Effectiveness of inhaler types for real-world asthma management: retrospective observational study using the GPRD

, , , , , , , & show all
Pages 37-47 | Published online: 28 Apr 2011

Abstract

Purpose:

Results of randomized controlled trials may not predict effectiveness of inhaled corticosteroids (ICS) in real-world clinical practice, where inhaler technique and device characteristics can influence effectiveness. We compared asthma outcomes for ICS delivered via three different inhaler devices: pressurized metered-dose inhaler (pMDI), breath-actuated MDI (BAI), and dry powder inhaler (DPI).

Patients and methods:

This retrospective database study evaluated 1-year outcomes for primary care patients with asthma aged 5–60 years prescribed their first ICS (initiation population) by pMDI (n = 39,746), BAI (n = 9809), or DPI (n = 6792), or their first ICS dose increase (step-up population) by pMDI (n = 6245), BAI (n = 1388), or DPI (n = 1536). Co-primary outcome measures were composite proxy measures of asthma control (no hospital attendance for asthma, oral corticosteroids, or antibiotics for lower respiratory infection) and severe exacerbations (unscheduled hospital admission, emergency room attendance, or oral corticosteroids). Outcomes were adjusted for potential confounding factors identified during a baseline year.

Results:

In the initiation population, adjusted odds ratios (95% confidence intervals [CI]) for asthma control, as compared with pMDIs, were significantly better for BAIs (1.08 [1.02–1.14]) and DPIs (1.13 [1.06–1.21]), while adjusted exacerbation rate ratios (95% CI) were 1.00 (0.93–1.08) and 0.88 (0.81–0.95), respectively. In the step-up population, adjusted odds of asthma control were 1.21 (1.05–1.39) for BAIs and 1.13 (0.99–1.29) for DPIs; adjusted exacerbation rate ratios were 0.83 (0.71–0.98) for BAIs and 0.85 (0.74–0.98) for DPIs, compared with pMDIs.

Conclusion:

Inhaler device selection may have a bearing on clinical outcomes. Differences in real-world effectiveness among these devices require closer evaluation in well-designed prospective trials.

Introduction

The findings of randomized controlled trials (RCTs) are integral to establishing the efficacy of therapies but may not predict their effectiveness in a real-world clinical setting, because RCTs are designed to maximize internal validity. Therefore, strict RCT inclusion criteria typically select idealized patient populations, free of comorbidities and with good adherence, and RCT protocols tend to require close patient monitoring at a level rarely possible or achieved in everyday clinical practice. Over 90% of patients with asthma in the community, such as those who smoke or have limited airway reversibility, do not meet eligibility criteria for most RCTs of asthma therapies.Citation1,Citation2

Several authors have expressed concern about the limited external validity, or generalizability, of many RCTs,Citation1Citation5 and there remains a need for effectiveness data to complement RCT findings. The use of diverse approaches to determine the appropriate use of therapeutic interventions is advocated by RawlinsCitation5 to replace an evidence hierarchy placing RCT results at the pinnacle of importance. Rigorously conducted observational studies can provide evidence to supplement that from RCTs.

Real-life data is particularly pertinent for inhaled therapies, for which additional factors such as delivery device characteristics and inhaler technique come into play and can influence the effectiveness of therapy.Citation6 Reviews of RCTs comparing asthma inhaler devices report no significant differences in clinical effectiveness according to device type.Citation7Citation9 However, patients enrolled in these trials usually received inhaler training and had to demonstrate and maintain proper inhaler technique throughout the trials. Yet in the real world, patients frequently make mistakes when using their inhaler devices,Citation6,Citation10Citation13 and errors in use of corticosteroid inhalers have been associated with poor asthma control.Citation14 Moreover, most inhaler RCTs are short term, and there is some evidence that, in the real world, inhaler technique deteriorates over time.Citation6 Indeed, results of an earlier observational study using a large primary care medical record database suggest that inhaler device choice does in fact affect asthma outcomes.Citation15

The objective of this retrospective database study was to compare outcomes for patients with asthma who were prescribed their first inhaled corticosteroid (ICS) therapy or an increase in ICS dose via pressurized metered-dose inhaler (pMDI), breath-actuated MDI (BAI), or dry powder inhaler (DPI). We examined respiratory-related clinical endpoints combined in composite measures of asthma control and exacerbations.

