Novel concept of detecting basal cell carcinoma in skin tissue using a continuous-wave millimeter-wave rectangular glass filled probe

&
Pages 275-285 | Published online: 22 Aug 2018

Abstract

Purpose

This article presents the study and simulation results of a millimeter (mm)-wave device for cancerous tissue detection. mm-Wave approach ensures cheaper equipment instead of the traditional terahertz (THz) frequency approach. A probe that could be implemented using inexpensive silicon technology is proposed, and it also permits integration of entire measuring tool for easy deployment. Skin cancer was chosen as it representŝ80% of all newly diagnosed cases and is the most common form of cancer in Australia. For an initial development and validation, due to data availability consideration in the open literature, basal cell carcinoma (BCC) was used for simulations.

Methods and results

A probe, using high-frequency signals in the upper mm-wave frequency spectrum (90–300 GHz) to maximize the lateral resolution (mm precision) and allows the detection of tumors located at up to 0.5 mm deep in the skin, is proposed. A frequency-dependent relativity permittivity and an equivalent conductivity of skins were calculated based on the double Debye parameters. For the first time, electromagnetic (EM) models were generated and used along with a high-frequency EM simulator, ANSYS HFSS, to demonstrate the sensitivity of the concept. The following two scenarios were studied: in scenario one, a BCC layer of different thicknesses (10–3000 μm) was located on the top of the normal skin and, in scenario two, the BCC was embedded in normal skin at depths from 10 to 3000 μm. Variability using ±10% of the corresponding dielectric property was also considered.

Conclusion

This study showed that the reflection coefficients vs frequency could capture useful information indicating the possible presence of BCC at mm-wave frequencies. Both magnitude and phase of the reflection coefficient were quantified, with two scenarios analyzed. It was found that a dual-band approach, 100–150 and 200–250 GHz, has the ability to highlight deviations from the normal skin.

Introduction

Skin cancer is the most common form of cancer in Australia, representinĝ80% of all newly diagnosed cases,Citation1 with melanoma accounted for >10% of all reported incidence, in 2014Citation2 and 2017.Citation3 The majority of skin cancers are detected by patients themselves, or by their partners, while some are detected by physicians during the routine clinical examinations. There is no doubt that early detection could significantly reduce complications or mortality.

There are many patients in Australia who are known to be at moderate-to-high risk of nonmelanoma skin cancer or who have pigmented lesions, which are at risk of developing into melanoma. Today, it is possible to use dermoscopy to determine if the lesions are benign or suspicious and if they require excision for biopsy for conclusive diagnosis. However, it is more often the case that patients have lesions, which do not fulfill the criteria of malignancy but need ongoing monitoring. As with objective blood tests, which are reliable means of monitoring patients at risk of complications (such as deteriorating kidney function), irrespective of the circumstances of their medical examination, ideally, there should exist an accurate mean of assessing the patients for skin cancer too. Under the present health care system, no such objective measure exists.

Prospective total skin photography has been advocated as one method of picking up the onset of melanoma in patients with many atypical moles since it is more often the case that melanoma arises as a new lesion rather than as a changed existing lesion. However, when one has hundreds of moles, many of which are located on the back, it is unlikely that either the patient or the doctor who sees the patient once every 6 months can investigate all lesions without expensive photography and detailed review. If supplementary, detailed dermoscopic images are taken to investigate certain moles, specialized equipment and software are required at extra cost.

Existing imaging techniques for skin cancer diagnosis (such as total body photography and sequential imaging at the macroscopic level) have a low resolution.Citation4,Citation5 At the microscopic level, where higher resolution imaging can be achieved, invasive biopsy samples need to be imaged to determine the type and stage of the skin cancer.

The everlasting desire to improve the quality of life has heightened the competition to develop imaging techniques for health monitoring. Currently, optical coherence tomography, ultrasound, near-IR, and Raman spectroscopies, MRI, positron emission tomography, and X-ray techniques are receiving the most attention.Citation6Citation14 Some of these techniques offer nonionizing solutions, good resolution, good acquisition speeds, and well-targeted contrast mechanisms. However, none of them meets all of these crucial criteria simultaneously. Besides, most of them require costly and bulky equipment and tools.

