvaluation of breast lesions with contrast-enhanced ultrasound using the microvascular imaging technique: initial observationsBreast200817553253918534851
  • LiXLiYZhuYFuLLiuPAssociation between enhancement patterns and parameters of contrast-enhanced ultrasound and microvessel distribution in breast cancerOncol Lett20181545643564929556301
  • XiaoXOuBYangHWuHLuoBBreast contrast-enhanced ultrasound: is a scoring system feasible? A preliminary study in ChinaPLoS One201498e10551725133534
  • HuQWangXYZhuSYKangLKXiaoYJZhengHYMeta-analysis of contrast-enhanced ultrasound for the differentiation of benign and malignant breast lesionsActa Radiol2015561253324436445
  • LiQHuMChenZMeta-analysis: contrast-enhanced ultrasound versus conventional ultrasound for differentiation of benign and malignant breast lesionsUltrasound Med Biol201844591992929530434
  • ZhangYRenHMeta-analysis of diagnostic accuracy of magnetic resonance imaging and mammography for breast cancerJ Cancer Res Ther201713586286829237918
  • ChenWZhengRBaadePDCancer statistics in China, 2015CA Cancer J Clin201666211513226808342
  • DenholmRDe StavolaBHipwellJHPre-natal exposures and breast tissue composition: findings from a British pre-birth cohort of young women and a systematic reviewBreast Cancer Res201618110227729066
  • KimSJParkYMJungSJSonographic appearances of juvenile fibroadenoma of the breastJ Ultrasound Med201433111879188425336474
  • ZhangJBreast Ultrasound DiagnosisBeijingPeople’s Medical Publishing House2012
  • Gathirua-MwangiWGZollingerTWMurageMJPradhanKRChampionVLAdult BMI change and risk of breast cancer: National Health and Nutrition Examination Survey (NHANES) 2005-2010Breast Cancer201522664865626350490
  • AhnJSchatzkinALaceyJVAdiposity, adult weight change, and postmenopausal breast cancer riskArch Intern Med2007167192091210217954804
  • LaiHWChenDRWuYCComparison of the diagnostic accuracy of magnetic resonance imaging with sonography in the prediction of breast cancer tumor size: a concordance analysis with histo-pathologically determined tumor sizeAnn Surg Oncol201522123816382325707494
  • GoubranHAKotbRRStakiwJEmaraMEBurnoufTRegulation of tumor growth and metastasis: the role of tumor microenvironmentCancer Growth Metastasis2014791824926201
  • WanCFDuJFangHLiFHZhuJSLiuQEnhancement patterns and parameters of breast cancers at contrast-enhanced us: correlation with prognostic factorsRadiology2012262245045922282183
  • ZhangJXCaiLSChenLDaiJLSongGHCEUS helps to rerate small breast tumors of BI-RADS category 3 and category 4BioMed Res Int20142014418
  • GolbabapourSPangWWGeorgeJPasupatiTAbdul-RahmanPSHashimOHChemically induced breast tumors in rats are detectable in early stages by contrast enhanced magnetic resonance imaging but not by changes in the acute-phase reactants in serumInt J Mol Sci20111221030104021541040
  • El SheikhHAlmetaherHAAbdallaHSShareefMMIntraductal breast masses: sonographic and mammographic predictors of malignancyThe Egyptian Journal of Radiology and Nuclear Medicine201546412711278
  • ChenYLGaoYChangCWangFZengWChenJJUltrasound shear wave elastography of breast lesions: correlation of anisotropy with clinical and histopathological findingsCancer Imaging20181811129622044
  • TelegrafoMRellaLStabile IanoraAAAngelelliGMoschettaMSupine breast us: how to correlate breast lesions from prone MRIBr J Radiol20168910592015049726689093
  • ParkJMYangLLaroiaAFrankenEAFajardoLLMissed and/or misinterpreted lesions in breast ultrasound: reasons and solutionsCan Assoc Radiol J2011621414920947291