Methods

This 2-year retrospective observational study comprised a baseline year for defining cohorts and potential confounding factors, followed consecutively by an outcome year starting on the index date when patients received a prescription for first ICS or an increase in dose of ICS. The data source for the study was the General Practice Research Database (GPRD), a large computerized database containing de-identified longitudinal medical record data from over 450 participating general practices located throughout the United Kingdom (UK), including England, Scotland, Wales, and Northern Ireland.Citation16 The GPRD is well-validated and has been used frequently for respiratory research.Citation17Citation20

Medical records from the GPRD were examined for a 10.5-year period when all inhaler devices of interest were available, beginning January 1997 and ending June 30, 2007. Patients with asthma and aged 5–60 years on the index date were included in the study if they were continuously registered at the same practice for at least 2 years including the 12 months before and 12 months after the index date; and the practice had to be judged by the GPRD as having up-to-standard data during that time.Citation16 Evidence of asthma was defined as a recorded diagnosis of asthma or two or more prescriptions for ICS for asthma at more than one time point during the outcome year. Prescribing information recorded in the GPRD includes dosage, quantity, indication, and instructions. Patients were excluded if their record contained a diagnostic code for chronic obstructive pulmonary disease or any chronic respiratory condition other than asthma or if they were prescribed more than one ICS or a combination inhaler with long-acting β2 agonist (LABA) on the index date.

Two separate analyses were undertaken for 1) those patients receiving a first ICS prescription (initiation population) and 2) those receiving an increased dose of ICS (step-up population), the latter having at least one recorded prescription for ICS during the baseline year. Patients were included if they received ICS using only one device type, namely, a pMDI, BAI, or DPI, during the outcome year. Cost effectiveness analyses were also carried out for both the initiation and step-up populations, but these data are not the focus of this paper and are published separately.Citation21

The GPRD Independent Scientific Advisory Committee approved the use of GPRD data for this study.

Outcome measures

All outcomes were predefined before reviewing the data. The two co-primary outcome measures were a proxy for asthma control and severe exacerbation rate. The “primary measure of asthma control” was a composite endpoint, defined as including all of the following:

  1. no recorded hospital attendance for asthma (neither admission nor attendance at the emergency department, the outpatient department, or out of hours);

  2. no prescription for oral corticosteroids (acute or maintenance); and

  3. no consultations, hospital admissions, or emergency department attendance for lower respiratory tract infection requiring antibiotics.

A severe asthma exacerbation was defined, in line with a recent ATS/ERS task force definition,Citation22 as unscheduled hospital admission or emergency department attendance for respiratory disease or a prescription for oral corticosteroids. Exacerbations on the index date were included within the outcome data.

Secondary outcomes included another composite measure, “asthma control plus short-acting β2 agonist (SABA) use”, which factored the additional criterion of minimal reliever medication use into the co-primary control proxy (ie, control required a maximum average daily use of 200 μg albuterol or 500 μg terbutaline). Other secondary endpoints were the disaggregated outcomes comprising the composite measures and changes in therapy, including increase in ICS dose or use of additional therapy.

The ICS doses are reported ex-valve as the chlorofluorocarbon-beclomethasone (CFC-BDP)-equivalent, with doses of budesonide (BUD), fluticasone propionate (FP), BDP in solution (Qvar®, Teva UK), and mometasone (MOM) converted as necessary in the following dose ratios relative to CFC-BDP: CFC-BDP:BUD:FP:Qvar:MOM = 1:1:2:2:2.

Statistical analyses

Descriptive statistics were used to compare baseline factors among cohorts. The Kruskal–Wallis test was used to compare skewed continuous data between cohorts whilst the χ2 test was used to compare categorical factors.

Patients prescribed pMDIs were defined as the reference cohort, and patients prescribed BAIs or DPIs as the comparison cohorts. Results were examined separately for each population (initiation and step-up populations) during the outcome year, the primary period of analysis. All analyses were specified a priori.