Hence, it would be advantageous if there were a device, which could be used with a validated read out score, to indicate if particular lesions have reached the “at-risk” stage. This is of particular importance because, in Australia, there is an extensive list of patients waiting to be seen by a dermatologist. As a result, a large number of lesions are unnecessarily removed because of medicolegal and unstated financial incentives.

To overcome the problems above, we propose here the design of a practical small handheld device able to detect the responses of skin conditions (normal and basal cell carcinoma [BCC]) from electromagnetic (EM) radiation at high mm-wave to low terahertz (THz) frequencies. This design represents a preamble for a follow-up sensitive, an integrated readout device, that can be made portable and that allows its deployment everywhere and anytime by any medical staff. A silicon millimeter (mm) mm-wave integrated circuit-type-based technology will ensure the reduced cost for mass deployment and implementation. This first study was conducted heavily relying on the existing open available data on BCC EM parameters. It is hoped that its success will give sufficient experience and background for other lethal cancers’ electronic devices’ studies.

Background

The high mm-wave to low THz signal range was chosen for our studies as it provides some unique features that make it particularly suitable for medical applications. For example, mm-waves have low generated photon energyCitation15,Citation16 not causing chemical damage to molecules or harmful ionization in biological tissues. Studies of high-power EM waves on human mammalian cells observed no significant differences between exposed and nonexposed cells, other than small temperature changes.Citation17 Also, mm-waves have very high absorption rates in water, marking them as responsive to characterize polar substances such as water and hydrated matter.Citation18,Citation19 Mm-waves’ manipulation entails also smaller device sizes that are beneficial for cheap mass production.

We propose a novel device that would permit designing of a system that could quantify the reflected signal from the exposed skin area. The proposed technique is based on shining the skin tissue to a mm-wave so that the complex dielectric properties can be evaluated from this interaction. For different real and imaginary values of different skins, different reflections and absorptions occur, providing unique signatures. McAuley et al and othersCitation20Citation22 showed that mm-wave is able to provide high contrast on tissue depending on their water content. This contrast has been exploited before to map the water distribution in plantsCitation23 and severely burnt skin samples.Citation21 Medical researchers have found that certain tumors have higher water content compared to healthy tissues.Citation24Citation27 Researchers also showed that the water absorption of the mm-wave signal increases significantly with temperature.Citation28 Also, it has been concluded that in vivo tissue generally has higher water content compared to ex vivo tissue.Citation29 Therefore, the in vivo losses are more significant than the equivalent ones in the ex vivo samples at the room temperature.

During the past 2 decades, groups from around the globe conducted research on imaging with mm-waves and THz waves. Most of their approaches used THz pulsed imaging (TPI) with optical mixing to generate higher mm-wave and THz signals and performed the analysis with finite difference time domain method.Citation20,Citation30Citation32 Currently, there is only one commercially available tissue imager, of great significance in the world, produced by TeraView.Citation33 The main drawbacks of using optical mixing systems are that they are bulky, expensive, and immovable. This is because these systems are primarily based on optical sources, mirrors, and lenses. To overcome these problems, more recent research shifts the focus to near-field systems that use mm-waves’ circuits. Sensors operating <40 GHz have been proposed and demonstrated; although they provide well-integrated solutions, their sensitivity is insufficient, due to the relatively large wavelengths.Citation34Citation36 Other systems, operating at much higher frequencieŝ90 GHz, have shown promising resultsCitation22,Citation37 but are very bulky and not suitable for integration in a readout device with the small form factor.

In recent years, mm-wave technology has undergone dramatic advancements mostly as a result of the progress in silicon transistors and integrated circuits, making them suitable for operation throughout the entire mm-wave spectrum (30–300 GHz). Several large volume silicon mm-wave products, such as automotive cruise control radarsCitation38,Citation39 and 60 GHz wireless USB3 transceivers operating at 5 Gb/s rates,Citation40 have demonstrated the feasibility and resilience of mm-wave silicon technology in the market. The costs of mm-wave silicon transceivers have also significantly diminished. Silicon mm-wave integrated circuits are now poised to take on and revolutionize a variety of new applications, where the medical imaging and diagnosis are among the most promising.