Odds ratios for the dichotomized definitions of asthma control were calculated using a binary logistic regression model with control as the dependent variable and cohort, together with potential confounding factors (year of index date, age, sex, socioeconomic status, and comorbidity and treatment with medication that could affect respiratory outcomes), as explanatory variables. Socioeconomic status was that assigned, in quintiles, by the GPRD to each practice using the Index of Multiple Deprivation as a proxy measure. The GPRD has linked the socioeconomic status to small areas using the practice postcode where possible. Comorbidities were included via the Charlson comorbidity index score,Citation23 a weighted index that accounts for number and severity of comorbidities, as calculated for each patient using ICD-9 matching algorithms produced by CliniClue software (http://www.cliniclue.com/software).

The selected confounding factors were variables that were significantly different (P < 0.10) between cohorts at baseline.

The total number of severe exacerbations in the outcome period was compared between cohorts using a Poisson regression model to obtain estimates of exacerbation rates relative to the pMDI cohort. The model was adjusted for over-dispersion using robust standard errors and adjustments for potential baseline confounders.

Data were analyzed using SPSS version 17 (SPSS, Chicago, IL, USA), and differences between cohorts were considered to be significant if P < 0.05 and trends if 0.05 ≤ P < 0.10.

Results

depicts the identification of eligible patients in the GPRD who received a first prescription for ICS or increased dose of ICS. Baseline characteristics and asthma-related medical resource use of patients in the three cohorts of the initiation and step-up populations are summarized in and .

Figure 1 Selection of eligible patients in the database.

Abbreviations: BAI, breath-actuated inhaler; COPD, chronic obstructive pulmonary disease; DPI, dry powder inhaler; ICS, inhaled corticosteroid; pMDI, pressurized metered-dose inhaler.
Figure 1 Selection of eligible patients in the database.

Table 1 Baseline characteristics of the initiation and step-up populations

Table 2 Asthma-related measures and medical resource use of the initiation and step-up populations during the baseline year

Patients receiving a first prescription for ICS: initiation population

There were several statistically significant but small baseline differences among the three cohorts ( and ). Given the minimal clinical significance of these differences, as adjudicated by clinicians in the research team, these factors were handled through statistical modeling, which allowed adjustment of outcomes for potential confounding factors rather than necessitating subgroup or matched analysis.

The dose of ICS prescribed at the index date was significantly different among the three cohorts, with the same median dose in all three (400 μg/day), and wider variability in the BAI and DPI cohorts (). Spacers were prescribed to 17.7%, 4.0%, and 0.7% of patients in pMDI, BAI, and DPI cohorts, respectively. A statistically significant, but clinically small, difference in ICS dose as received (based on prescriptions dispensed over the 12-month period) was also evident during the outcome year (see ). Thus, the profile of dosing across the treatment cohorts was fairly similar in terms of actual use, with some differences in proportions of patients prescribed the lowest starting doses.

Table 3 Inhaled corticosteroid type and doses prescribed for and used by patients receiving a first prescription or increased dose of ICS

The adjusted odds ratios for asthma control during the outcome year were significantly greater for patients prescribed a BAI or DPI relative to those prescribed a pMDI (). The adjusted rate ratio for severe exacerbations was significantly lower in the DPI, but not the BAI, cohort, relative to the pMDI cohort ().

Figure 2 Odds ratios (95% CI) for achieving the composite measures of asthma control and rate ratio (95% CI) for severe exacerbations during the outcome year for patients who received a prescription for first ICS (top panel) or increased dose of ICS (bottom panel) using a BAI or DPI, with pMDI cohort as comparator.