Methods

The suggested system consists of the following three parts: a signal generator–detector (eg, vector network analyser [VNA]), the proposed device itself, called “the probe”, and the sample under test (skin area). The general practitioners or qualified personnel can use the probe by scanning it over the visually examined area of investigation. The key innovation of this work consists in designing a probe that offers a small footprint for high lateral resolution sensing with sufficient sensitivity for high contrast between the normal and diseased skin and also in the EM modeling of the BCC using existing data from the literature. The system is designed based on the principle of continuous-wave (CW) near-field EM wave reflectometry. The generated mm-wave signal illuminates the skin sample and travels through the probe; a part of the signal penetrates the sample with the remainder reflected through the probe and reached back the detector. This reflected signal gives unique information (in terms of its amplitude and phase) at different frequencies when compared with the transmitted signal from the generator. The deviances of the magnitude and phase of different skin samples from the normal skin are determined in this article. In the suggested system, the signal generator–detector could be a commercially available VNA, with the final aim to fabricate a fully integrated circuit combining the VNA function with the probe scanning feature, as future work. The probe is designed so that it permits limited penetration depth within typical skin thickness (eg, <3 mm).Citation41,Citation42,Citation43 The normal dry skin and the most common type of skin cancers BCC were considered in this study. For simplicity, a flat and homogeneous skin was assumed. The assumptions are valid provided that the simulated skin area (1.5 mm in diameter) is much smaller than the typical average BCC size, which iŝ3–5 mm in diameter.Citation41 Also, the total skin thickness that consists of normal and BCC layers was limited here to 3 mm, as this is the typical maximum skin (epidermis–dermis) thickness across different body parts.Citation41,Citation42

There have been three steps in the present study. First, we calculated the variation of relative electrical permittivity and total effective conductivity functions vs frequencies. Second, the near-field reflectometry probe was designed. Finally, the combined probe-skin samples (treated as terminated/loaded probes) were studied and followed by their high frequency structure simulator (HFSS) simulations. The results accomplished in these steps are outlined later.

Step one: normal skin and BCC electrical parameters calculation

With CW EM simulations, typical commercially available software packages (eg, HFSS and CST) are unable to use complex frequency-dependent electric permittivity functions. Instead, they require the relative permittivity εr and the electric conductivity σ.

However, human skin dielectric properties are typically modeled as complex frequency-dependent permittivity ε*(ω), based on Debye theory. Pickwell et alCitation44 used double Debye theory and modeled frequency-dependent permittivity ε*(ω), from 0.1 to 4 THz for both normal skin and BCC, where ε*(ω) is given by ε*(ω)=ε+εsε21+jωτ1+ε2ε1+jωτ2(1) where εs, ε, ε2, τ1, and τ2 are the five double Debye parameters.

shows the double Debye parameters that Pickwell obtained for in vivo healthy and BCC skin samples. Pickwell was able to use these parameters to perform simulations in the time domain with optical techniques to verify their designs that ultimately resulted in producing the only commercially available clinical hardware for skin cancer detection using TPI.Citation33

Table 1 Double Debye parameters for normal skin and BCC for the calculation of the frequency-dependent relative electric permittivity εr and total effective conductivity (equivalent conductivity) σe

In our study, we perform EM modeling using Pickwell’s double Debye parameters. EquationEquation 1 was separated into the real (ε′) and imaginary (ε″) parts as in EquationEquation (2): ε*=εjε(2) The real part ε′ represents the relative permittivity or εr, and the imaginary part ε″ was further expressed as total effective conductivity (equivalent conductivity); permittivity is the ability of a substance to store electrical energy in an electric field. Electrical permittivity of free space is a universal constant ε0=8.854×10–12 F/m. Materials are characterized by the electrical permittivity εrε0=ε′ [F/m] and εr is the relative permittivity and is >1. Materials can also be characterized by losses, which include dielectric damping and conductive loss. Together they form ε″ and that can be expressed as the total effective conductivity equivalent conductivity.

σe of the material (skin) as in EquationEquation (3) ε=σe2πfε0(3) where ε0=8.85×10−12 F/m is the permittivity of free space and f is the frequency in Hz.

It should be noted that this total effective conductivity is not just the static (ionic) conductivity; it includes all the dielectric losses such as the standard static (ionic) conductivity and the associated loss (imaginary) terms from double Debye theory.