Notes: Adjustments were made for following baseline parameters: aSex, age, GERD diagnosis, NSAIDs, acetaminophen, asthma consultations excluding oral steroids, antibiotics, oral steroid prescriptions, ICS dose at index date, year of index date; bGERD diagnosis, NSAIDs, acetaminophen, asthma consultations excluding oral steroids, SABA dose, antibiotics, oral steroid prescriptions, hospital asthma definite, ICS dose at index date, year of index date; cAge, NSAIDs, acetaminophen, asthma consultations excluding oral steroids, SABA dose, antibiotics, oral steroid prescriptions, hospital admissions, year of index date, ICS dose at index date; dSex, NSAIDs, acetaminophen, baseline SABA dose, antibiotics, oral steroid prescriptions, ICS dose at index date; eAcetaminophen, asthma consultations excluding oral steroids, SABA dose, antibiotics, oral steroid prescriptions, ICS dose at index date; fSex, NSAIDs, asthma consultations excluding oral steroids, SABA dose, antibiotics, oral steroid prescriptions, average baseline ICS dose.
Abbreviations: BAI, breath-actuated inhaler; BMI, body mass index; DPI, dry powder inhaler; GERD, gastroesophageal reflux disease; ICS, inhaled corticosteroid; NSAIDs, nonsteroidal anti-inflammatory drugs; pMDI, pressurized metered-dose inhaler; SABA, short-acting β2 agonist.
Figure 2 Odds ratios (95% CI) for achieving the composite measures of asthma control and rate ratio (95% CI) for severe exacerbations during the outcome year for patients who received a prescription for first ICS (top panel) or increased dose of ICS (bottom panel) using a BAI or DPI, with pMDI cohort as comparator.

When restricted SABA use was added to the composite measure, the odds ratios for asthma control plus SABA use significantly favored patients in BAI and DPI cohorts (). Results for disaggregated outcomes of the composite measures and percentages of patients with change in therapy are depicted in .

Table 4 Outcomes for patients initiating ICS or receiving an increased dose of ICS via pressurized metered-dose inhaler, breath-actuated inhaler, or dry powder inhaler

Patients receiving an increased dose of ICS: step-up population

As for the initiation population, baseline characteristics and measures that were statistically significantly different among the three cohorts ( and ) were included in the outcome analyses as potential confounding factors.

The median dose of ICS on the index date was lowest in the BAI cohort (500 vs 800 μg/day in the other two cohorts), and proportionately more patients in the BAI cohort were prescribed a dose of 201–400 μg/day, with the same interquartile range for dose in all three cohorts (). Spacers were prescribed to 12.8%, 1.5%, and 0.8% of patients in pMDI, BAI, and DPI cohorts, respectively. As for the initiation population, while differences in ICS doses among the cohorts were statistically significant, the profile of doses received across the treatment cohorts was fairly similar, with the highest median ICS dose used in the pMDI cohort, which included proportionately more patients who received a dose ≥800 μg/day.

Over the outcome year, the adjusted odds of asthma control was significantly greater in the BAI cohort than in the pMDI cohort, while the odds for the DPI cohort was not significantly different when adjusted for confounders (). The adjusted rate ratios for severe exacerbations were significantly lower for both BAI and DPI cohorts ().

There were no significant differences among cohorts in the odds for asthma control plus SABA use (). Other secondary outcomes are summarized in .

Discussion

For the real-world primary care patients included in this database analysis, the odds of achieving our a priori definition of asthma control during the outcome year were significantly better for patients initiating ICS therapy via BAI or DPI compared with a pMDI, and for those receiving an ICS dose increase via BAI compared with a pMDI. The co-primary outcome measure, rate of severe exacerbations, was significantly lower for those in the initiation population using a DPI and for those in the step-up population using a BAI or DPI, again as compared with a pMDI.

The goal of this study was to assess whether or not the asthma inhaler device is of material importance in the prescribing of ICS to patients in a real-world setting. The study results suggest that the inhaler device does in fact matter. Findings were generally consistent for BAIs and DPIs when compared with pMDIs and as prescribed for both initiation and step-up populations. Overall, the signal was stronger – in terms of the two co-primary outcome measures – with a DPI for patients initiating ICS and with a BAI for patients prescribed an increased dose of ICS.