These calculations have been performed for each frequency point, ranging from 95 to 300 GHz, and are plotted in . In order to account for variability, ±10% of the BCC’s relative permittivity εr and the total effective conductivity σe were also included.

Figure 1 (A) Extracted frequency-dependent relative permittivity εr and (B) total effective conductivity (equivalent conductivity) σe (S/m) for normal dry skin and BCC using the double Debye parameters by Pickwell for frequencies ranging from 95 to 300 GHz.

Abbreviation: BCC, basal cell carcinoma.
Figure 1 (A) Extracted frequency-dependent relative permittivity εr and (B) total effective conductivity (equivalent conductivity) σe (S/m) for normal dry skin and BCC using the double Debye parameters by Pickwell for frequencies ranging from 95 to 300 GHz.

Step two: probe considerations

A typical coaxial type probe could be used as the sensor. Although some coaxial probes are commercially available, they are not suitable for this application as they suffer from having relatively large footprints.Citation45 Typically >3 mm in diameter, the commercial probes, which will also sense surrounding normal skin due to the fringing field, are not able to provide sufficient discrimination between normal skin, the BCC properties, and the surrounding normal skin.Citation46 illustrates the proposed rectangular waveguide probe that supports the transverse electric TE10 mode for the sensing signal and that is easy to fabricate. The rectangular waveguide probe sits on the skin area of interest at one end and is connected to the VNA (generator and detector) through a coplanar waveguide transition at the other end. Front and back views of the skin sample in contact with the probe are shown in .

Figure 2 Probe sitting on the skin sample of interest.

Notes: (A) Front view of the probe connected to the VNA via a CPW transition. (B) Back view showing the rectangular waveguide behind the skin sample.
Abbreviations: CPW, coplanar waveguide; VNA, vector network analyser; WG, waveguide.
Figure 2 Probe sitting on the skin sample of interest.

A higher operational frequency, >90 GHz, is selected based on the lateral resolution and penetration considerations. In rectangular waveguide design, higher frequencies offer smaller waveguide cross-section due to waveguide cutoff frequencyCitation44 and also provide shallower EM wave penetration in the skin. The lateral resolution and EM penetration must be adequately tackled. This is because BCC can be very shallow with a thickness in sub-mm range and can have a lateral dimensions of order of mm2. Glass was selected as the waveguide filling material due to its wide commercial availability for microfabrication. Glass also has a dielectric constant of 5.5, which is the closest to 5.6 value that is attributed to the normal skin,Citation47 allowing the minimization of the impedance mismatch between the probe and normal skin at the contact interface. This permits the mm-wave signal to penetrate into the skin and allows to take place only when mismatchCitation46 occurs. As BCC exhibits departure from the dielectric constant and conductivity values of the normal skin, a mismatch will be encountered and this can be quantified by measuring the reflection coefficient.

With the selection of sensing frequencies >95 GHz and glass as filling material for the probe, a minimal footprint, as small as 0.7 mm (W)×0.35 mm (H) can be achieved. Also, as the sensing frequency is >95 GHz and operation is in the near-field EM radiation condition, the system is interference immune. It will offer interference avoidance from existing far-field communication and military systems, operating at frequencies <100 GHz.

Step three: probe loaded with skin samples study

An entire combination of the probe loaded with different skin samples was simulated in HFSS. The EM signal was excited from the generator (the VNA), has traveled through the probe, was reflected from the skin, and has returned back to the detector (the VNA). Two possible occurring scenarios were considered in order to emulate different BCC conditions.

In scenario 1, a shallow BCC layer of various thicknesses is located on the top of the normal skin, as shown in . shows the EM simulation set-up and the TE10 mode propagation for scenario 1.

Figure 3 Scenario 1: (A) BCC layer (of different thicknesses) on the surface of normal skin; (B) the corresponding EM simulation setup; and (C) illustration of the TE10 propagation mode. Scenario 2: (D) BCC layer embedded beneath normal skin at different depths; (E) the corresponding EM simulation setup; and (F) illustration of the TE10 propagation mode.

Abbreviations: BCC, basal cell carcinoma; EM, electromagnetic.
Figure 3 Scenario 1: (A) BCC layer (of different thicknesses) on the surface of normal skin; (B) the corresponding EM simulation setup; and (C) illustration of the TE10 propagation mode. Scenario 2: (D) BCC layer embedded beneath normal skin at different depths; (E) the corresponding EM simulation setup; and (F) illustration of the TE10 propagation mode.

In scenario 2, a 500 μm thick BCC layer is embedded under the normal skin at various depths, as shown in . shows the EM simulation set-up and the illustration of the TE10 mode propagation for scenario 2.

Results and discussion

The HFSS-simulated reflected signals were analyzed in terms of the magnitude (|S11| in dB) and phase (∠S11 in degrees) of the reflection coefficient, for both scenarios 1 and 2. These results were compared with the magnitude |S11| and phase ∠S11 of the reflection coefficient for the normal skin that are given in .

Figure 4 HFSS simulation results: magnitude |S11| (dB) and phase ∠S11 (degree) of the reflection coefficient for probe loaded with the normal skin.

Figure 4 HFSS simulation results: magnitude |S11| (dB) and phase ∠S11 (degree) of the reflection coefficient for probe loaded with the normal skin.

We note that for frequency ranging between 150 and 300 GHz, the magnitude |S11| takes values <–20 dB. Therefore, in order to effectively emphasize the effectiveness of the proposed system, the results were presented as differences between the magnitudes and phases of the reflection coef ficients where the BCC values were considered first and then the normal skin, ie, Δ|S11| = |S11|BCC – |S11|normal skin and Δ∠S11 = ∠S11BCC – ∠S11normal skin. The study results for scenarios 1 and 2 are illustrated in and , where ±10% variations in relative permittivity εr and total effective conductivity σe were introduced (eg, εr=90% and 110% and σe=90% and 110% of the original calculated functions), to account for possible variability.

Figure 5 Simulation results of scenario 1 (BCC layer with thickness 10, 50, 100, 500, 1000, and 3000 μm on the surface of healthy skin): (A–E) shows differences in magnitude (Δ|S11|) and (F–J) shows phase (Δ∠S11) of the reflection coefficient, with ±10% variation of BCC’s εr and σe.

Abbreviation: BCC, basal cell carcinoma.
Figure 5 Simulation results of scenario 1 (BCC layer with thickness 10, 50, 100, 500, 1000, and 3000 μm on the surface of healthy skin): (A–E) shows differences in magnitude (Δ|S11|) and (F–J) shows phase (Δ∠S11) of the reflection coefficient, with ±10% variation of BCC’s εr and σe.

Figure 6 Simulation results of scenario 2 (BCC layer embedded in healthy skin at depths of 10, 50, 100, 500, 1000, and 3000 μm): (A–E) shows differences in magnitude (Δ|S11|) and (F–J) shows phase (Δ∠S11) of the reflection coefficient, with ±10% variation of BCC’s εr and σe.

Abbreviation: BCC, basal cell carcinoma.
Figure 6 Simulation results of scenario 2 (BCC layer embedded in healthy skin at depths of 10, 50, 100, 500, 1000, and 3000 μm): (A–E) shows differences in magnitude (Δ|S11|) and (F–J) shows phase (Δ∠S11) of the reflection coefficient, with ±10% variation of BCC’s εr and σe.

For scenario 1, shows the differences in magnitude of the reflection coefficient Δ|S11|, while illustrates the phase differences Δ∠S11 of BCC of different thicknesses (10, 50, 100, 500, 1000, and 3000 μm) and normal skin, over the entire frequency range, with ±10% variations in relative permittivity εr and conductivity σe of the BCC. According to these results, even a very shallow BCC layer resulted in a distinguishable difference in magnitude response compared to normal skin. This is even more pronounced for the difference in phase response, where other than the 10 μm thin BCC layer, a large phase deviation from the normal skin phase is observed, in particular >200 GHz.

shows the simulation results for the second scenario, where a 500 μm thick BCC layer is embedded in a normal skin layer at progressive depths 10, 50, 100, 500, 1000, and 3000 μm with ±10% variations in relative permittivity εr and conductivity σe of the BCC. Clearly, notable differences in magnitude |S11| () and phase ∠S11 () values could be captured for all BCC layers embedded at depths from 10 to 100 μm, within these variations. In fact, the shallower the BCC layer is embedded under normal skin tissue, the more differences to the Δ|S11| and Δ∠S11 from those for the normal skin. It should be noted that according to the simulations, as the BCC layer is embedded >500 μm under the normal skin, the responses from both magnitude and phase become similar to normal skin, across the entire 95–300 GHz frequency range. This is, in fact, a desirable feature for the proposed probe. According to researchers,Citation48 the device should only detect unusual tissues that are embedded in the epidermis layer, which has a typical thickness of <100 μm across body sites.