While results of RCTs indicate little difference in the effectiveness of inhaler devices when used correctly and predominantly over the short-term,Citation8,Citation9 our findings suggest that real-world factors influence the effectiveness of these devices. Correct inhaler use is integral to the effectiveness of inhaled therapy,Citation6,Citation24Citation26 and it is possible that correct use is easier to learn and maintain with BAIs and DPIs than with pMDIs. Other factors that could influence effectiveness of therapy include adherence to therapy, patient preferences, and physician practices and preferences. As reflected in the prescribing patterns captured by this study, pMDIs are the most commonly prescribed inhaler devices in the UK, perhaps in part because they have been available for longer. Thus, another possible explanation for the observed differences, purely speculative, is that patients prescribed a BAI or DPI rather than a pMDI constitute a particular subgroup of patients in the view of their physicians and thus are monitored more closely, with resultant better outcomes, than those prescribed the more common pMDIs.

The delivery of medication to the target airways from an inhaler device depends on proper preparation of the device coupled with correct inhalation technique, both of which are device-specific. The correct use of pMDIs requires actuation of the dose synchronized with an inhalation that is slow and deep; and the most common mistakes made with pMDIs are failure to coordinate actuation with inhalation and too rapid an inhalation.Citation6,Citation25,Citation26 By contrast, dose emission from both BAIs and DPIs is actuated by the patient’s inhalation, which should be slow and deep for a BAI and, for a DPI, sharp and forceful to aerosolize the dry powder. Some patients may not be able to achieve sufficient inspiratory flow to actuate certain types of DPI.Citation27 Failure to exhale before actuation and to breath-hold after inhalation are mistakes common to all devices.Citation10,Citation12

Observational studies evaluating patients’ abilities to use inhaler devices correctly show inconsistent results, with some reporting mistakes to be more common with pMDIs than DPIs and BAIsCitation11,Citation12 and others reporting similar prevalence of inhaler misuse with pMDIs and DPIs.Citation13,Citation28 While coordination of actuation and inhalation is a recognized challenge with pMDIs and many consider these to be a difficult device to use,Citation6 all inhaler device types have specific requirements for proper preparation and use. Measures that can promote correct inhaler technique include provision of inhaler device training, repeat training with each revisit, and prescribing of the same type of inhaler device for both controller and reliever medications.Citation25,Citation26 Moreover, patients may have natural inspiratory patterns that accommodate one device better than another; this should be taken into consideration when choosing the inhaler type.Citation25

The strengths of this study include the large numbers of patients with available data from a validated data source. The 12-month outcome period minimized the effect of seasonal changes and allowed us to record measurable differences in less frequent outcomes such as severe asthma exacerbations. Study outcome measures were chosen prospectively. Because of the large patient numbers, even baseline differences among inhaler cohorts that were small from a clinical perspective were often statistically significantly different. We included baseline characteristics that were statistically significantly different among the three cohorts as potential confounding factors in the outcome analyses.

Severe exacerbations were identified in the database using two parameters proposed by a recent joint task force of the American Thoracic Society and European Respiratory Society to define exacerbations,Citation22 namely, unplanned or emergency care for asthma or a course of oral corticosteroid, both markers of poor asthma control. The primary measure of asthma control was a composite proxy that captured the absence in the database of recorded exacerbations or lower respiratory tract infection requiring antibiotics. The requirement for no recorded antibiotics for lower respiratory infection was based on the rationale that asthma exacerbations can be confused clinically for respiratory infection.Citation29,Citation30 A weakness of this proxy measure is evident in that three quarters (76%) and over half (61%) of patients in the initiation and step-up populations, respectively, met the primary asthma control definition during the baseline year before their health-care provider prescribed first ICS or an increased dose of ICS. Of course, the asthma control measure was not designed to supply information for a specific point in time, such as the date of the index prescription, but rather to summarize clinical information for a 1-year period.

The secondary composite measure was designed to add another aspect of asthma control to the primary outcome measure, namely, prescriptions for reliever therapy (SABA) as an indicator of asthma symptoms. The adjusted results for this measure significantly favored the BAI and DPI cohorts in the initiation but not the step-up population, as compared with the pMDI cohorts. We do not have a ready explanation for this difference between populations but note that SABA prescriptions are an imprecise measure of actual SABA use,Citation22 as patients may obtain extra prescriptions to have reliever therapy readily available in different places frequented during the day.