The difference in magnitude shows <2 dB values >200 GHz for scenario 1, and the difference in magnitude shows >3 dB values <200 GHz for scenario 2, as shown in and , respectively. As magnitude |S11| for the reference normal skin is <–20 dB >200 GHz, the absolute magnitudes for both scenarios are onlŷ–20 dB, which correspond to very small values. This is not the case for the differences in phase, Δ∠S11 data, where significant values were obtained, rendering the phase as the preferred parameter for our analysis. Therefore, a dual-band operation (95–150 and 200–250 GHz) is ideal, where magnitude and phase are the preferred parameters, respectively, for the BCC detection. Dual-band approach will not pose any difficulty in silicon technology implementation.Citation49

With the proposed concept, the following advantages could be achieved: The small footprint of the probe for high lateral resolution (~mm) sensing; Controlled sensing depth into the skin, suitable for 100 μm deep embedded BCC; High contrast between normal skin and BCC, for satisfactory sensitivity; Cheap silicon technology implementation; Integration with on-chip VNA for an ultimate small, hand-held portable device that can easily be deployed anywhere.

The present study reveals that the proposed technique has the potential to detect BCC presence; however, at this stage, it cannot provide absolute location and thickness information. In fact, as suggested in Truong et al,Citation50 normal skin and BCC could result in very different dielectric properties depending on the person, the body parts, modeling, and measurement techniques. Therefore, this proposed technique is valid and offers useful information only when BCC is referenced to the adjacent normal skin. Future studies including detailed mathematical modeling and classifications are required in order to further determine the thickness, the location, and dielectric properties over a frequency range.

Conclusion

Studies and simulation results of a mm-wave rectangular glass filled probe using CW were carried out for the detection of skin conditions. In this study, normal skin and BCC skin EM models, with ±10% variation in the BCC complex dielectric properties, were created and simulated with a high-frequency EM simulator ANSYS HFSS. The simulation results validated that the technique provides a satisfactory sensitivity to detect BCC while providing a small footprint and adequate sensing depth. Depending on the frequency range, either amplitude or phase could be the preferred parameter for the detection of maximum contrast between BCC and normal skin. We concluded that a dual-band system in silicon implementation could be ideal for this application. While a VNA was proposed as a transmitter and receiver in conjunction with the probe, the authors hope that the present work can be extended to an advanced fabricated VNA probe integrated into a single small device in the future, providing a preamble to a compact handheld portable device that can be easily deployed and used. It should be noted that although this concept was validated in simulations, a final decisive conclusion of the usefulness of this concept should only be made after performing in vivo experiments. Hence the next step to validate this concept is to perform a preclinical study using animal modelsCitation51 consisting BCC and normal control skins followed by in vitro study with human skin samples. A statistical study should also be performed to validate its sensitivity and specificity.

Disclosure

The authors report no conflicts of interest in this work.