Our study limitations include those inherent to any observational study using retrospective data. While results were adjusted for multiple potential confounding factors, there remains the possibility of confounding factors not accounted for, such as particle size of ICS and lung deposition profiles, as well as factors that cannot be adjusted for, such as health-care professional preference; this can compromise the internal validity of the study. Moreover, our study data were recorded when CFC propellants were still in common use; there may be issues related to the newer hydrofluoroalkane (HFA) propellants that are not captured by this study. Thus, we consider our study results to be hypothesis-generating and suggest that they be interpreted in conjunction with findings from RCTs, pragmatic trials, and other observational studies.

Conclusion

Results of this retrospective database study reflect real-world effectiveness of ICS via three different delivery devices and suggest that selection of inhaler device type has a bearing on clinical asthma outcomes. The odds of asthma control were mostly better, and severe exacerbation rates lower, for patients initiating or increasing ICS therapy via BAI or DPI rather than a pMDI. These differences among inhaler devices require closer evaluation in well-designed prospective trials.

Acknowledgements

We thank Martha Joseph and Linda Kemp for their work on the initial analyses.

Access to data from the General Practice Research Database was funded by Merck and Co., Inc., and the analysis was funded by Teva Pharmaceuticals Ltd.

Disclosures

David Price has consultant arrangements with Boehringer Ingelheim, GlaxoSmithKline, Merck, Mundipharma, Novartis, and Teva. He or his research team have received grants and support for research in respiratory disease from the following organizations in the last 5 years: UK National Health Service, Aerocrine, AstraZeneca, Boehringer Ingelheim, GlaxoSmithKline, Merck, Mundipharma, Novartis, Nycomed, Pfizer, and Teva. He has spoken for AstraZeneca, Boehringer Ingelheim, Chiesi, GlaxoSmithKline, Merck, Mundipharma, Pfizer, and Teva. He has shares in AKL Ltd which produces phytopharmaceuticals. He is the sole owner of Research in Real Life Ltd.

John Haughney has received reimbursements for attending symposia, fees for speaking and organizing educational events, funds for research, or fees for consulting from AstraZeneca, Boehringer-Ingelheim, GlaxoSmith-Kline, Merck, Sharp and Dohme, Mundipharma, Novartis, Nycomed, Sanofi-Aventis, and Teva.

Elizabeth V Hillyer has done freelance writing work for Merck, Aerocrine, and Teva Sante (France).

Amanda J Lee receives payment for statistical consultancy from Research in Real Life Ltd.

Neil Barnes has lectured for or done consultancy for GlaxoSmithKline, AstraZeneca, Chiesi, Boehringer, Teva, and Nycomed and has received research support from GlaxoSmithKline, Novartis, and Schering-Plough.

All other authors report no conflicts of interest in this work.