References

  • AIHW, AACRCancer in Australia: An Overview, 20062007 Contract No.: CAN 32
  • AIHWCancer in Australia: In Brief 20142015 Contract No.: CAN 89
  • HirthGImproving public health relating to ultra-violet radiation exposure – innovations and plans at ArpansaScience and the Art of Radiation ProtectionWollongong, Australia2017
  • MalvehyJPuigSFollow-up of melanocytic skin lesions with digital total-body photography and digital dermoscopy: a two-step methodClin Dermatol200220329730412074871
  • MenziesSWEmeryJStaplesMImpact of dermoscopy and short-term sequential digital dermoscopy imaging for the management of pigmented lesions in primary care: a sequential intervention trialBr J Dermatol200916161270127719747359
  • McBainCAHenryAMSykesJX-ray volumetric imaging in image-guided radiotherapy: the new standard in on-treatment imagingInt J Radiat Oncol Biol Phys200664262563416343802
  • KimballABSummersRMTurnerMMagnetic resonance imaging detection of occult skin and subcutaneous abnormalities in juvenile dermatomyositis: implications for diagnosis and therapyArthritis Rheum20004381866187310943878
  • BarataCRuelaMFranciscoMMendoncaTMarquesJSTwo systems for the detection of melanomas in dermoscopy images using texture and color featuresIEEE Sys J201483965979
  • MendiEYogurtcularCSezginYBayrakCAutomatic mobile segmentation of dermoscopy images using density based and fuzzy c-means clusteringIEEE Int Symp Med Meas Appl201416
  • DoTTZhouYZhengHCheungNMKohDEarly melanoma diagnosis with mobile imagingConf Proc IEEE Eng Med Biol Soc201420146752675710.1109/EMBC.2014.694517825571546
  • AbuzaghlehOBarkanaBDFaezipourMAutomated skin lesion analysis based on color and shape geometry feature set for melanoma early detection and preventionIEEE Long Island Systems, Applications and Technology Conference (LISAT)Farmingdale, NY, USA2014
  • WangYXuDYangSXingDToward in vivo biopsy of melanoma based on photoacoustic and ultrasound dual imaging with an integrated detectorBiomed Opt Express20167227928626977339
  • GodoySERamirezDAMyersSADynamic infrared imaging for skin cancer screeningInfrared Phys Technol201570147152
  • MadaniIDuthoyWDerieCPositron emission tomography-guided, focal-dose escalation using intensity-modulated radiotherapy for head and neck cancerInt J Radiat Oncol Biol Phys200768112613517448871
  • BerryEWalkerGCFitzgeraldAJDo in vivo terahertz imaging systems comply with safety guidelines?J Laser Appl2003153192198
  • ZeniOGalleranoGPPerrottaACytogenetic observations in human peripheral blood leukocytes following in vitro exposure to THz radiation: a pilot studyHealth Phys200792434935717351499
  • JansenEDWilminkGJThomasRJQuantitative investigation of the bioeffects associated with terahertz radiationSPIE BiOS2010756275620L
  • LaneJASaxtonJADielectric dispersion in pure polar liquids at very high radio-frequencies. I. Measurements on water, methyl and ethyl alcoholsProc Royal Soc A Math Phys Eng Sci19522131114400408
  • FedericiJFSchulkinBHuangFTHz imaging and sensing for security applications—explosives, weapons and drugsSemicond Sci Technol2005207S266S280
  • McAuleyIYoungLGradzielMMillimetre-wave and terahertz imaging systems with medical applicationsConference Digest of the 2006 Joint 31st International Conference on Infrared and Millimeter Waves and 14th International Conference on Terahertz Electronics2006371
  • MittlemanDMGuptaMNeelamaniRRecent advances in terahertz imagingAppl Phys B Lasers Opt199968610851094
  • TopferFEmtestamLOberhammerJDermatological verification of micromachined millimeter-wave skin-cancer probeIEEE MTT-S Int Microw Symposium (IMS)2014
  • HadjiloucasSKaratzasLSBowenJWMeasurements of leaf water content using terahertz radiationIEEE Trans Microw Theory Tech1999472142149
  • RossKFAGordonREWater in malignant tissue, measured by cell refractometry and nuclear magnetic resonanceJ Microsc198212817217143438
  • SuntzeffVCarruthersCThe water content in the epidermis of mice undergoing carcinogenesis by methylcholanthreneCancer Res1946657457720996757