References

  • HerlandKAkselsenJPSkjonsbergOHBjermerLHow representative are clinical study patients with asthma or COPD for a larger “real life” population of patients with obstructive lung disease?Respir Med200599111915672843
  • TraversJMarshSWilliamsMExternal validity of randomised controlled trials in asthma: to whom do the results of the trials apply?Thorax20076221922317105779
  • Van SpallHGTorenAKissAFowlerRAEligibility criteria of randomized controlled trials published in high-impact general medical journals: a systematic sampling reviewJAMA20072971233124017374817
  • RothwellPMExternal validity of randomised controlled trials: “to whom do the results of this trial apply?”Lancet2005365829315639683
  • RawlinsMDDe Testimonio: On the evidence for decisions about the use of therapeutic interventionsHarveian OrationOct-2008 edLondonRoyal College of Physicians2008
  • VirchowJCCromptonGKDal NegroRImportance of inhaler devices in the management of airway diseaseRespir Med2008102101917923402
  • BrocklebankDRamFWrightJComparison of the effectiveness of inhaler devices in asthma and chronic obstructive airways disease: a systematic review of the literatureHealth Technol Assess20015114911701099
  • BrocklebankDWrightJCatesCSystematic review of clinical effectiveness of pressurised metered dose inhalers versus other hand held inhaler devices for delivering corticosteroids in asthmaBMJ200132389690011668133
  • DolovichMBAhrensRCHessDRDevice selection and outcomes of aerosol therapy: evidence-based guidelines: American College of Chest Physicians/American College of Asthma, Allergy, and ImmunologyChest200512733537115654001
  • LavoriniFMagnanADubusJCEffect of incorrect use of dry powder inhalers on management of patients with asthma and COPDRespir Med200810259360418083019
  • LenneyJInnesJACromptonGKInappropriate inhaler use: assessment of use and patient preference of seven inhalation devices. EDICIRespir Med20009449650010868714
  • MolimardMRaherisonCLignotSDepontFAbouelfathAMooreNAssessment of handling of inhaler devices in real life: an observational study in 3811 patients in primary careJ Aerosol Med20031624925414572322
  • MelaniASZanchettaDBarbatoNInhalation technique and variables associated with misuse of conventional metered-dose inhalers and newer dry powder inhalers in experienced adultsAnn Allergy Asthma Immunol20049343944615562882
  • GiraudVRocheNMisuse of corticosteroid metered-dose inhaler is associated with decreased asthma stabilityEur Respir J20021924625111866004
  • PriceDThomasMMitchellGNiziolCFeatherstoneRImprovement of asthma control with a breath-actuated pressurised metred dose inhaler (BAI): a prescribing claims study of 5556 patients using a traditional pressurised metred dose inhaler (MDI) or a breath-actuated deviceRespir Med200397121912556005
  • General Practice Research Database http://www.gprd.com/home/default.asp. Accessed January 7, 2011.
  • JickSSKayeJAVasilakis-ScaramozzaCValidity of the General Practice Research DatabasePharmacotherapy20032368668912741446
  • SorianoJBMaierWCVisickGPrideNBValidation of general practitioner-diagnosed COPD in the UK General Practice Research DatabaseEur J Epidemiol2001171075108012530765
  • ThomasMvon ZiegenweidtJLeeAJPriceDHigh-dose inhaled corticosteroids versus add-on long-acting beta-agonists in asthma: an observational studyJ Allergy Clin Immunol2009123116121e1018986690
  • PriceDMartinRJBarnesNPrescribing practices and asthma control with hydrofluoroalkane-beclomethasone and fluticasone: a real-world observational studyJ Allergy Clin Immunol2010126511518e1e1020692026
  • KempLHaughneyJBarnesNCost-effectiveness analysis of corticosteroid inhaler devices in primary care asthma management: a real world observational studyClinicoEconomics Outcomes Res201027585
  • ReddelHKTaylorDRBatemanEDAn official American Thoracic Society/European Respiratory Society statement: asthma control and exacerbations: standardizing endpoints for clinical asthma trials and clinical practiceAm J Respir Crit Care Med2009180599919535666
  • CharlsonMEPompeiPAlesKLMacKenzieCRA new method of classifying prognostic comorbidity in longitudinal studies: development and validationJ Chronic Dis1987403733833558716
  • CochraneMGBalaMVDownsKEMauskopfJBen-JosephRHInhaled corticosteroids for asthma therapy: patient compliance, devices, and inhalation techniqueChest200011754255010669701
  • HaughneyJPriceDBarnesNCVirchowJCRocheNChrystynHChoosing inhaler devices for people with asthma: current knowledge and outstanding research needsRespir Med20101041237124520472415
  • LavoriniFLevyMLCorriganCCromptonGThe ADMIT series – Issues in Inhalation Therapy. 6) Training tools for inhalation devicesPrim Care Respir J20101933534121049263
  • van der PalenJPeak inspiratory flow through diskus and turbuhaler, measured by means of a peak inspiratory flow meter (In-Check DIAL)Respir Med20039728528912645837
  • SestiniPCappielloVAlianiMPrescription bias and factors associated with improper use of inhalersJ Aerosol Med20061912713616796537
  • KozyrskyjALDahlMEUngarWJBeckerABLawBJAntibiotic treatment of wheezing in children with asthma: what is the practice?Pediatrics2006117e1104e111016740813
  • VanderweilSGTsaiCLPelletierAJInappropriate use of antibiotics for acute asthma in United States emergency departmentsAcad Emerg Med20081573674318627585