  • ChenJHAvramHECrooksLEArakawaMKaufmanLBritoACIn vivo relaxation times and hydrogen density at 0.063-4.85 T in rats with implanted mammary adenocarcinomasRadiology199218424274341620841
  • ScheppsJLFosterKRThe UHF and microwave dielectric properties of normal and tumour tissues: variation in dielectric properties with tissue water contentPhys Med Biol1980256114911597208627
  • RonneCThraneLAstrandPOInvestigation of the temperature dependence of dielectric relaxation in liquid water by THz reflection spectroscopy and molecular dynamics simulationJ Chem Phys19971071453195331
  • SasakiKWakeKWatanabeSMeasurement of the dielectric properties of the epidermis and dermis at frequencies from 0.5 GHz to 110 GHzPhys Med Biol201459164739474725082800
  • SiebertKJQuastHLeonhardtRContinuous-wave all-optoelectronic terahertz imagingAppl Phys Lett2002801630033005
  • WoodwardRMWallaceVPArnoneDDLinfieldEHPepperMTera-hertz pulsed imaging of skin cancer in the time and frequency domainJ Biol Phys2003292–325725923345842
  • WallaceVPFitzgeraldAJPickwellETerahertz pulsed spectroscopy of human basal cell carcinomaAppl Spectrosc200660101127113317059664
  • TeraViewStudy to Distinguish Between Healthy Tissue and Basal Cell Carcinoma2017
  • KimJMOhDYoonJIn vitro and in vivo measurement for biological applications using micromachined probeIEEE Trans Microw Theory Tech2005531134153421
  • KimJMChoSKimNPlanar type micromachined probe with low uncertainty at low frequenciesSensor Actuat A Phys20071391–2111117
  • ByoungjoongKParkJHChoJNovel low-cost planar probes with broadside apertures for nondestructive dielectric measurement of biological materials at microwave frequenciesIEEE Trans Microw Theory Tech2005531134143
  • DanylyukSPoppeUKadlecFBroadband microwave-to-terahertz near-field imagingIEEE/MTT-S International Microwave Symposium, 2007Honolulu, HI, USA2007
  • HaschJTopakESchnabelRMillimeter-wave technology for automotive radar sensors in the 77 GHz frequency bandIEEE Trans Microw Theory Tech2012603845860
  • Infineon [webpage on the Internet] Available from: https://www.infi-neon.comAccessed June 26, 2017
  • SiBEAMSiBEAM Introduces USB 3.0 802.11ad Reference Design Delivering Wireless Connectivity at Multi-Gigabit Speeds Available from: http://www.sibeam.com/en/News/2016/20160104SiBEAMUSB3080211ad.aspxAccessed June 26, 2017
  • ShusterSAMBlackMMMcVitieEVAThe influence of age and sex on skin thickness, skin collagen and densityBr J Dermatol19759366396431220811
  • LaurentAMistrettaFBottigioliDEchographic measurement of skin thickness in adults by high frequency ultrasound to assess the appropriate microneedle length for intradermal delivery of vaccinesVaccine200725346423643017640778
  • CignaETaralloMMarucciaMSorvilloVPollastriniAScuderiNBasal cell carcinoma: 10 years of experienceJ Skin Cancer2011201147636221151696
  • PickwellEFitzgeraldAJColeBESimulating the response of terahertz radiation to basal cell carcinoma using ex vivo spectroscopy measurementsJ Biomed Opt200510606402116409086
  • MeaneyPMGregoryAPSeppalaJLahtinenTOpen-ended coaxial dielectric probe effective penetration depth determinationIEEE Trans Microw Theory Tech201664391592327346890
  • PozarDMMicrowave Engineering3 edNew YorkWiley2004
  • HasgallPAGennaroFDBaumgartnerCwebpage on the InternetIT’IS Database for Thermal and Electromagnetic Parameters of Biological Tissues 20152018 Available from: www.itis.ethz.ch/databaseAccessed June 26, 2018
  • Sandby-MollerJPoulsenTWulfHCEpidermal thickness at different body sites: relationship to age, gender, pigmentation, blood content, skin type and smoking habitsActa Derm Venereol200383641041314690333
  • ShopovSBalteanuAHaschJA 234–261-GHz 55-nm SiGe BiCMOS Signal Source with 5.4–7.2 dBm Output Power, 1.3% DC-to-RF efficiency, and 1-GHz Divided-Down OutputIEEE J Solid State Circuits201651920542065
  • TruongBCTuanHDFitzgeraldAJWallaceVPNguyenHTHigh correlation of double Debye model parameters in skin cancer detectionConf Proc IEEE Eng Med Biol Soc2014201471872125570059
  • ChaudharySCWaseemMRanaMNaproxen inhibits UVB-induced basal cell and squamous cell carcinoma development in Ptch1(±)/SKH-1 hairless micePhotochem Photobiol20179341016102428329421